Photon-Counting Detector CT With Quantum Iterative Reconstruction: Impact on Liver Lesion Detection and Radiation Dose Reduction.


Journal

Investigative radiology
ISSN: 1536-0210
Titre abrégé: Invest Radiol
Pays: United States
ID NLM: 0045377

Informations de publication

Date de publication:
01 04 2023
Historique:
pubmed: 13 9 2022
medline: 16 3 2023
entrez: 12 9 2022
Statut: ppublish

Résumé

To assess image noise, diagnostic performance, and potential for radiation dose reduction of photon-counting detector (PCD) computed tomography (CT) with quantum iterative reconstruction (QIR) in the detection of hypoattenuating and hyperattenuating focal liver lesions compared with energy-integrating detector (EID) CT. A medium-sized anthropomorphic abdominal phantom with liver parenchyma and lesions (diameter, 5-10 mm; hypoattenuating and hyperattenuating from -30 HU to +90 HU at 120 kVp) was used. The phantom was imaged on ( a ) a third-generation dual-source EID-CT (SOMATOM Force, Siemens Healthineers) in the dual-energy mode at 100 and 150 kVp with tin filtration and ( b ) a clinical dual-source PCD-CT at 120 kVp (NAEOTOM Alpha, Siemens). Scans were repeated 10 times for each of 3 different radiation doses of 5, 2.5, and 1.25 mGy. Datasets were reconstructed as virtual monoenergetic images (VMIs) at 60 keV for both scanners and as linear-blended images (LBIs) for EID-CT. For PCD-CT, VMIs were reconstructed with different strength levels of QIR (QIR 1-4) and without QIR (QIR-off). For EID-CT, VMIs and LBIs were reconstructed using advanced modeled iterative reconstruction at a strength level of 3. Noise power spectrum was measured to compare image noise magnitude and texture. A channelized Hotelling model observer was used to assess diagnostic accuracy for lesion detection. The potential for radiation dose reduction using PCD-CT was estimated for the QIR strength level with the highest area under the curve compared with EID-CT for each radiation dose. Image noise decreased with increasing QIR level at all radiation doses. Using QIR-4, noise reduction was 41%, 45%, and 59% compared with EID-CT VMIs and 12%, 18%, and 33% compared with EID-CT LBIs at 5, 2.5, and 1.25 mGy, respectively. The peak spatial frequency shifted slightly to lower frequencies at higher QIR levels. Lesion detection accuracy increased at higher QIR levels and was higher for PCD-CT compared with EID-CT VMIs. The improvement in detection with PCD-CT was strongest at the lowest radiation dose, with an area under the receiver operating curve of 0.917 for QIR-4 versus 0.677 for EID-CT VMIs for hyperattenuating lesions, and 0.900 for QIR-4 versus 0.726 for EID-CT VMIs for hypoattenuating lesions. Compared with EID-CT LBIs, detection was higher for QIR 1-4 at 2.5 mGy and for QIR 2-4 at 1.25 mGy (eg, 0.900 for QIR-4 compared with 0.854 for EID-CT LBIs at 1.25 mGy). Radiation dose reduction potential of PCD-CT with QIR-4 was 54% at 5 mGy compared with VMIs and 39% at 2.5 mGy compared with LBIs. Compared with EID-CT, PCD-CT with QIR substantially improved focal liver lesion detection, especially at low radiation dose. This enables substantial radiation dose reduction while maintaining diagnostic accuracy.

Identifiants

pubmed: 36094810
doi: 10.1097/RLI.0000000000000925
pii: 00004424-202304000-00001
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

245-252

Informations de copyright

Copyright © 2022 Wolters Kluwer Health, Inc. All rights reserved.

Déclaration de conflit d'intérêts

Conflicts of interest and sources of funding: none declared.

Références

Si-Mohamed SA, Greffier J, Miailhes J, et al. Comparison of image quality between spectral photon-counting CT and dual-layer CT for the evaluation of lung nodules: a phantom study. Eur Radiol . 2022;32:524–532.
Mergen V, Ried E, Allmendinger T, et al. Epicardial adipose tissue attenuation and fat attenuation index: phantom study and in vivo measurements with photon-counting detector CT. AJR Am J Roentgenol . 2022;218:822–829.
Sartoretti T, Landsmann A, Nakhostin D, et al. Quantum iterative reconstruction for abdominal photon-counting detector CT improves image quality. Radiology . 2022;303:339–348.
Sartoretti T, Racine D, Mergen V, et al. Quantum iterative reconstruction for low-dose ultra-high-resolution photon-counting detector CT of the lung. Diagnostics (Basel) . 2022;12:522.
Rotzinger DC, Racine D, Becce F, et al. Performance of spectral photon-counting coronary CT angiography and comparison with energy-integrating-detector CT: objective assessment with model observer. Diagnostics (Basel) . 2021;11:2376.
Euler A, Higashigaito K, Mergen V, et al. High-pitch photon-counting detector computed tomography angiography of the aorta: intraindividual comparison to energy-integrating detector computed tomography at equal radiation dose. Invest Radiol . 2022;57:115–121.
Jungblut L, Kronenberg D, Mergen V, et al. Impact of contrast enhancement and virtual monoenergetic image energy levels on emphysema quantification: experience with photon-counting detector computed tomography. Invest Radiol . 2022;57:359–365.
Decker JA, Bette S, Lubina N, et al. Low-dose CT of the abdomen: initial experience on a novel photon-counting detector CT and comparison with energy-integrating detector CT. Eur J Radiol . 2022;148:110181.
Bartlett DJ, Koo CW, Bartholmai BJ, et al. High-resolution chest computed tomography imaging of the lungs: impact of 1024 matrix reconstruction and photon-counting detector computed tomography. Invest Radiol . 2019;54:129–137.
Jungblut L, Blüthgen C, Polacin M, et al. First performance evaluation of an artificial intelligence-based computer-aided detection system for pulmonary nodule evaluation in dual-source photon-counting detector CT at different low-dose levels. Invest Radiol . 2022;57:108–114.
Grunz J-P, Petritsch B, Luetkens KS, et al. Ultra-low-dose photon-counting CT imaging of the paranasal sinus with tin prefiltration: how low can we go? [published online ahead of print May 6, 2022]. Invest Radiol .
Thomsen FSL, Horstmeier S, Niehoff JH, et al. Effective spatial resolution of photon counting CT for imaging of trabecular structures is superior to conventional clinical CT and similar to high resolution peripheral CT. Invest Radiol . 2022;57:620–626.
Mergen V, Racine D, Jungblut L, et al. Virtual noncontrast abdominal imaging with photon-counting detector CT [published online ahead of print June 7, 2022]. Radiology .
Sartoretti T, Mergen V, Higashigaito K, et al. Virtual noncontrast imaging of the liver using photon-counting detector computed tomography: a systematic phantom and patient study. Invest Radiol . 2022;57:488–493.
Husarik DB, Gordic S, Desbiolles L, et al. Advanced virtual monoenergetic computed tomography of hyperattenuating and hypoattenuating liver lesions: ex-vivo and patient experience in various body sizes. Invest Radiol . 2015;50:695–702.
De Cecco CN, Caruso D, Schoepf UJ, et al. A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions. Eur Radiol . 2018;28:3393–3404.
Lenga L, Lange M, Arendt CT, et al. Measurement reliability and diagnostic accuracy of virtual monoenergetic dual-energy CT in patients with colorectal liver metastases. Acad Radiol . 2020;27:e168–e175.
Voss BA, Khandelwal A, Wells ML, et al. Impact of dual-energy 50-keV virtual monoenergetic images on radiologist confidence in detection of key imaging findings of small hepatocellular carcinomas using multiphase liver CT [published online ahead of print November 1, 2021]. Acta Radiol .
Greffier J, Frandon J, Hamard A, et al. Impact of iterative reconstructions on image quality and detectability of focal liver lesions in low-energy monochromatic images. Phys Med . 2020;77:36–42.
Euler A, Stieltjes B, Szucs-Farkas Z, et al. Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages. Eur Radiol . 2017;27:5252–5259.
Jensen K, Andersen HK, Smedby Ö, et al. Quantitative measurements versus receiver operating characteristics and visual grading regression in CT images reconstructed with iterative reconstruction: a phantom study. Acad Radiol . 2018;25:509–518.
Mileto A, Zamora DA, Alessio AM, et al. CT detectability of small low-contrast hypoattenuating focal lesions: iterative reconstructions versus filtered back projection. Radiology . 2018;289:443–454.
Hanson GJ, Michalak GJ, Childs R, et al. Low kV versus dual-energy virtual monoenergetic CT imaging for proven liver lesions: what are the advantages and trade-offs in conspicuity and image quality? A pilot study. Abdom Radiol (NY) . 2018;43:1404–1412.
Cester D, Eberhard M, Alkadhi H, et al. Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance. Quant Imaging Med Surg . 2022;12:726–741.
Oppenheimer J, Bressem KK, Elsholtz FHJ, et al. Can optimized model-based iterative reconstruction improve the contrast of liver lesions in CT? Acta Radiol . 1987;2022:2841851211070119.
Krauss B, Grant KL, Schmidt BT, et al. The importance of spectral separation: an assessment of dual-energy spectral separation for quantitative ability and dose efficiency. Invest Radiol . 2015;50:114–118.
Ellmann S, Kammerer F, Allmendinger T, et al. Advanced modeled iterative reconstruction (ADMIRE) facilitates radiation dose reduction in abdominal CT. Acad Radiol . 2018;25:1277–1284.
Kataria B, Nilsson Althén J, Smedby Ö, et al. Image quality and potential dose reduction using advanced modeled iterative reconstruction (ADMIRE) in abdominal CT—a review. Radiat Prot Dosimetry . 2021;195(3–4):177–187.
Ba A, Abbey CK, Baek J, et al. Inter-laboratory comparison of channelized Hotelling observer computation. Med Phys . 2018;45:3019–3030.
Miéville FA, Gudinchet F, Brunelle F, et al. Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments—a phantom approach. Phys Med . 2013;29:99–110.
Solomon JB, Christianson O, Samei E. Quantitative comparison of noise texture across CT scanners from different manufacturers. Med Phys . 2012;39:6048–6055.
Burgess AE. Statistically defined backgrounds: performance of a modified nonprewhitening observer model. J Opt Soc Am A Opt Image Sci Vis . 1994;11:1237–1242.
Eckstein M, Bartroff J, Abbey C, et al. Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks. Opt Express . 2003;11:460–475.
Racine D, Ba AH, Ott JG, et al. Objective assessment of low contrast detectability in computed tomography with channelized Hotelling observer. Phys Med . 2016;32:76–83.
Brankov JG. Evaluation of the channelized Hotelling observer with an internal-noise model in a train-test paradigm for cardiac SPECT defect detection. Phys Med Biol . 2013;58:7159–7182.
Wunderlich A, Noo F, Gallas BD, et al. Exact confidence intervals for channelized Hotelling observer performance in image quality studies. IEEE Trans Med Imaging . 2015;34:453–464.
Verdun FR, Racine D, Ott JG, et al. Image quality in CT: from physical measurements to model observers. Phys Med . 2015;31:823–843.
Viry A, Aberle C, Lima T, et al. Assessment of task-based image quality for abdominal CT protocols linked with national diagnostic reference levels. Eur Radiol . 2022;32:1227–1237.
Schindera ST, Odedra D, Raza SA, et al. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology . 2013;269:511–518.
Viry A, Aberle C, Racine D, et al. Effects of various generations of iterative CT reconstruction algorithms on low-contrast detectability as a function of the effective abdominal diameter: a quantitative task-based phantom study. Phys Med . 2018;48:111–118.
Jensen CT, Wagner-Bartak NA, Vu LN, et al. Detection of colorectal hepatic metastases is superior at standard radiation dose CT versus reduced dose CT. Radiology . 2019;290:400–409.
Baker ME, Dong F, Primak A, et al. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol . 2012;199:8–18.
Solomon J, Mileto A, Ramirez-Giraldo JC, et al. Diagnostic performance of an advanced modeled iterative reconstruction algorithm for low-contrast detectability with a third-generation dual-source multidetector CT scanner: potential for radiation dose reduction in a multireader study. Radiology . 2015;275:735–745. https://pubmed.ncbi.nlm.nih.gov/25751228/ . Accessed April 13, 2022.
Chen B, Yu L, Leng S, et al. Predicting detection performance with model observers: Fourier domain or spatial domain? Proc SPIE Int Soc Opt Eng . 2016;9783:978326.
Solomon J, Samei E. Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography. J Med Imaging (Bellingham) . 2016;3:035506.
Oostveen LJ, Boedeker K, Shin D, et al. Visibility of noise texture changes in CT images. In: Proc. SPIE 12035, Medical Imaging 2022: Image Perception, Observer Performance, and Technology Assessment, 120351B . 2022. https://doi.org/10.1117/12.2610736 .
Euler A, Solomon J, Marin D, et al. A third-generation adaptive statistical iterative reconstruction technique: phantom study of image noise, spatial resolution, lesion detectability, and dose reduction potential. AJR Am J Roentgenol . 2018;210:1301–1308.
Ghetti C, Palleri F, Serreli G, et al. Physical characterization of a new CT iterative reconstruction method operating in sinogram space. J Appl Clin Med Phys . 2013;14:4347.
Morsbach F, Desbiolles L, Raupach R, et al. Noise texture deviation: a measure for quantifying artifacts in computed tomography images with iterative reconstructions. Invest Radiol . 2017;52:87–94.

Auteurs

Damien Racine (D)

From the Institute of Radiation Physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne.

Victor Mergen (V)

Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich.

Anaïs Viry (A)

From the Institute of Radiation Physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne.

Matthias Eberhard (M)

Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich.

Fabio Becce (F)

Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.

David C Rotzinger (DC)

Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.

Hatem Alkadhi (H)

Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich.

André Euler (A)

Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH