Remote sensing array (RSA) for linac beam monitoring.

IMRT QA MLC beam monitoring detector arrays dosimetry linac QA sensors

Journal

Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220

Informations de publication

Date de publication:
24 02 2022
Historique:
received: 23 08 2021
accepted: 08 02 2022
pubmed: 9 2 2022
medline: 6 5 2022
entrez: 8 2 2022
Statut: epublish

Résumé

The purpose of the present work is to evaluate the feasibility of a novel real-time beam monitoring device for medical linacs which remotely senses charge carriers produced in air by the beam without intersecting and attenuating the beamline. The primary goal is to elaborate a theoretical concept of a possible detector geometry and underlying physical model that allows for determination of clinically relevant beam data in real time, namely MLC leaf positions and dose rate. The detector consists of two opposing electrode arrays arranged in two possible orientations around the beamline. Detection of charge carriers is governed by electromagnetic principles described by Shockley-Ramo theorem. Ions produced by ionization of the air column upstream of patient move laterally in an external electric field. According to the method of images, mirror charges and mirror currents are formed in the strip electrodes. Determination of MU rate and MLC positions using the measured signal requires solution of an inverse problem. In the present work we adopted a Least-Square approach and characterized detector response and sensitivity to detection of beam properties for different electrode geometries and MLC shapes. Results were dependent on MLC field shape and the leaf position within the active volume. The accuracy of determination of leaf positions were in the sub-mm range (up to 0.25-1 mm). Additionally, detector sensitivity was quantified by simulating ions/pulse delivered with a radiation transport deterministic computation in 1D in CEPXS/ONEDANT. For a 6 MV linac pulse, signal amplitude per pulse was estimated to be in the lower pA to fA range. We computationally demonstrated feasibility of the remote sensing detector capable of measuring beam parameters such as MLC leaf positions and dose range for each pulse. Future work should focus on optimizing the electrode geometry to increase sensitivity and better reconstruction algorithms to provide more accurate solutions of the inverse problem.

Identifiants

pubmed: 35134790
doi: 10.1088/1361-6560/ac530d
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

© 2022 Institute of Physics and Engineering in Medicine.

Auteurs

Robert Lauber (R)

Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States of America.
Heidelberg University, Heidelberg, Germany.
University of Massachusetts Lowell, MA, United States of America.
Department of Radiation Oncology, University Hospital Bonn, Rheinische Friedrich-Wilhelms-Universität Bonn, Bonn, Germany.

Davide Brivio (D)

Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States of America.

Erno Sajo (E)

University of Massachusetts Lowell, MA, United States of America.

Jürgen Hesser (J)

Heidelberg University, Heidelberg, Germany.

Piotr Zygmanski (P)

Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States of America.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH