Improved imaging using Mn He-α x rays at OMEGA EP.
Journal
The Review of scientific instruments
ISSN: 1089-7623
Titre abrégé: Rev Sci Instrum
Pays: United States
ID NLM: 0405571
Informations de publication
Date de publication:
01 Sep 2021
01 Sep 2021
Historique:
entrez:
2
10
2021
pubmed:
3
10
2021
medline:
3
10
2021
Statut:
ppublish
Résumé
In this paper, we report on a crystal based x-ray imaging system fielded at the OMEGA EP laser facility. This new system has a pointing accuracy of +/100 μm, a temporal resolution down to 100 ps (depending on backlighter characteristics), variable magnification, and a spatial resolution of 21.9 µm at the object plane at a magnification of 15×. The system is designed to use a crystal along the crystal plane that satisfies the Bragg condition for the x ray of interest. The thin crystal is then bent into a spherical geometry and attached to a glass backing substrate to hold it in the diagnostic, and the x rays are imaged onto a charge coupled device. We report on data acquired with the new Los Alamos National Laboratory supplied spherical quartz crystal to image the Mn He-α 6.15 keV line emission.
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM