Initial Experience With a Novel Sheathless Guiding Catheter (Hyperion SheathLess) for Transradial Coronary Intervention.
sheathless guiding catheter
slender device
transradial approach
Journal
The Journal of invasive cardiology
ISSN: 1557-2501
Titre abrégé: J Invasive Cardiol
Pays: United States
ID NLM: 8917477
Informations de publication
Date de publication:
07 2022
07 2022
Historique:
pubmed:
24
6
2022
medline:
6
7
2022
entrez:
23
6
2022
Statut:
ppublish
Résumé
The Hyperion SheathLess guiding catheter (Asahi Intecc) is a novel guiding system for transradial percutaneous coronary intervention (PCI). Compared with the previous model (SheathLess Eaucath; Asahi Intecc), improvements achieved with the Hyperion SheathLess catheter are a more flexible tip and a smaller outer diameter of the catheter while maintaining the diameter of the inner lumen. The aim of this study was to report our initial experience with this guiding catheter. A total of 127 consecutive patients undergoing transradial PCI using a 6-Fr Hyperion SheathLess guiding catheter (n = 78) or a 7-Fr catheter (n = 49) were analyzed. Procedural success was 99.2%, with 1 patient (0.8%) requiring conversion to femoral access. No patients were noted to have coronary ostial dissection. Doppler ultrasound examination of the radial artery at 30-day follow-up was available in 111 of 127 patients, with 3 patients (2.7%; all 3 patients in the 7-Fr catheter group) found to have radial artery occlusion. Radial spasm occurred in 3 patients (2.4%; all 3 patients in the 7-Fr catheter group). Forearm hematoma occurred in 10 patients (7.9% total; 4 grade-1 patients [3.2%], 5 grade-2 patients [3.9%], and 1 grade-3 patient [0.8%]). The use of the Hyperion SheathLess guiding catheter for transradial PCI was feasible and associated with a low rate of procedure-related complications, including coronary ostial dissection. This sheathless guiding catheter could be a valuable option to further decrease the risk of procedure-related complications.
Sections du résumé
BACKGROUND
The Hyperion SheathLess guiding catheter (Asahi Intecc) is a novel guiding system for transradial percutaneous coronary intervention (PCI). Compared with the previous model (SheathLess Eaucath; Asahi Intecc), improvements achieved with the Hyperion SheathLess catheter are a more flexible tip and a smaller outer diameter of the catheter while maintaining the diameter of the inner lumen. The aim of this study was to report our initial experience with this guiding catheter.
METHODS
A total of 127 consecutive patients undergoing transradial PCI using a 6-Fr Hyperion SheathLess guiding catheter (n = 78) or a 7-Fr catheter (n = 49) were analyzed.
RESULTS
Procedural success was 99.2%, with 1 patient (0.8%) requiring conversion to femoral access. No patients were noted to have coronary ostial dissection. Doppler ultrasound examination of the radial artery at 30-day follow-up was available in 111 of 127 patients, with 3 patients (2.7%; all 3 patients in the 7-Fr catheter group) found to have radial artery occlusion. Radial spasm occurred in 3 patients (2.4%; all 3 patients in the 7-Fr catheter group). Forearm hematoma occurred in 10 patients (7.9% total; 4 grade-1 patients [3.2%], 5 grade-2 patients [3.9%], and 1 grade-3 patient [0.8%]).
CONCLUSION
The use of the Hyperion SheathLess guiding catheter for transradial PCI was feasible and associated with a low rate of procedure-related complications, including coronary ostial dissection. This sheathless guiding catheter could be a valuable option to further decrease the risk of procedure-related complications.