INVESTIGATION OF THE TL CHARACTERISTICS OF COMPOSITE PEAK 5 IN THE GLOW CURVE OF LIF:MG,TI (TLD-100) USING NATURALLY AND FURNACE-COOLED SAMPLES FOLLOWING THE 400°C PRE-IRRADIATION ANNEAL.
Journal
Radiation protection dosimetry
ISSN: 1742-3406
Titre abrégé: Radiat Prot Dosimetry
Pays: England
ID NLM: 8109958
Informations de publication
Date de publication:
03 Nov 2021
03 Nov 2021
Historique:
received:
11
07
2021
revised:
08
08
2021
accepted:
10
08
2021
pubmed:
1
9
2021
medline:
6
11
2021
entrez:
31
8
2021
Statut:
ppublish
Résumé
The cooling rate to room temperature following the 400°C pre-irradiation anneal is known to affect the thermoluminescent properties of LiF:Mg,Ti (TLD-100) as a result of migration and clustering of defects during the cooling down process. In this investigation the dose response over an extended dose range from 0.01 to 7000 Gy in both naturally cooled and the much slower furnace-cooled samples has been measured. Glow curve deconvolution based on first-order kinetics is employed to extract the dose response of the various glow peaks. Of especial interest is the behaviour of glow peaks 4, 5a and 5 as a function of dose. The idea is to modify the supralinear dose response of peak 5 from 1 to 30 Gy to a linear behaviour. This dose range is important for clinical therapy and a linear dose response is of substantial advantage leading to both improved accuracy and precision.
Identifiants
pubmed: 34463339
pii: 6360342
doi: 10.1093/rpd/ncab129
doi:
Substances chimiques
Lithium Compounds
0
Fluorides
Q80VPU408O
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
53-59Subventions
Organisme : PAZY Foundation
Organisme : Israeli Ministry of Science and Technology
ID : 3-16316
Informations de copyright
© The Author(s) 2021. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.