Emissivity-compensated radiation thermometry method using a concave mirror.


Journal

The Review of scientific instruments
ISSN: 1089-7623
Titre abrégé: Rev Sci Instrum
Pays: United States
ID NLM: 0405571

Informations de publication

Date de publication:
01 May 2021
Historique:
entrez: 10 7 2021
pubmed: 11 7 2021
medline: 11 7 2021
Statut: ppublish

Résumé

In this work, a method to simultaneously measure the temperature and emissivity of any specimen, including metals and semiconductor materials, was developed using a concave mirror and a polarized radiometer. The ratio of the enhanced radiance after reflection off the mirror to the original radiance of a specimen allowed for emissivity determination in the range 0.1-0.9, with a variation of 8%. The combined standard uncertainty, u