Classification of X-Ray Attenuation Properties of Additive Manufacturing and 3D Printing Materials Using Computed Tomography From 70 to 140 kVp.
3D printing
additive manufacturing
computed tomography
phantom materials
quality control
radiographic phantoms
x-ray
Journal
Frontiers in bioengineering and biotechnology
ISSN: 2296-4185
Titre abrégé: Front Bioeng Biotechnol
Pays: Switzerland
ID NLM: 101632513
Informations de publication
Date de publication:
2021
2021
Historique:
received:
08
09
2021
accepted:
25
10
2021
entrez:
16
12
2021
pubmed:
17
12
2021
medline:
17
12
2021
Statut:
epublish
Résumé
Additive manufacturing and 3D printing is particularly useful in the production of phantoms for medical imaging applications including determination and optimization of (diagnostic) image quality and dosimetry. Additive manufacturing allows the leap from simple slab and stylized to (pseudo)-anthropomorphic phantoms. This necessitates the use of materials with x-ray attenuation as close as possible to that of the tissues or organs mimicked. X-ray attenuation properties including their energy dependence were determined for 35 printing materials comprising photocured resins and thermoplastic polymers. Prior to measuring x-ray attenuation in CT from 70 to 140 kVp, printing parameters were thoroughly optimized to ensure maximum density avoiding too low attenuation due to microscopic or macroscopic voids. These optimized parameters are made available. CT scanning was performed in a water filled phantom to guarantee defined scan conditions and accurate HU value determination. The spectrum of HU values covered by polymers printed using fused deposition modeling reached from -258 to +1,063 at 120 kVp (-197 to +1,804 at 70 kVp, to -266 to +985 at 140 kVp, respectively). Photocured resins covered 43 to 175 HU at 120 kVp (16-156 at 70, and 57-178 at 140 kVp). At 120 kVp, ASA mimics water almost perfectly (+2 HU). HIPS (-40 HU) is found close to adipose tissue. In all photocurable resins, and 17 printing filaments HU values decreased with increasing beam hardness contrary to soft tissues except adipose tissue making it difficult to mimic water or average soft tissue in phantoms correctly over a range of energies with one single printing material. Filled filaments provided both, the HU range, and an appropriate energy dependence mimicking bone tissues. A filled material with almost constant HU values was identified potentially allowing mimicking soft tissues by reducing density using controlled under-filling. The measurements performed in this study can be used to design phantoms with a wide range of x-ray contrasts, and energy dependence of these contrasts by combining appropriate materials. Data provided on the energy dependence can also be used to correct contrast or contrast to noise ratios from phantom measurements to real tissue contrasts or CNRs.
Identifiants
pubmed: 34912790
doi: 10.3389/fbioe.2021.763960
pii: 763960
pmc: PMC8666890
doi:
Types de publication
Journal Article
Langues
eng
Pagination
763960Informations de copyright
Copyright © 2021 Ma, Buschmann, Unger and Homolka.
Déclaration de conflit d'intérêts
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Références
J Digit Imaging. 2020 Apr;33(2):456-464
pubmed: 31520278
Front Bioeng Biotechnol. 2020 May 08;8:385
pubmed: 32457883
Phys Med Biol. 2018 Sep 06;63(17):175020
pubmed: 29999497
Med Phys. 1992 Mar-Apr;19(2):325-7
pubmed: 1584125
Med Phys. 2015 Jul;42(7):4116-26
pubmed: 26133612
Br J Radiol. 1977 Nov;50(599):814-21
pubmed: 588903
Biomed Tech (Berl). 1988 Apr;33(4):73-6
pubmed: 3395653
Med Phys. 2018 Oct;45(10):4600-4606
pubmed: 30144100
Med Phys. 2020 Apr;47(4):1509-1522
pubmed: 32026482
Phys Med Biol. 2017 Apr 21;62(8):3158-3174
pubmed: 28192281
Phys Med Biol. 2002 Aug 21;47(16):2907-16
pubmed: 12222854
Med Phys. 2011 Feb;38(2):891-6
pubmed: 21452726
Med Eng Phys. 2011 Jun;33(5):590-6
pubmed: 21239209
Phys Med Biol. 2002 Feb 7;47(3):N47-52
pubmed: 11848127
Phys Med Biol. 2017 Jan 10;62(3):758-780
pubmed: 28072573
Sci Rep. 2019 Oct 7;9(1):14357
pubmed: 31591433
Eur Radiol. 2018 Jul;28(7):2951-2959
pubmed: 29460076
J Digit Imaging. 2017 Oct;30(5):572-575
pubmed: 28168529
Med Phys. 2018 Jun 22;:
pubmed: 29933508
Radiology. 2004 Jul;232(1):115-25
pubmed: 15220497
Phys Med. 2017 Jun;38:111-118
pubmed: 28610691
Appl Radiat Isot. 1999 Jan;50(1):247-59
pubmed: 10028641
Phys Med Biol. 2019 Oct 23;64(21):215005
pubmed: 31469105
Phys Eng Sci Med. 2020 Jun;43(2):701-710
pubmed: 32524450
Med Phys. 2018 Apr;45(4):1614-1621
pubmed: 29493803