Variability of respiratory rate measurements in neonates- every minute counts.


Journal

BMC pediatrics
ISSN: 1471-2431
Titre abrégé: BMC Pediatr
Pays: England
ID NLM: 100967804

Informations de publication

Date de publication:
03 01 2022
Historique:
received: 22 09 2021
accepted: 21 12 2021
entrez: 4 1 2022
pubmed: 5 1 2022
medline: 15 3 2022
Statut: epublish

Résumé

Respiratory rate is difficult to measure, especially in neonates who have an irregular breathing pattern. The World Health Organisation recommends a one-minute count, but there is limited data to support this length of observation. We sought to evaluate agreement between the respiratory rate (RR) derived from capnography in neonates, over 15 s, 30 s, 120 s and 300 s, against the recommended 60 s. Neonates at two hospitals in Nairobi were recruited and had capnograph waveforms recorded using the Masimo Rad 97. A single high quality 5 min epoch was randomly chosen from each subject. For each selected epoch, the mean RR was calculated using a breath-detection algorithm applied to the waveform. The RR in the first 60 s was compared to the mean RR measured over the first 15 s, 30 s, 120 s, full 300 s, and last 60 s. We calculated bias and limits of agreement for each comparison and used Bland-Altman plots for visual comparisons. A total of 306 capnographs were analysed from individual subjects. The subjects had a median gestation age of 39 weeks with slightly more females (52.3%) than males (47.7%). The majority of the population were term neonates (70.1%) with 39 (12.8%) having a primary respiratory pathology. There was poor agreement between all the comparisons based on the limits of agreement [confidence interval], ranging between 11.9 [- 6.79 to 6.23] breaths per minute in the one versus 2 min comparison, and 34.7 [- 17.59 to 20.53] breaths per minute in the first versus last minute comparison. Worsening agreement was observed in plots with higher RRs. Neonates have high variability of RR, even over a short period of time. A slight degradation in the agreement is noted over periods shorter than 1 min. However, this is smaller than observations done 3 min apart in the same subject. Longer periods of observation also reduce agreement. For device developers, precise synchronization is needed when comparing devices to reduce the impact of RR variation. For clinicians, where possible, continuous or repeated monitoring of neonates would be preferable to one time RR measurements.

Sections du résumé

BACKGROUND
Respiratory rate is difficult to measure, especially in neonates who have an irregular breathing pattern. The World Health Organisation recommends a one-minute count, but there is limited data to support this length of observation. We sought to evaluate agreement between the respiratory rate (RR) derived from capnography in neonates, over 15 s, 30 s, 120 s and 300 s, against the recommended 60 s.
METHODS
Neonates at two hospitals in Nairobi were recruited and had capnograph waveforms recorded using the Masimo Rad 97. A single high quality 5 min epoch was randomly chosen from each subject. For each selected epoch, the mean RR was calculated using a breath-detection algorithm applied to the waveform. The RR in the first 60 s was compared to the mean RR measured over the first 15 s, 30 s, 120 s, full 300 s, and last 60 s. We calculated bias and limits of agreement for each comparison and used Bland-Altman plots for visual comparisons.
RESULTS
A total of 306 capnographs were analysed from individual subjects. The subjects had a median gestation age of 39 weeks with slightly more females (52.3%) than males (47.7%). The majority of the population were term neonates (70.1%) with 39 (12.8%) having a primary respiratory pathology. There was poor agreement between all the comparisons based on the limits of agreement [confidence interval], ranging between 11.9 [- 6.79 to 6.23] breaths per minute in the one versus 2 min comparison, and 34.7 [- 17.59 to 20.53] breaths per minute in the first versus last minute comparison. Worsening agreement was observed in plots with higher RRs.
CONCLUSIONS
Neonates have high variability of RR, even over a short period of time. A slight degradation in the agreement is noted over periods shorter than 1 min. However, this is smaller than observations done 3 min apart in the same subject. Longer periods of observation also reduce agreement. For device developers, precise synchronization is needed when comparing devices to reduce the impact of RR variation. For clinicians, where possible, continuous or repeated monitoring of neonates would be preferable to one time RR measurements.

Identifiants

pubmed: 34980049
doi: 10.1186/s12887-021-03087-z
pii: 10.1186/s12887-021-03087-z
pmc: PMC8722355
doi:

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

16

Informations de copyright

© 2021. The Author(s).

Références

Ann Emerg Med. 2005 Jan;45(1):68-76
pubmed: 15635313
BMJ Open. 2020 Apr 12;10(4):e035184
pubmed: 32284391
J Clin Epidemiol. 1994 Sep;47(9):1043-9
pubmed: 7730907
EClinicalMedicine. 2019 Jun 27;12:6-7
pubmed: 31388657
Paediatr Respir Rev. 2014 Jun;15(2):163-9
pubmed: 23523390
Am J Respir Crit Care Med. 2019 Apr 15;199(8):1036-1037
pubmed: 30673505
Paediatr Respir Rev. 2015 Sep;16(4):276-84
pubmed: 26364005
Anesth Analg. 2019 Aug;129(2):399-408
pubmed: 30234539
BMJ. 2012 Jul 03;345:e4249
pubmed: 22761089
Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:3131-4
pubmed: 23366589
Clin Chest Med. 2014 Sep;35(3):457-67
pubmed: 25156762
EClinicalMedicine. 2019 Jun 10;12:20-30
pubmed: 31388660
Arch Dis Child. 1991 Oct;66(10):1199-203
pubmed: 1953002
Early Hum Dev. 1981 Feb;5(1):39-54
pubmed: 7193566
JAMA. 2017 Aug 1;318(5):462-471
pubmed: 28763554

Auteurs

Catherine Muthoni Njeru (CM)

The Aga Khan University Hospital, 3rd Parklands Avenue, Nairobi, Kenya. njerucathy@gmail.com.

J Mark Ansermino (JM)

The University of British Columbia, 4480 Oak Street, Vancouver, British Columbia, Canada.

William M Macharia (WM)

The Aga Khan University Hospital, 3rd Parklands Avenue, Nairobi, Kenya.

Dustin T Dunsmuir (DT)

The University of British Columbia, 4480 Oak Street, Vancouver, British Columbia, Canada.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH