Drug-Coated Balloons: Drugs Beyond Paclitaxel?
coronary heart disease
drug coated balloons
endovascular therapies
in-stent stenosis
neointima formation
paclitaxel
Journal
Frontiers in bioscience (Landmark edition)
ISSN: 2768-6698
Titre abrégé: Front Biosci (Landmark Ed)
Pays: Singapore
ID NLM: 101612996
Informations de publication
Date de publication:
10 10 2022
10 10 2022
Historique:
received:
13
06
2022
revised:
06
09
2022
accepted:
15
09
2022
entrez:
7
11
2022
pubmed:
8
11
2022
medline:
9
11
2022
Statut:
ppublish
Résumé
Although controversially discussed, paclitaxel is the only clinically proven drug that inhibits restenosis when released from drug-coated balloons (DCBs). Limus drugs are currently being explored as alternatives. The aim of the preclinical studies was to investigate drug candidates beyond paclitaxel considered for balloon coating. Drugs were tested with respect to dissolution in organic solvents, coating on balloons, and drug transfer to the vessel wall. Inhibition of neointimal proliferation was tested in the porcine model of coronary in-stent stenosis. Intravascular drug treatment was achieved by DCBs at the time of stent implantation. Coating had to be adjusted for each drug. Doses on the balloons ranged from 1.0 to 8.6 μg/mm2 balloon surface. Satisfactory amounts of drug ranging from 5% to 29% of initial doses were transferred into the vessel wall. Angiographic parameters such as late lumen loss (LLL) at 4 weeks did not show reduction of in-stent neointimal proliferation by treatment with arsenic trioxide (0.87 ± 0.44 mm), betamethasone dipropionate (1.00 ± 0.54 mm), bortezomib (1.74 ± 0.46 mm), green tea extract (1.24 ± 0.51 mm), fantolon, an epothilone (0.86 ± 0.61 mm), methotrexate (1.09 ± 0.72 mm), and thalidomide (1.59 ± 0.55 mm) compared to treatment with uncoated balloons (1.07 ± 0.60 mm), while coatings with paclitaxel reliably reduced in-stent stenosis (LLL = 0.36 ± 0.25 mm). Despite the proven antiproliferative and/or anti-inflammatory effect of the drugs, none of the coatings significantly reduced LLL compared to uncoated balloons and thus, based on the results presented here, none of the tested coatings may be considered a substitute for the paclitaxel-based coatings currently in clinical use.
Sections du résumé
BACKGROUND
Although controversially discussed, paclitaxel is the only clinically proven drug that inhibits restenosis when released from drug-coated balloons (DCBs). Limus drugs are currently being explored as alternatives. The aim of the preclinical studies was to investigate drug candidates beyond paclitaxel considered for balloon coating.
METHODS
Drugs were tested with respect to dissolution in organic solvents, coating on balloons, and drug transfer to the vessel wall. Inhibition of neointimal proliferation was tested in the porcine model of coronary in-stent stenosis. Intravascular drug treatment was achieved by DCBs at the time of stent implantation.
RESULTS
Coating had to be adjusted for each drug. Doses on the balloons ranged from 1.0 to 8.6 μg/mm2 balloon surface. Satisfactory amounts of drug ranging from 5% to 29% of initial doses were transferred into the vessel wall. Angiographic parameters such as late lumen loss (LLL) at 4 weeks did not show reduction of in-stent neointimal proliferation by treatment with arsenic trioxide (0.87 ± 0.44 mm), betamethasone dipropionate (1.00 ± 0.54 mm), bortezomib (1.74 ± 0.46 mm), green tea extract (1.24 ± 0.51 mm), fantolon, an epothilone (0.86 ± 0.61 mm), methotrexate (1.09 ± 0.72 mm), and thalidomide (1.59 ± 0.55 mm) compared to treatment with uncoated balloons (1.07 ± 0.60 mm), while coatings with paclitaxel reliably reduced in-stent stenosis (LLL = 0.36 ± 0.25 mm).
CONCLUSIONS
Despite the proven antiproliferative and/or anti-inflammatory effect of the drugs, none of the coatings significantly reduced LLL compared to uncoated balloons and thus, based on the results presented here, none of the tested coatings may be considered a substitute for the paclitaxel-based coatings currently in clinical use.
Identifiants
pubmed: 36336862
pii: S2768-6701(22)00642-6
doi: 10.31083/j.fbl2710283
doi:
Substances chimiques
Paclitaxel
P88XT4IS4D
Coated Materials, Biocompatible
0
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
283Informations de copyright
© 2022 The Author(s). Published by IMR Press.
Déclaration de conflit d'intérêts
US and BSche are shareholder of InnoRa GmbH. TH, SBie, ML, NB, DS, BK are employees of InnoRa.