Pyroelectric detector-based method for low uncertainty spectral irradiance and radiance responsivity calibrations in the infrared using tunable lasers.


Journal

Applied optics
ISSN: 1539-4522
Titre abrégé: Appl Opt
Pays: United States
ID NLM: 0247660

Informations de publication

Date de publication:
10 Apr 2022
Historique:
entrez: 26 4 2022
pubmed: 27 4 2022
medline: 27 4 2022
Statut: ppublish

Résumé

The standard uncertainty of detector-based radiance and irradiance responsivity calibrations in the short-wave infrared (SWIR) traditionally has been limited to around 1% or higher by the poor spatial uniformity of detectors used to transfer the scale from radiant power. Pyroelectric detectors offer a solution that avoids the spatial uniformity uncertainty but also introduces additional complications due to alternating current (AC) measurement techniques. Herein, a new, to the best of our knowledge, method for low uncertainty irradiance responsivity calibrations in the SWIR is presented. An absolute spectral irradiance responsivity scale was placed on two pyroelectric detectors (PED) at wavelengths

Identifiants

pubmed: 35471271
pii: 470955
doi: 10.1364/AO.455412
pmc: PMC9887653
mid: NIHMS1855426
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2957-2966

Subventions

Organisme : Intramural NIST DOC
ID : 9999-NIST
Pays : United States

Références

Appl Opt. 2006 Jan 20;45(3):511-8
pubmed: 16463736
Appl Opt. 2006 Nov 10;45(32):8218-37
pubmed: 17068565
Appl Opt. 2007 Jul 10;46(20):4228-36
pubmed: 17579678
Rev Sci Instrum. 2018 Sep;89(9):091301
pubmed: 30278699

Auteurs

Classifications MeSH