Influence of screw head diameter on ex vivo fixation of equine lateral condylar fractures with 5.5 mm cortical screws.


Journal

Veterinary surgery : VS
ISSN: 1532-950X
Titre abrégé: Vet Surg
Pays: United States
ID NLM: 8113214

Informations de publication

Date de publication:
May 2022
Historique:
revised: 01 02 2022
received: 29 01 2021
accepted: 13 02 2022
pubmed: 19 3 2022
medline: 27 4 2022
entrez: 18 3 2022
Statut: ppublish

Résumé

To determine the influence of screw head diameter on equine condylar fracture fixation with 5.5 mm cortical screws. Ex vivo, biomechanical study, blinded, matched-pair design. Fifteen pairs of equine third metacarpal (MC3) bones. Lateral condylar fractures were simulated by parasagittal osteotomies and repaired pairwise by 2 × 5.5 mm cortical screws of 8 mm (standard) or 10 mm (modified) head diameter. Interfragmentary compression at maximum screw insertion torque was measured. The instrumented specimens were pairwise stratified for biomechanical testing under the following modalities (n = 5): (1) screw insertion torque to failure, (2) quasi-static axial load to failure, and (3) cyclic axial load to 2 mm displacement followed by failure. Tests (1) and (2) were analyzed for yield, maximum, and failure torque/angle and load/displacement, respectively. Number of cycles to 2 mm displacement and failure was assessed from test (3). Maximum insertion torque was greater, and failure angle smaller, when constructs repaired with modified screws were tested (8.1 ± 0.5 vs. 7.4 ± 0.5 Nm; P = .0047 and 550 ± 104 vs. 1130 ± 230; P = .008). Axial yield (7118 ± 707 vs. 5740 ± 2267 N; P = .043) and failure load (12 347 ± 3359 vs. 8695 ± 2277 N; P = .043) were greater for specimens repaired with modified screws. No difference was detected between constructs in the number of cycles to 2 mm displacement. Condylar MC3 osteotomies repaired with modified 5.5 mm cortical screws sustained greater maximal hand torque insertion, smaller insertion failure angle and 1.4 fold greater quasi-static failure forces than constructs repaired with standard 5.5 mm screws. Use of modified screws with larger heads may improve the fixation of condylar fractures in horses. These results provide evidence to justify clinical evaluation in horses undergoing fracture repair.

Identifiants

pubmed: 35302250
doi: 10.1111/vsu.13797
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

576-591

Informations de copyright

© 2022 American College of Veterinary Surgeons.

