What is the failure rate of constrained liners in complex revision total hip arthroplasty?

Acetabular bone loss Constrained acetabular liner Dual mobility cup ENDO classification Femoral bone loss Hip dislocation Paprosky classification Periprosthetic infection inter observer reliability Posterior approach Re-revision THA Revision hip arthroplasty

Journal

Archives of orthopaedic and trauma surgery
ISSN: 1434-3916
Titre abrégé: Arch Orthop Trauma Surg
Pays: Germany
ID NLM: 9011043

Informations de publication

Date de publication:
Mar 2023
Historique:
received: 22 10 2021
accepted: 06 03 2022
pubmed: 5 4 2022
medline: 3 3 2023
entrez: 4 4 2022
Statut: ppublish

Résumé

Recurrent hip dislocation after multiple revision total hip arthroplasty is a severe complication. Therefore, constrained acetabular liners (CL) have been used during salvage procedures. We report our experience of constrained liners in a re-revision setting with focus on re-dislocation. We also evaluated acetabular and femoral bone loss as potential risk factor. Between January 2013 and December 2016, 65 patients were treated in a single institution for revision and re-revision hip arthroplasty using CL. The indication for using a CL was a high risk of re-redislocation after multiple recurrent hip dislocation including failed Dual Mobility Cups (DMC). Compromising soft tissue defects as well as severe bone defect were therefore regarded as high risks. Thirty-eight patients (77.6%) underwent a minimum of three surgical procedures before the index revision procedure. Sixteen patients (24.6%) were excluded as they were lost to follow-up, expired before minimum follow-up or refused study participation, leaving 49 patients in the analysis (75.4%). The mean follow-up was 62 months (44-74; SD = 7.7). We assessed the following potential risk factors for revision or dislocation: type of surgical setting (septic/aseptic), BMI, cup inclination angle, size of liner used and acetabular and femoral bone loss according to Paprosky classification. The primary endpoints were dislocation or repeat revision for any reason. Of the 49 patients, we found an overall re-revision rate of 40.8% (20/49) and a dislocation rate of 30.6% (15/49). There were no significant differences among the surgical re-revision rate or dislocation rate as a factor of patient characteristics. In terms of bone loss, there was a trend towards higher revision rates for increasing acetabular and femoral bone loss, but without statistical significance. We found the use of a constrained liner in a re-revision setting still bears a high risk of re-revision and re-dislocation. Therefore we restrained from using constrained liners in favour of Dual mobility cups. In this study there was no significant higher dislocation rate in the subgroup of periprosthetic infection. Furthermore the rigid design of a constrained liner bears the known risk of structural failure of acetabular reconstruction implants. Severe acetabular or femoral bone defects seem to have an impact on the revision rate, but not on the dislocation rate with regards to the restored offset and center of the hip. Results have to be taken into context such that the study population inherently has a predisposition for poorer outcomes. Indications should be strongly filtered for patients at high risk for recurrent hip joint dislocation including failed DMCs with only limited bone loss and moderate soft tissue defects. Our modification to the existing classification with a high inter and intraobserver reliability will make future studies more comparable regarding revisions and bone stock loss. Still further research using objective and reproducible parameters is needed to better analyze data especially in the background of complex revision hip arthroplasty.

