[Research on Patient Monitoring Alarms in ICU and NICU].
alarm fatigue
artificial intelligence
critical care
patient monitor
status alarm
Journal
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
ISSN: 1671-7104
Titre abrégé: Zhongguo Yi Liao Qi Xie Za Zhi
Pays: China
ID NLM: 9426153
Informations de publication
Date de publication:
30 Sep 2021
30 Sep 2021
Historique:
entrez:
10
10
2021
pubmed:
11
10
2021
medline:
13
10
2021
Statut:
ppublish
Résumé
The patient monitors were used to explore the alarm actuality in a ICU and NICU to investigate the awareness and reaction of medical staff to alarms. A series of surveys and interviews were taken to acquire clinicians' feelings and attitudes to monitoring alarms. The researchers were scheduled to track the alarms with annotations, and collect the alarm data of patient monitors using central monitoring system. A total of 235 387 and 67 783 alarms occurred in ICU and NICU respectively. The average alarm rate was about 142 alarms/patient-day in ICU and 96 alarms/patient-day in NICU. There remains alarm fatigue in ICU and NICU, the main reason is the large number of false alarms and clinically irrelevant alarms. In addition, patient monitor is still in the level of threshold alarms or combined alarms, the data integrity and intelligence level need to be improved in future.