Résumé

Current approaches to detect the positions and orientations of directional deep brain stimulation (DBS) electrodes rely on radiative imaging data. In this study, we aim to present an improved version of a radiation-free method for magnetic detection of the position and the orientation (MaDoPO) of directional electrodes based on a series of magnetoencephalography (MEG) measurements and a possible future solution for optimized results using emerging on-scalp MEG systems. A directional DBS system was positioned into a realistic head-torso phantom and placed in the MEG scanner. A total of 24 measurements of 180 s each were performed with different predefined electrode configurations. Finite element modeling and model fitting were used to determine the position and orientation of the electrode in the phantom. Related measurements were fitted simultaneously, constraining solutions to the a priori known geometry of the electrode. Results were compared with the results of the high-quality CT imaging of the phantom. The accuracy in electrode localization and orientation detection depended on the number of combined measurements. The localization error was minimized to 2.02 mm by considering six measurements with different non-directional bipolar electrode configurations. Another six measurements with directional bipolar stimulations minimized the orientation error to 4°. These values are mainly limited due to the spatial resolution of the MEG. Moreover, accuracies were investigated as a function of measurement time, number of sensors, and measurement direction of the sensors in order to define an optimized MEG device for this application. Although MEG introduces inaccuracies in the detection of the position and orientation of the electrode, these can be accepted when evaluating the benefits of a radiation-free method. Inaccuracies can be further reduced by the use of on-scalp MEG sensor arrays, which may find their way into clinics in the foreseeable future.

Sections du résumé

BACKGROUND BACKGROUND
Current approaches to detect the positions and orientations of directional deep brain stimulation (DBS) electrodes rely on radiative imaging data. In this study, we aim to present an improved version of a radiation-free method for magnetic detection of the position and the orientation (MaDoPO) of directional electrodes based on a series of magnetoencephalography (MEG) measurements and a possible future solution for optimized results using emerging on-scalp MEG systems.
METHODS METHODS
A directional DBS system was positioned into a realistic head-torso phantom and placed in the MEG scanner. A total of 24 measurements of 180 s each were performed with different predefined electrode configurations. Finite element modeling and model fitting were used to determine the position and orientation of the electrode in the phantom. Related measurements were fitted simultaneously, constraining solutions to the a priori known geometry of the electrode. Results were compared with the results of the high-quality CT imaging of the phantom.
RESULTS RESULTS
The accuracy in electrode localization and orientation detection depended on the number of combined measurements. The localization error was minimized to 2.02 mm by considering six measurements with different non-directional bipolar electrode configurations. Another six measurements with directional bipolar stimulations minimized the orientation error to 4°. These values are mainly limited due to the spatial resolution of the MEG. Moreover, accuracies were investigated as a function of measurement time, number of sensors, and measurement direction of the sensors in order to define an optimized MEG device for this application.
CONCLUSION CONCLUSIONS
Although MEG introduces inaccuracies in the detection of the position and orientation of the electrode, these can be accepted when evaluating the benefits of a radiation-free method. Inaccuracies can be further reduced by the use of on-scalp MEG sensor arrays, which may find their way into clinics in the foreseeable future.

Identifiants

pubmed: 35053829
pii: brainsci12010086
doi: 10.3390/brainsci12010086
pmc: PMC8774199
pii:
doi:

Types de publication

Journal Article

Langues

eng

Subventions

Organisme : Deutsche Forschungsgemeinschaft
ID : SFB1261

Références

Front Hum Neurosci. 2021 Apr 19;15:644593
pubmed: 33953663
Stereotact Funct Neurosurg. 2018;96(5):335-341
pubmed: 30481772
Mov Disord. 2016 Aug;31(8):1240-3
pubmed: 27241197
Zentralbl Neurochir. 2008 Aug;69(3):144-7
pubmed: 18666049
IEEE Trans Biomed Eng. 1992 Jun;39(6):541-57
pubmed: 1601435
Stereotact Funct Neurosurg. 2018;96(6):364-369
pubmed: 30566953
Comput Intell Neurosci. 2011;2011:156869
pubmed: 21253357
Oper Neurosurg (Hagerstown). 2019 Apr 1;16(4):465-470
pubmed: 29920586
Brain Stimul. 2020 Mar - Apr;13(2):378-385
pubmed: 31786180
Neuron. 2013 Feb 6;77(3):406-24
pubmed: 23395370
Front Physiol. 2019 Jan 04;9:1788
pubmed: 30662407
J Neurosci Methods. 2017 Jan 30;276:38-45
pubmed: 27887969
Neuroimage. 2020 Nov 1;221:117157
pubmed: 32659354
J Neural Eng. 2021 Mar 02;18(2):
pubmed: 33503598
Sensors (Basel). 2021 Apr 04;21(7):
pubmed: 33916581
Brain. 2014 Jul;137(Pt 7):2015-26
pubmed: 24844728
Brain Stimul. 2008 Jan;1(1):7-15
pubmed: 19142235
J Neural Eng. 2021 Oct 14;18(5):
pubmed: 34598173
Front Neurol. 2020 Oct 30;11:593798
pubmed: 33193061
Front Neurol. 2021 Nov 08;12:785529
pubmed: 34819915
Neuromodulation. 2021 May 27;:
pubmed: 34047410
Hum Brain Mapp. 2021 Oct 15;42(15):4844-4856
pubmed: 34327772
AJNR Am J Neuroradiol. 2017 Jun;38(6):1111-1116
pubmed: 28385887
Neuroimage. 2020 Oct 1;219:116995
pubmed: 32480036
Med Phys. 2017 Sep;44(9):4463-4473
pubmed: 28639387
Neuroimage. 2015 Feb 15;107:127-135
pubmed: 25498389
Nature. 2018 Mar 29;555(7698):657-661
pubmed: 29562238
Stereotact Funct Neurosurg. 2018;96(5):347-348
pubmed: 30304735
Neuroimage. 2019 Jan 1;184:293-316
pubmed: 30179717
Brain Sci. 2021 Oct 31;11(11):
pubmed: 34827449

Auteurs

Mevlüt Yalaz (M)

Microwave Engineering, Christian-Albrechts-Universität zu Kiel, 24143 Kiel, Germany.

Nicholas Maling (N)

Boston Scientific Corporation, Santa Clarita, CA 91355, USA.

Günther Deuschl (G)

Department of Neurology, Christian-Albrechts-Universität zu Kiel, 24105 Kiel, Germany.

León M Juárez-Paz (LM)

Boston Scientific Corporation, Santa Clarita, CA 91355, USA.

Markus Butz (M)

Institute of Clinical Neuroscience and Medical Psychology, Medical Faculty, Heinrich-Heine-Universität Düsseldorf, 40225 Düsseldorf, Germany.

Alfons Schnitzler (A)

Institute of Clinical Neuroscience and Medical Psychology, Medical Faculty, Heinrich-Heine-Universität Düsseldorf, 40225 Düsseldorf, Germany.

Ann-Kristin Helmers (AK)

Department of Neurosurgery, Christian-Albrechts-Universität zu Kiel, 24105 Kiel, Germany.

Michael Höft (M)

Microwave Engineering, Christian-Albrechts-Universität zu Kiel, 24143 Kiel, Germany.

Classifications MeSH