Analytical precision assessment for microplastic analyses.
Infrared
Microplastics
Microspectroscopy
Precision
Quality assurance
Journal
Chemosphere
ISSN: 1879-1298
Titre abrégé: Chemosphere
Pays: England
ID NLM: 0320657
Informations de publication
Date de publication:
Oct 2022
Oct 2022
Historique:
received:
23
03
2022
revised:
03
06
2022
accepted:
07
06
2022
pubmed:
14
6
2022
medline:
28
6
2022
entrez:
13
6
2022
Statut:
ppublish
Résumé
The need for improved microplastic (MP) data accuracy has been widely reported, but MP precision issues have been investigated less thoroughly. This work demonstrates how initial and continuing assessments of a laboratory's analytical precision can be used for establishing laboratory repeatability for MP analyses. These precision estimates can be reported along with MP results to ensure their quality and compare them meaningfully to other data. Re-analyses of reference MP samples can be used to assess and compare precision between different laboratories. A multi-lab precision exercise was demonstrated using infrared (IR) standard test methods performed on reference samples consisting of low-concentration MP spikes in both clean water and wastewater matrices. Each lab repeated their IR analyses 7 times and calculated relative standard deviations (RSD) for each detected polymer type using a standardized template. All labs' MP methods yielded generally repeatable results, though RSDs were consistently higher for lower MP counts. The reported range of total MP counts per sample was 8-33 particles, and the observed RSDs were 0.1-0.6. These RSDs were the same or lower than the expected imprecision due to random (Poisson) counting error alone, suggesting that these automated methods did not contribute any additional variability, and had slightly better reproducibility than expected for independent recounts. The wastewater matrix exhibited numerous interfering particles but did not yield more variability than the clean water matrix. The low-count design is a worst case for precision but is appropriate for some real-world sample concentrations. In practice, labs could generate separate references for precision assessment at multiple MP ranges (e.g., high, medium, and low.) The RSDs obtained from this data can be used to generate QC charts, detect changes in analyst performance, compare to Poisson error to identify additional sources of imprecision, and determine target filtration and instrumental parameters for MP analyses.
Identifiants
pubmed: 35697113
pii: S0045-6535(22)01788-X
doi: 10.1016/j.chemosphere.2022.135295
pii:
doi:
Substances chimiques
Microplastics
0
Plastics
0
Waste Water
0
Water Pollutants, Chemical
0
Water
059QF0KO0R
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
135295Informations de copyright
Copyright © 2022 Elsevier Ltd. All rights reserved.