Dose coefficients for organ dosimetry in tomosynthesis imaging of adults and pediatrics across diverse protocols.
Monte Carlo
Task Group 321
XCAT
patient specific
radiation dosimetry and risk
tomosynthesis
Journal
Medical physics
ISSN: 2473-4209
Titre abrégé: Med Phys
Pays: United States
ID NLM: 0425746
Informations de publication
Date de publication:
Aug 2022
Aug 2022
Historique:
revised:
16
04
2022
received:
25
01
2021
accepted:
31
05
2022
pubmed:
13
6
2022
medline:
19
8
2022
entrez:
12
6
2022
Statut:
ppublish
Résumé
The gold-standard method for estimation of patient-specific organ doses in digital tomosynthesis (DT) requires protocol-specific Monte Carlo (MC) simulations of radiation transport in anatomically accurate computational phantoms. Although accurate, MC simulations are computationally expensive, leading to a turnaround time in the order of core hours for simulating a single exam. This limits their clinical utility. The purpose of this study is to overcome this limitation by utilizing patient- and protocol-specific MC simulations to develop a comprehensive database of air-kerma-normalized organ dose coefficients for a virtual population of adult and pediatric patient models over an expanded set of exam protocols in DT for retrospective and prospective estimation of radiation dose in clinical tomosynthesis. A clinically representative virtual population of 14 patient models was used, with pediatric models (M and F) at ages 1, 5, 10, and 15 and adult patient models (M and F) with body mass index (BMIs) at 10th, 50th, and 90th percentiles of the US population. A graphics processing unit (GPU)-based MC simulation framework was used to simulate organ doses in the patient models, incorporating the scanner-specific configuration of a clinical DT system (VolumeRad, GE Healthcare, Waukesha, WI, USA) and an expanded set of exam protocols, including 21 distinct acquisition techniques for imaging a variety of anatomical regions (head and neck, thorax, spine, abdomen, and knee). Organ dose coefficients (h The database of organ dose coefficients (h This study developed a database of organ dose coefficients for a virtual population of 14 adult and pediatric XCAT patient models over a set of 21 exam protocols in DT. Using empirical measurements of air kerma in the clinic, these organ dose coefficients enable practical retrospective and prospective patient-specific radiation dosimetry. The computation of air-kerma-normalized patient-specific effective doses further enables the comparison of radiation burden to the patient populations between protocols and between imaging modalities (e.g., DT and projection radiography), as presented in this study.
Identifiants
pubmed: 35690885
doi: 10.1002/mp.15798
pmc: PMC9536505
mid: NIHMS1838253
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
5439-5450Subventions
Organisme : NIBIB NIH HHS
ID : P41 EB028744
Pays : United States
Organisme : NIBIB NIH HHS
ID : R01 EB001838
Pays : United States
Informations de copyright
© 2022 American Association of Physicists in Medicine.
Références
Med Phys. 2014 Apr;41(4):042101
pubmed: 24694149
Ann ICRP. 2007;37(2-4):1-332
pubmed: 18082557
Med Phys. 2009 Nov;36(11):4878-80
pubmed: 19994495
J Radiol Prot. 2021 Jun 01;41(2):
pubmed: 33571972
Med Phys. 2013 Apr;40(4):043701
pubmed: 23556927
Med Phys. 2022 Aug;49(8):5439-5450
pubmed: 35690885
Med Phys. 2015 Aug;42(8):4719-26
pubmed: 26233199
Eur J Radiol. 2009 Nov;72(2):244-51
pubmed: 19616909
Med Phys. 2014 Feb;41(2):023901
pubmed: 24506654
Med Phys. 2009 Dec;36(12):5480-7
pubmed: 20095260
JAMA Netw Open. 2020 Jul 1;3(7):e2011792
pubmed: 32721031
Med Phys. 2010 Sep;37(9):4902-15
pubmed: 20964209
Phys Med Biol. 2011 Apr 21;56(8):2347-65
pubmed: 21427484
J Med Imaging (Bellingham). 2017 Jul;4(3):031207
pubmed: 28804729
Med Phys. 2008 Dec;35(12):5821-8
pubmed: 19175138
AJR Am J Roentgenol. 2021 Mar;216(3):824-834
pubmed: 33474986
Radiographics. 2016 May-Jun;36(3):735-50
pubmed: 27163590
J Am Coll Radiol. 2016 Nov;13(11S):R67-R73
pubmed: 27814818
Eur Radiol. 2021 Sep;31(9):7022-7030
pubmed: 33624163
Med Phys. 2022 Feb;49(2):891-900
pubmed: 34902159
Med Phys. 2015 Jul;42(7):3788-800
pubmed: 26133581