Références

Schlich T. Introduction: the AO. In: Schlich T, ed. Surgery, Science and Industry. Science. Palgrave_Macmillan; 2012:1-5.
Nixon A. Chapter_4: general consideration in selecting cases for fracture repair. In: Nixon A, ed. Equine Fracture Repair. 2nd ed. Wiley; 2020:35-43.
Auer J. Chapter 77: principles of fracture treatment. In: Auer J, Stick J, eds. Equine surgery. 5th ed. Elsevier; 2019:1277-1314.
Stewart S, Richardson D, Boston R, Schaer TP. Risk factors associated with survival to hospital discharge of 54 horses with fractures of the radius. Vet Surg. 2015;44:1036-1041.
Peloso JG, Mundy GD, Cohen ND. Prevalence of, and factors associated with, musculoskeletal racing injuries of thoroughbreds. J Am Vet Med Assoc. 1994;204:620-626.
Parkin TD, Clegg PD, French NP, et al. Risk of fatal distal limb fractures among thoroughbreds involved in the five types of racing in the United Kingdom. Vet Rec. 2004;154:493-497.
Ramzan PH, Palmer L. Musculoskeletal injuries in thoroughbred racehorses: a study of three large training yards in Newmarket, UK (2005-2007). Vet_J. 2011;187:325-329.
Richardson D, Ortved K. Chapter 93: third metacarpal/metatarsal bones. In: Auer J, Stick J, eds. Equine Surgery. 5th ed. Elsevier; 2019:1617-1635.
Martin GS. Factors associated with racing performance of thoroughbreds undergoing lag screw repair of condylar fractures of the third metacarpal or metatarsal bone. J Am vet Med Assoc. 2000;217:1870-1877.
Shigley JE. Chapter 8: screws, fasteners, and the Design of Nonpermanent Joints. In: Budynas R, Nisbett K, eds. Shigley's Mechanical Engineering Design. 9th ed. McGraw-Hill Education; 2011:410-451.
Bishop JA, Behn AW, Castillo TN. The biomechanical significance of washer use with screw fixation. J Orthop Trauma. 2014;28:114-117.
Egol KA, Kubiak EN, Fulkerson E, et al. Biomechanics of locked plates and screws. J_Orthop_Trauma. 2004;8:18.
Implants for surgery - metal bone screws with conical under-surface of head - Dimensions, ISO 9268:1988.
Labens R, Khairuddin NH, Murray M, Jermyn K, Ahmad RS. In vitro comparison of linear vs triangular screw configuration to stabilize complete uniarticular parasagittal fractures of the proximal phalanx in horses. Vet Surg. 2019;48:96-104.
Johnson K, Smith F. Axial compression generated by cortical and cancellous lag screws in the equine distal phalanx. Vet_J. 2003;166:159-163.
Carpenter RS, Galuppo LD, Stover SM. Effect of diameter of the drill hole on torque of screw insertion and pushout strength for headless tapered compression screws in simulated fractures of the lateral condyle of the equine third metacarpal bone. Am_J_Vet_Res. 2006;67:895-900.
Galuppo LD, Stover SM, Jensen DG. A biomechanical comparison of equine third metacarpal condylar bone fragment compression and screw pushout strength between headless tapered variable pitch and AO cortical bone screws. Vet_Surg. 2002;31:201-210.
Fletcher JWA, Wenzel L, Neumann V, et al. Surgical performance when inserting non-locking screws: a systematic review. EFORT Open Rev. 2020;5:26-36.
Zoppa ALV, Santoni B, Puttlitz CM, Cochran K, Hendrickson DA. Arthrodesis of the equine proximal interphalangeal joint: a biomechanical comparison of 3-hole 4.5mm locking compression plate and 3-hole 4.5mm narrow dynamic compression plate, with two transarticular 5.5mm cortex screws. Vet Surg. 2011;40:253-259.
Haerdi C, Costa RD, Auer JA, Linke B, Steiner A. Mechanical comparison of 3 different clamp and 2 different rod types of a new veterinary internal fixation system, 4.5/5.5-mm VetFix. Vet Surg. 2003;32:431-438.
Thomas RL, Bouazza-Marouf K, Taylor GJ. Automated surgical screwdriver: automated screw placement. Proc Inst Mech Eng H. 2008;222:817-827.
Behring JK, Gjerdet NR, Mølster A. Slippage between screwdriver and bone screw. Clin Orthop Relat Res (1976-2007). 2002;404:368-372.
Brooks CR, Choudhury A. Chapter 2: mechanical aspects and macroscopic fracture-surface orientation. In: Brooks CR, Choudhury A, eds. Failure Analysis of Engineering Materials. 1st ed. McGraw-Hill Education; 2002:73-119.
Nunamaker DM, Perren SM. Force measurements in screw fixation. J_Biomech. 1976;9:669-675.
Hughes AN, Jordan BA. The mechanical properties of surgical bone screws and some aspects of insertion practice. Injury. 1972;4:25-38.
Lewis AJ, Sod GA, Burba DJ, et al. Compressive forces achieved in simulated equine third metacarpal bone lateral condylar fractures of varying fragment thickness with Acutrak plus screw and 4.5 mm AO cortical screws. Vet_Surg. 2010;39:78-82.
Moulin N, François I, Coté N, et al. Surgical repair of propagating condylar fractures of the third metacarpal/metatarsal bones with cortical screws placed in lag fashion in 26 racehorses (2007-2015). Equine_Vet_ J. 2018;50:629-635.
Turner AS, Mills EJ, Gabel AA. In vivo measurement of bone strain in the horse. Am_J_Vet_Res. 1975;36:1573-1579.
Rybicki EF, Mills EJ. In vivo and analytical studies of forces and moments in equine long bones. J_Biomech. 1977;10:701-705.
Rahm C, Ito K, Auer J. Screw fixation in lag fashion of equine cadaveric metacarpal and metatarsal condylar bone specimens: a biomechanical comparison of shaft and cortex screws. Vet_Surg. 2000;29:564-571.
Galuppo LD, Stover SM, Jensen DG, Willits NH. A biomechanical comparison of headless tapered variable pitch and AO cortical bone screws for fixation of a simulated lateral condylar fracture in equine third metacarpal bones. Vet Surg. 2001;30:332-340.
LA MD, Stover SM, Coleman K. Limb loading activity of adult horses confined to box stalls in an equine hospital barn. Am_J_Vet_Res. 2000;61:234-237.
Sod GA, Mitchell CF, Hubert JD, et al. In vitro biomechanical comparison of equine proximal interphalangeal joint arthrodesis techniques. Vet_Surg. 2007;36:792-799.
Sod GA, Riggs LM, Mitchell CF, et al. An in vitro biomechanical comparison of equine proximal interphalangeal joint arthrodesis techniques. Vet_Surg. 2010;39:83-90.
Sod GA, Riggs LM, Mitchell CF, et al. An in vitro biomechanical comparison of equine proximal interphalangeal joint arthrodesis techniques: an axial positioned dynamic compression plate and two abaxial transarticular cortical screws inserted in lag fashion versus three parallel transarticular cortical screws inserted in lag fashion. Vet Surg. 2010;39:83-90.
Gudehus T, Sod GA, Riggs LM, et al. An in vitro biomechanical comparison of equine proximal interphalangeal joint arthrodesis techniques: two parallel transarticular headless tapered variable pitch screws versus two parallel transarticular AO cortical bone screws inserted in lag fashion. Vet_Surg. 2011;40:261-265.
Merkens HW, Schamhardt HC. Distribution of ground reaction forces of the concurrently loaded limbs of the Dutch Warmblood horse at the normal walk. Equine Vet_J. 1988;20:209-213.

Auteurs

Caroline Constant (C)

AO Research Institute Davos, Davos, Switzerland.

Ivan Zderic (I)

AO Research Institute Davos, Davos, Switzerland.

Daniel Arens (D)

AO Research Institute Davos, Davos, Switzerland.

Brenna Pugliese (B)

AO Research Institute Davos, Davos, Switzerland.

Dominic Gehweiler (D)

AO Research Institute Davos, Davos, Switzerland.

Boyko Gueorguiev-Rüegg (B)

AO Research Institute Davos, Davos, Switzerland.

Stephan Zeiter (S)

AO Research Institute Davos, Davos, Switzerland.

Articles similaires

Robotic Surgical Procedures Animals Humans Telemedicine Models, Animal

Odour generalisation and detection dog training.

Lyn Caldicott, Thomas W Pike, Helen E Zulch et al.
1.00
Animals Odorants Dogs Generalization, Psychological Smell
Animals TOR Serine-Threonine Kinases Colorectal Neoplasms Colitis Mice
Animals Tail Swine Behavior, Animal Animal Husbandry

Classifications MeSH