Typical doses and typical values for fluoroscopic diagnostic and interventional procedures.
diagnostic reference level
dosimetric comparison
dosimetry
fluoroscopy
interventional radiology
typical doses
typical values
Journal
Journal of radiological protection : official journal of the Society for Radiological Protection
ISSN: 1361-6498
Titre abrégé: J Radiol Prot
Pays: England
ID NLM: 8809257
Informations de publication
Date de publication:
09 Mar 2022
09 Mar 2022
Historique:
received:
25
10
2021
accepted:
07
02
2022
pubmed:
8
2
2022
medline:
1
4
2022
entrez:
7
2
2022
Statut:
epublish
Résumé
To implement typical doses (TD) and typical values (TV) for fluoroscopic diagnostic and interventional procedures. A total of 3811 fluoroscopic procedures performed within 34 months on three devices were included in this retrospective study. Dose-, patient- and procedure-related information were extracted using the institutional dose management system (DMS). TD/TV were defined as median dose and calculated for the five most frequent procedures per device for dose area product (DAP), cumulative air kerma (CAK) and fluoroscopy time (FT). National diagnostic reference levels and other single facility studies were compared to our results. Additionally, the five procedures with the highest doses of each device were analysed. To evaluate the data coverage of the DMS compared to the picture archiving and communication system (PACS), procedure lists were extracted from the PACS and compared to the procedure information extracted from the DMS. TD/TV for 15 procedures were implemented. Among all devices, TD for DAP ranged between 0.6 Gycm
Identifiants
pubmed: 35130526
doi: 10.1088/1361-6498/ac5294
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
Creative Commons Attribution license.