Sections du résumé

BACKGROUND BACKGROUND
Recurrent hip dislocation after multiple revision total hip arthroplasty is a severe complication. Therefore, constrained acetabular liners (CL) have been used during salvage procedures. We report our experience of constrained liners in a re-revision setting with focus on re-dislocation. We also evaluated acetabular and femoral bone loss as potential risk factor.
METHODS METHODS
Between January 2013 and December 2016, 65 patients were treated in a single institution for revision and re-revision hip arthroplasty using CL. The indication for using a CL was a high risk of re-redislocation after multiple recurrent hip dislocation including failed Dual Mobility Cups (DMC). Compromising soft tissue defects as well as severe bone defect were therefore regarded as high risks. Thirty-eight patients (77.6%) underwent a minimum of three surgical procedures before the index revision procedure. Sixteen patients (24.6%) were excluded as they were lost to follow-up, expired before minimum follow-up or refused study participation, leaving 49 patients in the analysis (75.4%). The mean follow-up was 62 months (44-74; SD = 7.7). We assessed the following potential risk factors for revision or dislocation: type of surgical setting (septic/aseptic), BMI, cup inclination angle, size of liner used and acetabular and femoral bone loss according to Paprosky classification. The primary endpoints were dislocation or repeat revision for any reason.
RESULTS RESULTS
Of the 49 patients, we found an overall re-revision rate of 40.8% (20/49) and a dislocation rate of 30.6% (15/49). There were no significant differences among the surgical re-revision rate or dislocation rate as a factor of patient characteristics. In terms of bone loss, there was a trend towards higher revision rates for increasing acetabular and femoral bone loss, but without statistical significance.
CONCLUSIONS CONCLUSIONS
We found the use of a constrained liner in a re-revision setting still bears a high risk of re-revision and re-dislocation. Therefore we restrained from using constrained liners in favour of Dual mobility cups. In this study there was no significant higher dislocation rate in the subgroup of periprosthetic infection. Furthermore the rigid design of a constrained liner bears the known risk of structural failure of acetabular reconstruction implants. Severe acetabular or femoral bone defects seem to have an impact on the revision rate, but not on the dislocation rate with regards to the restored offset and center of the hip. Results have to be taken into context such that the study population inherently has a predisposition for poorer outcomes. Indications should be strongly filtered for patients at high risk for recurrent hip joint dislocation including failed DMCs with only limited bone loss and moderate soft tissue defects. Our modification to the existing classification with a high inter and intraobserver reliability will make future studies more comparable regarding revisions and bone stock loss. Still further research using objective and reproducible parameters is needed to better analyze data especially in the background of complex revision hip arthroplasty.

Identifiants

pubmed: 35377048
doi: 10.1007/s00402-022-04419-z
pii: 10.1007/s00402-022-04419-z
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

