Fully automated guideline-compliant diameter measurements of the thoracic aorta on ECG-gated CT angiography using deep learning.

Deep learning aortic aneurysm computed tomography angiography dimensional measurement accuracy observer variation time management

Journal

Quantitative imaging in medicine and surgery
ISSN: 2223-4292
Titre abrégé: Quant Imaging Med Surg
Pays: China
ID NLM: 101577942

Informations de publication

Date de publication:
Oct 2021
Historique:
received: 09 02 2021
accepted: 27 05 2021
entrez: 4 10 2021
pubmed: 5 10 2021
medline: 5 10 2021
Statut: ppublish

Résumé

Manually performed diameter measurements on ECG-gated CT-angiography (CTA) represent the gold standard for diagnosis of thoracic aortic dilatation. However, they are time-consuming and show high inter-reader variability. Therefore, we aimed to evaluate the accuracy of measurements of a deep learning-(DL)-algorithm in comparison to those of radiologists and evaluated measurement times (MT). We retrospectively analyzed 405 ECG-gated CTA exams of 371 consecutive patients with suspected aortic dilatation between May 2010 and June 2019. The DL-algorithm prototype detected aortic landmarks (deep reinforcement learning) and segmented the lumen of the thoracic aorta (multi-layer convolutional neural network). It performed measurements according to AHA-guidelines and created visual outputs. Manual measurements were performed by radiologists using centerline technique. Human performance variability (HPV), MT and DL-performance were analyzed in a research setting using a linear mixed model based on 21 randomly selected, repeatedly measured cases. DL-algorithm results were then evaluated in a clinical setting using matched differences. If the differences were within 5 mm for all locations, the cases was regarded as coherent; if there was a discrepancy >5 mm at least at one location (incl. missing values), the case was completely reviewed. HPV ranged up to ±3.4 mm in repeated measurements under research conditions. In the clinical setting, 2,778/3,192 (87.0%) of DL-algorithm's measurements were coherent. Mean differences of paired measurements between DL-algorithm and radiologists at aortic sinus and ascending aorta were -0.45±5.52 and -0.02±3.36 mm. Detailed analysis revealed that measurements at the aortic root were over-/underestimated due to a tilted measurement plane. In total, calculated time saved by DL-algorithm was 3:10 minutes/case. The DL-algorithm provided coherent results to radiologists at almost 90% of measurement locations, while the majority of discrepent cases were located at the aortic root. In summary, the DL-algorithm assisted radiologists in performing AHA-compliant measurements by saving 50% of time per case.

Sections du résumé

BACKGROUND BACKGROUND
Manually performed diameter measurements on ECG-gated CT-angiography (CTA) represent the gold standard for diagnosis of thoracic aortic dilatation. However, they are time-consuming and show high inter-reader variability. Therefore, we aimed to evaluate the accuracy of measurements of a deep learning-(DL)-algorithm in comparison to those of radiologists and evaluated measurement times (MT).
METHODS METHODS
We retrospectively analyzed 405 ECG-gated CTA exams of 371 consecutive patients with suspected aortic dilatation between May 2010 and June 2019. The DL-algorithm prototype detected aortic landmarks (deep reinforcement learning) and segmented the lumen of the thoracic aorta (multi-layer convolutional neural network). It performed measurements according to AHA-guidelines and created visual outputs. Manual measurements were performed by radiologists using centerline technique. Human performance variability (HPV), MT and DL-performance were analyzed in a research setting using a linear mixed model based on 21 randomly selected, repeatedly measured cases. DL-algorithm results were then evaluated in a clinical setting using matched differences. If the differences were within 5 mm for all locations, the cases was regarded as coherent; if there was a discrepancy >5 mm at least at one location (incl. missing values), the case was completely reviewed.
RESULTS RESULTS
HPV ranged up to ±3.4 mm in repeated measurements under research conditions. In the clinical setting, 2,778/3,192 (87.0%) of DL-algorithm's measurements were coherent. Mean differences of paired measurements between DL-algorithm and radiologists at aortic sinus and ascending aorta were -0.45±5.52 and -0.02±3.36 mm. Detailed analysis revealed that measurements at the aortic root were over-/underestimated due to a tilted measurement plane. In total, calculated time saved by DL-algorithm was 3:10 minutes/case.
CONCLUSIONS CONCLUSIONS
The DL-algorithm provided coherent results to radiologists at almost 90% of measurement locations, while the majority of discrepent cases were located at the aortic root. In summary, the DL-algorithm assisted radiologists in performing AHA-compliant measurements by saving 50% of time per case.

Identifiants

pubmed: 34603980
doi: 10.21037/qims-21-142
pii: qims-11-10-4245
pmc: PMC8408788
doi:

Types de publication

Journal Article

Langues

eng

Pagination

4245-4257

Informations de copyright

2021 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Déclaration de conflit d'intérêts

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://dx.doi.org/10.21037/qims-21-142). JS is an employee of Siemens Healthineers and received personal fees, RK is a consultant for Siemens Healthineers. JS and RK both helped in installation and maintance of the software but were not involved in study design, data analysis or interpretation. They report that they have a patent US2020/0160527Al pending to Siemens Healthineers. The other authors have no conflict of interest to declare.

Références

Gen Thorac Cardiovasc Surg. 2019 Jan;67(1):59-65
pubmed: 29030719
Circulation. 2010 Apr 6;121(13):e266-369
pubmed: 20233780
Med Image Anal. 2012 Aug;16(6):1187-201
pubmed: 22795524
Int J Cardiovasc Imaging. 2013 Feb;29(2):479-88
pubmed: 22864960
Radiographics. 2004 Jan-Feb;24(1):287-97
pubmed: 14730052
Ann Thorac Surg. 2002 Jan;73(1):17-27; discussion 27-8
pubmed: 11834007
J Am Coll Cardiol. 2014 Oct 21;64(16):1725-39
pubmed: 25323262
JAMA. 1998 Dec 9;280(22):1926-9
pubmed: 9851478
Circulation. 2006 Dec 12;114(24):2611-8
pubmed: 17145990
Eur J Radiol. 2021 Jan;134:109424
pubmed: 33259990
AJR Am J Roentgenol. 2009 May;192(5):W255-63
pubmed: 19380532
Radiology. 2020 Mar;294(3):487-489
pubmed: 31891322
Circulation. 2000 Nov 7;102(19 Suppl 3):III35-9
pubmed: 11082359
Clin Imaging. 2016 Sep-Oct;40(5):936-43
pubmed: 27203287
IEEE Trans Pattern Anal Mach Intell. 2019 Jan;41(1):176-189
pubmed: 29990011
Acad Radiol. 2015 Sep;22(9):1191-8
pubmed: 26210525
J Thorac Cardiovasc Surg. 2013 May;145(5):1326-33
pubmed: 23312977
Eur J Vasc Endovasc Surg. 2013 Mar;45(3):241-7
pubmed: 23318135
Eur Radiol. 2019 Sep;29(9):4613-4623
pubmed: 30673817
Ann Thorac Surg. 2011 Sep;92(3):904-12
pubmed: 21723533
J Thorac Imaging. 2018 Nov;33(6):358-365
pubmed: 29979239
Am Heart J. 2011 Jul;162(1):38-46.e1
pubmed: 21742088
Cardiovasc Diagn Ther. 2017 Jun;7(3):317-324
pubmed: 28567357
J Am Coll Cardiol. 2010 Mar 2;55(9):841-57
pubmed: 20185035
JAMA. 2011 Sep 14;306(10):1104-12
pubmed: 21917581

Auteurs

Maurice Pradella (M)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Thomas Weikert (T)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Jonathan I Sperl (JI)

Siemens Healthineers, Siemensstraße 3, 91301 Forchheim, Germany.

Rainer Kärgel (R)

Siemens Healthineers, Siemensstraße 3, 91301 Forchheim, Germany.

Joshy Cyriac (J)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Rita Achermann (R)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Alexander W Sauter (AW)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Jens Bremerich (J)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Bram Stieltjes (B)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Philipp Brantner (P)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.
Regional Hospitals Rheinfelden and Laufenburg, Riburgerstrasse 12, 4310 Rheinfelden, Switzerland.

Gregor Sommer (G)

Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031 Basel, Switzerland.

Classifications MeSH