Preliminary Design and Development of a Mechanical Ventilator Using Industrial Automation Components for Rapid Deployment During the COVID-19 Pandemic.

covid 19 industrial automation mechanical ventilator rapid development real time operation

Journal

Cureus
ISSN: 2168-8184
Titre abrégé: Cureus
Pays: United States
ID NLM: 101596737

Informations de publication

Date de publication:
Dec 2021
Historique:
accepted: 12 12 2021
entrez: 17 1 2022
pubmed: 18 1 2022
medline: 18 1 2022
Statut: epublish

Résumé

The novel coronavirus disease 2019 (COVID-19) created a shortage of mechanical ventilators in the healthcare sector, resulting in rationed distribution, ethical dilemmas, and high mortalities. This technical report outlines the design and product outcome of a mechanical ventilator based on readily available off-the-shelf components, minimizing the dependence on manufacturing facilities. The ventilator was designed to operate in both hospitals and remote locations, having the ability to operate off various gas pressures and low voltage supplies. Due to the COVID-19 restrictions, the challenges of developing a device in an online setting with minimal manufacturing assistance were explored. Within a 10-day period, the team designed, prototyped, and conducted preliminary feasibility testing on the mechanical ventilator. The proposed design was not intended to replace, or be used as a medically approved ventilator, but demonstrates the ability to exploit off-the-shelf components to enable fast development and assembly.

Identifiants

pubmed: 35036217
doi: 10.7759/cureus.20386
pmc: PMC8752376
doi:

Types de publication

Journal Article

Langues

eng

Pagination

e20386

Informations de copyright

Copyright © 2021, DeBoer et al.

Déclaration de conflit d'intérêts

The authors have declared that no competing interests exist.

Références

Ann Intensive Care. 2017 Dec;7(1):68
pubmed: 28624991
F1000Res. 2020 Mar 30;9:218
pubmed: 32411358
JAMA. 2020 May 12;323(18):1839-1841
pubmed: 32215647
N Engl J Med. 2020 Apr 30;382(18):e41
pubmed: 32212516
Med Devices Sens. 2020 Jun 05;:e10106
pubmed: 32838208
Cureus. 2015 Nov 02;7(11):e371
pubmed: 26677421
Health Devices. 1998 Sep-Oct;27(9-10):363-4
pubmed: 9809257
Can J Respir Ther. 2020 Sep 28;56:42-51
pubmed: 33043131
Crit Care. 2015 Apr 01;19:133
pubmed: 25888116
EuroMediterr J Environ Integr. 2020;5(2):23
pubmed: 32391413
Int J Qual Health Care. 2021 Jan 12;33(Supplement_1):4-10
pubmed: 32780821

Auteurs

Benjamin DeBoer (B)

Faculty of Engineering and Applied Science, Ontario Tech University, Oshawa, CAN.

Ahmad Barari (A)

Faculty of Engineering and Applied Science, Ontario Tech University, Oshawa, CAN.

Mika Nonoyama (M)

Faculty of Health Sciences, Ontario Tech University, Oshawa, CAN.

Adam Dubrowski (A)

Faculty of Health Sciences, Ontario Tech University, Oshawa, CAN.

Marco Zaccagnini (M)

Respiratory Therapy, McGill University, Montreal, CAN.

Ali Hosseini (A)

Faculty of Engineering and Applied Science, Ontario Tech University, Oshawa, CAN.

Classifications MeSH