Repeatability and Agreement of a New Scheimpflug Device and a Hartmann-Shack Aberrometer With a Ray-Tracing Aberrometer in Normal, Keratoconus, and CXL Groups.


Journal

Journal of refractive surgery (Thorofare, N.J. : 1995)
ISSN: 1938-2391
Titre abrégé: J Refract Surg
Pays: United States
ID NLM: 9505927

Informations de publication

Date de publication:
Mar 2022
Historique:
entrez: 11 3 2022
pubmed: 12 3 2022
medline: 16 3 2022
Statut: ppublish

Résumé

To assess repeatability and agreement of a Scheimpflug topographer and a Hartmann-Shack aberrometer with a ray-tracing aberrometer in normal, keratoconus, and corneal cross-linking (CXL) groups. In this cross-sectional, observational study, normal, keratoconus, and CXL groups with 125 eyes in each of the three groups were included. All eyes underwent three sets of measurements, at a single visit, using the Pentacam AXL Wave (Oculus Optikgeräte GmbH) and iTrace (Tracey Technologies). Keratometry, pachymetry, objective refraction, and total ocular aberrations including root mean square (RMS) lower order aberrations (LOAs), RMS higher order aberrations (HOAs), RMS coma, and spherical aberrations (SA) were analyzed. Objective assessment of haze after CXL was done using the Pentacam AXL Wave. Repeatability was assessed using within-subject standard deviation (Sw), test-retest variability, within-subject coefficient of variation (COV), and intraclass correlation coefficient (ICC). Bland-Altman analysis assessed 95% limits of agreement and correlation coefficient ( The Pentacam AXL Wave had Sw for keratometry of 0.12 in the normal group and 0.15 in the keratoconus group and lower (poorer) Sw of 0.17 in the CXL group. For pachymetry, Sw was 9.18, 9.53, and 10.11 in the normal, keratoconus, and CXL groups, respectively. Total aberrations had ICCs ranging from 0.88 for RMS HOAs to 0.95 for SA in the normal group, 0.86 for RMS HOAs to 0.92 for SA in the keratoconus group, and 0.72 for RMS HOAs to 0.82 for SA (poorer) in the CXL group. In the normal group, mean differences between the two devices were nonsignificant for all parameters except SA (0.011 ± 0.038 μm, The Pentacam AXL Wave showed comparable repeatability in the normal and keratoconus groups, but was poorer in the CXL group, more so with increasing corneal haze. Both devices can be used interchangeably in normal eyes, except for SA, but not in eyes with keratoconus or CXL for aberration measurements.

Identifiants

pubmed: 35275005
doi: 10.3928/1081597X-20220110-01
doi:

Types de publication

Journal Article Observational Study

Langues

eng

Sous-ensembles de citation

IM

Pagination

201-208

Auteurs

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH