Comparison of two new integrated SS-OCT tomography and biometry devices.


Journal

Journal of cataract and refractive surgery
ISSN: 1873-4502
Titre abrégé: J Cataract Refract Surg
Pays: United States
ID NLM: 8604171

Informations de publication

Date de publication:
01 11 2022
Historique:
received: 04 01 2022
accepted: 16 05 2022
pubmed: 25 5 2022
medline: 3 11 2022
entrez: 24 5 2022
Statut: ppublish

Résumé

To evaluate the agreement in biometry measurements and intraocular lens (IOL) power calculations between 2 integrated swept-source optical coherence tomography (SS-OCT) tomography/biometry devices: the Eyestar 900 and the Anterion. Tel Aviv Medical Center, Tel Aviv, Israel. Retrospective comparative study. 113 eyes of 66 consecutive patients who were evaluated prior to cataract surgery using both devices on the same visit were included. Measurements of axial length, anterior and posterior keratometry, anterior chamber depth, corneal diameter (CD), central corneal thickness, and lens thickness were recorded from both devices and analyzed for agreement and correlation. IOL power calculations using Barrett Universal II formula were compared between the devices in addition to prediction error (PE) and the mean absolute error (MAE). The mean age was 71.6 ± 9.8 years. Clinically, the mean differences in measured variables were small. Measurement correlation and agreement between the devices were excellent for all variables, except for CD (Spearman r = 0.763, P < .001, and Bland-Altman 95% limits of agreement -0.21 to +0.65 mm). Good agreement was found between the devices in IOL power calculations, PE, and MAE, and the agreement was better when using the Eyestar 900 reflective keratometry measurements than the Eyestar 900 SS-OCT keratometry measurements (IOL power agreement within ±0.5 diopter (D) in 90.5% [n = 95] and 81.0% [n = 85], respectively [ P = .076], and within ±0.25 D in 64.8% [n = 68] and 44.8% [n = 47], respectively [ P = .005]). The 2 integrated SS-OCT tomography/biometry devices had good agreement in biometry measurements and IOL power calculations. Agreement was better when the Eyestar 900 reflective anterior keratometry measurements were used rather than the Eyestar 900 SS-OCT anterior keratometry measurements. Measurements of CD had reduced agreement and were not interchangeable.

Identifiants

pubmed: 35608316
doi: 10.1097/j.jcrs.0000000000000974
pii: 02158034-202211000-00010
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

1277-1284

Informations de copyright

Copyright © 2022 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.

Déclaration de conflit d'intérêts

Disclosures: A. Abulafia is a consultant for Haag-Streit AG and received lecture fees from Haag-Streit AG. All other authors have no conflicts of interest.

Références

Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg 2015;41:2224–2232
Sabatino F, Matarazzo F, Findl O, Maurino V. Comparative analysis of 2 swept-source optical coherence tomography biometers. J Cataract Refract Surg 2019;45:1124–1129
Fişuş AD, Hirnschall ND, Findl O. Comparison of 2 swept-source optical coherence tomography-based biometry devices. J Cataract Refract Surg 2021;47:87–92
Pfaeffli OA, Weber A, Hoffer KJ, Savini G, Baenninger PB, Thiel MA, Taroni L, Muller L. Agreement of IOL power calculation between IOLMaster 700 and Anterion swept source optical coherence tomography-based biometers. J Cataract Refract Surg 2022;48:535–541.
Panthier C, Rouger H, Gozlan Y, Moran S, Gatinel D. Comparative analysis of 2 biometers using swept-source optical coherence tomography technology. J Cataract Refract Surg 2022;48:26–31
Karakosta A, Vassilaki M, Plainis S, Elfadl NH, Tsilimbaris M, Moschandreas J. Choice of analytic approach for eye-specific outcomes: one eye or two? Am J Ophthalmol 2012;153:571–579
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310
Sorkin N, Rosenblatt A, Barequet IS. Predictability of biometry in patients undergoing cataract surgery. Optom Vis Sci 2016;93:1545–1551
Thibos LN, Wheeler W, Horner D. Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci 1997;74:367–375
Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schutze J, Buehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg 2016;42:76–83
McAlinden C, Wang Q, Gao R, Zhao W, Yu A, Li Y, Guo Y, Huang J. Axial length measurement failure rates with biometers using swept-source optical coherence tomography compared to partial-coherence interferometry and optical low-coherence interferometry. Am J Ophthalmol 2017;173:64–69
Tu R, Yu J, Savini G, Ye J, Ning R, Xiong J, Chen S, Huang J. Agreement between two optical biometers based on large coherence length SS-OCT and Scheimpflug imaging/partial coherence interferometry. J Refract Surg 2020;36:459–465
Salouti R, Nowroozzadeh MH, Tajbakhsh Z, Bagheri M, Ghoreyshi M, Azizzadeh M, Razeghinejad MR. Agreement of corneal diameter measurements obtained by a swept-source biometer and a Scheimpflug-based topographer. Cornea 2017;36:1373–1376
Vega Y, Gershoni A, Achiron A, Tuuminen R, Weinberger Y, Livny E, Nahum Y, Bahar I, Elbaz U. High agreement between Barrett Universal II calculations with and without utilization of optional biometry parameters. J Clin Med 2021;10:542–547
Kose B. Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of posterior corneal curvature. J Cataract Refract Surg 2022;48:185–189
Mi H, Tan N, Ang M, Htoon HM, Mehta JS. Comparison of anterior and posterior topographic analysis between 3 imaging systems. J Cataract Refract Surg 2015;41:2533–2545
Wylęgała A, Mazur R, Bolek B, Wylęgała E. Reproducibility, and repeatability of corneal topography measured by Revo NX, Galilei G6 and Casia 2 in normal eyes. PLoS One 2020;15:e0230589
Lu AQ, Poulsen A, Cui D, Seeger C, Lehman E, Scott IU, Pantanelli SM. Repeatability and comparability of keratometry measurements obtained with swept-source optical coherence and combined dual Scheimpflug-Placido disk-based tomography. J Cataract Refract Surg 2020;46:1637–1643

Auteurs

Nir Sorkin (N)

From the Department of Ophthalmology, Tel Aviv Medical Center and Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel (Sorkin, Achiron, Abumanhal, Cohen, Gutfreund, Mandelblum, Varssano, Levinger); Department of Ophthalmology, Shaare Zedek Medical Center and the Hebrew University-Hadassah Medical School, Jerusalem, Israel (Abulafia).

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH