Focal osteolysis and corrosion at the junction of Precice Stryde intramedullary lengthening device : preliminary clinical, radiological, and metallurgic analysis of 57 lengthened segments.
Intramedullary lengthening nail
Osteolysis
Stryde
Journal
Bone & joint research
ISSN: 2046-3758
Titre abrégé: Bone Joint Res
Pays: England
ID NLM: 101586057
Informations de publication
Date de publication:
Jul 2021
Jul 2021
Historique:
entrez:
16
7
2021
pubmed:
17
7
2021
medline:
17
7
2021
Statut:
ppublish
Résumé
This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors' institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Adjacent to the nail's telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail's telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article:
Identifiants
pubmed: 34269599
doi: 10.1302/2046-3758.107.BJR-2021-0146.R1
pmc: PMC8333033
doi:
Types de publication
Journal Article
Langues
eng
Pagination
425-436Références
Acta Orthop. 2021 Jun 8;:1-7
pubmed: 34102950
J Bone Joint Surg Am. 2001 Apr;83(4):537-48
pubmed: 11315782
Acta Orthop. 2021 Mar 24;:1-6
pubmed: 33757381
J Bone Joint Surg Am. 2015 Sep 2;97(17):1432-40
pubmed: 26333739
Bone Joint J. 2016 Oct;98-B(10):1382-1388
pubmed: 27694593
J Pediatr Orthop. 2021 Jan;41(1):e44-e49
pubmed: 32947442
Bone Joint J. 2019 Sep;101-B(9):1168-1176
pubmed: 31474141
Arch Orthop Trauma Surg. 2021 May 13;:
pubmed: 33983528
J Pediatr Orthop. 2020 Jul;40(6):e420-e423
pubmed: 32501902
Clin Orthop Relat Res. 1997 Apr;(337):281-90
pubmed: 9137201
Bone Joint J. 2021 Jun;103-B(6):1168-1172
pubmed: 34058877
Clin Orthop Relat Res. 1996 Dec;(333):51-75
pubmed: 8981882
JBJS Case Connect. 2021 Mar 8;11(1):
pubmed: 33684083
Acta Orthop. 2014 Jun;85(3):293-8
pubmed: 24758320
Clin Orthop Relat Res. 2014 Dec;472(12):3869-78
pubmed: 24682741
Injury. 2001 Dec;32 Suppl 4:SD129-39
pubmed: 11812486
Acta Orthop. 2021 Feb;92(1):120-127
pubmed: 33106069
Acta Orthop. 2019 Feb;90(1):81-87
pubmed: 30371122
J Pediatr Orthop. 2019 Oct;39(9):458-465
pubmed: 31503231
Spine Deform. 2018 Jan;6(1):48-53
pubmed: 29287817
Expert Rev Med Devices. 2015 May;12(3):231-49
pubmed: 25692375
SICOT J. 2017;3:19
pubmed: 29785927
Dtsch Arztebl Int. 2020 Jun 12;117(24):405-411
pubmed: 32865491
Arch Orthop Trauma Surg. 2021 Jun;141(6):899-905
pubmed: 32458076
Pathol Res Pract. 2014 Dec;210(12):779-86
pubmed: 25454771
Bone Joint Res. 2019 Apr 2;8(3):146-155
pubmed: 30997040
BMC Musculoskelet Disord. 2017 Dec 11;18(1):523
pubmed: 29228956
Bone Joint Res. 2018 Aug 4;7(7):476-484
pubmed: 30123497
Clin Orthop Relat Res. 1997 Oct;(343):135-43
pubmed: 9345218
Acta Orthop. 2020 Dec;91(6):761-769
pubmed: 32835564
Clin Orthop Relat Res. 2012 Apr;470(4):1221-31
pubmed: 22143986