1671-1678

Informations de copyright

© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Références

Berry DJ, von Knoch M, Schleck CD, Harmsen WS (2004) The cumulative long-term risk of dislocation after primary Charnley total hip arthroplasty. J Bone Joint Surg 86:9–14
doi: 10.2106/00004623-200401000-00003 pubmed: 14711939
Sanchez-Sotelo J, Berry DJ (2001) Epidemiology of instability after total hip replacement. Orthop Clin North Am 32:543–552
doi: 10.1016/S0030-5898(05)70225-X pubmed: 11689368
Woo RY, Morrey BF (1982) Dislocations after total hip arthroplasty. J Bone Joint Surg 64:1295–306
doi: 10.2106/00004623-198264090-00004 pubmed: 7142237
Carter AH, Sheehan EC, Mortazavi SMJ, Purtill JJ, Sharkey PF, Parvizi J (2011) Revision for recurrent instability: what are the predictors of failure? J Arthroplasty 26:46–52
doi: 10.1016/j.arth.2011.03.021 pubmed: 21550768
Alberton GM, High WA, Morrey BF (2002) Dislocation after revision total hip arthroplasty an analysis of risk factors and treatment options. J Bone Joint Surg 84:1788–92
doi: 10.2106/00004623-200210000-00008 pubmed: 12377909
Wetters NG, Murray TG, Moric M, Sporer SM, Paprosky WG, Della Valle CJ (2013) Risk factors for dislocation after revision total hip arthroplasty. Clin Orthop Relat Res 471:410–416
doi: 10.1007/s11999-012-2561-7 pubmed: 22956236
Hailer NP, Weiss RJ, Stark A, Kärrholm J (2012) Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocation: 228 patients from the Swedish hip Arthroplasty register. Acta Orthop 83:566–571
doi: 10.3109/17453674.2012.742395 pubmed: 23116439 pmcid: 3555442
Kung PL, Ries MD (2007) Effect of femoral head size and abductors on dislocation after revision THA. Clin Orthop Relat Res 465:170–174
doi: 10.1097/BLO.0b013e318159a983 pubmed: 17891038
van Sikes C, Lai LP, Schreiber M, Mont MA, Jinnah RH, Seyler TM (2008) Instability after total hip arthroplasty: treatment with large femoral heads vs constrained liners. J Arthroplast 23:59–63
doi: 10.1016/j.arth.2008.06.032
Charissoux J-L, Asloum Y, Marcheix P-S (2014) Surgical management of recurrent dislocation after total hip arthroplasty. Orthop Traumatol Surg Res 100:S25-34
doi: 10.1016/j.otsr.2013.11.008 pubmed: 24434366
Lombardi AV, Mallory TH, Kraus TJ, Vaughn BK (1991) Preliminary report on the S-ROM constraining acetabular insert: a retrospective clinical experience. Orthopedics 14:297–303
pubmed: 2020628
Kaper BP, Bernini PM (1998) Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg 80:561–5
doi: 10.2106/00004623-199804000-00013 pubmed: 9563386
Jones SA (2018) Constrained acetabular liners. J Arthroplast 33:1331–1336
doi: 10.1016/j.arth.2018.01.026
Su EP, Pellicci PM (2004) The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res 420:122–9
doi: 10.1097/00003086-200403000-00017
Williams JT, Ragland PS, Clarke S (2007) Constrained components for the unstable hip following total hip arthroplasty: a literature review. Int Orthop 31:273–277
Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG (2005) High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplast 20:103–107
doi: 10.1016/j.arth.2005.05.005
Jo S, Jimenez Almonte JH, Sierra RJ (2015) The cumulative risk of re-dislocation after revision tha performed for instability increases close to 35% at 15 years. J Arthroplast 30:1177–1182
doi: 10.1016/j.arth.2015.02.001
Berend KR, Lombardi AV, Mallory TH, Adams JB, Russell JH, Groseth KL (2005) The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplast 20:93–102
doi: 10.1016/j.arth.2005.06.001
Yun AG, Padgett D, Pellicci P, Dorr LD (2005) Constrained acetabular liners: mechanisms of failure. J Arthroplasty 20:536–541
doi: 10.1016/j.arth.2004.12.054 pubmed: 16124973
Unter Ecker N, Kocaoğlu H, Zahar A, Haasper C, Gehrke T, Citak M (2021) What is the dislocation and revision rate of dual-mobility cups used in complex revision THAs? Clin Orthop Relat Res 479:280–285
doi: 10.1097/CORR.0000000000001467 pubmed: 32898046
Guyen O, Lewallen DG, Cabanela ME (2008) Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg 90:1553–60
doi: 10.2106/JBJS.G.00317 pubmed: 18594105
Engelbrecht E, Heinert K (1987) Klassifikation und Behandlungsrichtlinien von Knochensubstanzverlusten bei Revisionsoperationen am Hüftgelenk — mittelfristige Ergebnisse. In: ENDO-Klinik (eds) Primär- und Revisions-Alloarthroplastik Hüft- und Kniegelenk. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-71832-8_20
Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. J Arthroplast 9:33–44
doi: 10.1016/0883-5403(94)90135-X
Paprosky WG (1990) Femoral defect classification: clinical application. Orthop Rev Oct 43(2):379–380
Bangdiwala SI, Shankar V (2013) The agreement chart. BMC Med Res Methodol 13:97
doi: 10.1186/1471-2288-13-97 pubmed: 23890315 pmcid: 3733724
Christian K, Wenhao Ch, Georges B (2021) Outcome and risk factors of failures associated with revision total hip arthroplasty for recurrent dislocation. Arch Orthop Trauma Surg. https://doi.org/10.1007/s00402-021-03814-2

Auteurs

Niklas Unter Ecker (N)

Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Holstenstrasse 2, 22767, Hamburg, Germany. Nuescience@web.de.

Pongsiri Piakong (P)

Institute of Orthopaedics, Lerdsin Hospital, Bangkok, Thailand.

Giorgio Delgado (G)

Department of Orthopedics, Philippine General Hospital, University of the Philippines Manila, Taft Avenue, 1000, Manila, Philippines.

Thorsten Gehrke (T)

Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Holstenstrasse 2, 22767, Hamburg, Germany.

Mustafa Citak (M)

Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Holstenstrasse 2, 22767, Hamburg, Germany.

Malte Ohlmeier (M)

Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Holstenstrasse 2, 22767, Hamburg, Germany.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs