[Application value of liquid crystal digital display goniometer in total hip arthroplasty].
Acetabular anteversion
Arthroplasty, replacement, hip
Digital display goniometer
Direct anterior approach
Journal
Zhongguo gu shang = China journal of orthopaedics and traumatology
ISSN: 1003-0034
Titre abrégé: Zhongguo Gu Shang
Pays: China
ID NLM: 9815790
Informations de publication
Date de publication:
25 Jul 2021
25 Jul 2021
Historique:
entrez:
28
7
2021
pubmed:
29
7
2021
medline:
30
7
2021
Statut:
ppublish
Résumé
To investigate the application value of liquid crystal digital display goniometer in total hip arthroplasty. From January 2018 to December 2019, 83 patients underwent primary total hip arthroplasty, including 28 males and 55 females, aged 42 to 81 (70.4±7.9) years. There were 63 cases of femoral neck fracture and 20 cases of avascular necrosis of femoral head. All patients used liquid crystal digital goniometer to control the anteversion of acetabular cup prosthesis during operation, and CT scanning was used to measure the anteversion of acetabular cup after operation. The two methods were compared to understand the accuracy of using liquid crystal digital goniometer. Postoperative CT measurement showed that the acetabular anteversion of all patients was in the safe area advocated by Lewinnek. The anteversion angle of acetabular cup measured by liquid crystal digital goniometer was 14.20(12.80 to 15.40)°, and the anteversion angle of acetabular cup measured by postoperative CT scan was 14.20 (13.40 to 15.50)°. There was no significant difference between the two ( It is an accurate and reliable method to control the anteversion of acetabular cup with liquid crystal digital display angle instrument, which has a good auxiliary reference value.
Identifiants
pubmed: 34318632
doi: 10.12200/j.issn.1003-0034.2021.07.002
doi:
Types de publication
Journal Article
Langues
chi
Sous-ensembles de citation
IM