Three-dimensional Winston-Lutz test using reusable polyvinyl alcohol-iodide (PVA-I) radiochromic gel dosimeter.
PVA-I
SRS
Winston–Lutz test
gel dosimeter
radiochromic gel
Journal
Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220
Informations de publication
Date de publication:
01 10 2021
01 10 2021
Historique:
received:
26
05
2021
accepted:
16
09
2021
pubmed:
17
9
2021
medline:
16
4
2022
entrez:
16
9
2021
Statut:
epublish
Résumé
Medical linear-accelerator-based stereotactic radiosurgery (SRS) using a stereotactic apparatus or image-guided radiotherapy system for intracranial lesions is performed widely in clinical practice. In general, Winston-Lutz (WL) tests using films or electric portal imaging devices (EPIDs) have been performed as pre-treatment and routine quality assurance (QA) for the abovementioned treatment. Two-dimensional displacements between the radiation isocentre and mechanical isocentre are analysed from the test; therefore, it is difficult to identify the three-dimensional (3D) isocentre position intuitively. In this study, we developed an innovative 3D WL test for SRS-QA using a novel radiochromic gel dosimeter based on a polyvinyl alcohol-iodide (PVA-I) complex that can be reused after annealing. A WL gel phantom that was consisted of the PVA-I gel dosimeter poured into a tall acrylic container and an embedded small tungsten sphere was used as a position detector. A flatbed scanner was used to analyse the isocentre position. The measured 3D isocentre accuracy from the gel-based WL test was within 0.1 mm compared with that obtained from the EPID-based WL test. Furthermore, excellent reusability of the WL gel phantom was observed in long-term SRS isocentre verification, in which clinical SRS cases involving repeated irradiation and annealing were analysed. These results demonstrate the high accuracy and reliable evaluation of the isocentre position using an innovative test. In addition, the clinical-based routine SRS-QA using the PVA-I gel dosimeter demonstrates a highly convenience while affording an easy and fast analysis process.
Identifiants
pubmed: 34530407
doi: 10.1088/1361-6560/ac279d
doi:
Substances chimiques
Iodides
0
Polyvinyl Alcohol
9002-89-5
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
© 2021 Institute of Physics and Engineering in Medicine.