How Do Classic (Static) RSA and Patient Motion Artifacts Affect the Assessment of Migration of a TKA Tibial Component? An In Vitro Study.


Journal

Clinical orthopaedics and related research
ISSN: 1528-1132
Titre abrégé: Clin Orthop Relat Res
Pays: United States
ID NLM: 0075674

Informations de publication

Date de publication:
01 Feb 2023
Historique:
received: 22 02 2022
accepted: 21 09 2022
pmc-release: 01 02 2024
pubmed: 19 11 2022
medline: 25 1 2023
entrez: 18 11 2022
Statut: ppublish

Résumé

Classic (static) Roentgen stereophotogrammetric analysis (RSA) is the current gold standard to assess, in vivo, the migration of total joint arthroplasty components. To prevent potential patient motion artifacts during the acquisition of paired radiostereometric images, images must be taken by simultaneously firing both X-ray tubes. However, the influence of nonsynchronized RSA paired images or patient motion artifacts on the precision of RSA and the assessment of implant migration is not well understood. We assessed (1) the effect of possible patient motion on the precision of RSA and (2) apparent differences in implant migration among axes (in-plane and out-of-plane translations and in-plane and out-of-plane rotations) of possible motion artifacts. Radiographs of two tibial knee arthroplasty components, each fixed in two bone-implant models as a customized phantom, were taken in a uniplanar measurement setup. We evaluated both model-based (implant models from reversed engineering) and marker-based (additional attached implant markers) RSA approaches. Between the simulated reference and follow-up examinations, we used one of the bone-implant models to simulate patient motion and the other to simulate no patient motion in parallel. Two defined protocols were followed for each of the bone-implant models: no-motion and simulated motion protocols. RSA image pairs were analyzed using a model-based RSA software package (MBRSA 4.1, RSA core ). Precision was calculated through repeat examinations, and migration of the two components was assessed for comparison of the components with each other. Measurements were taken along the medial-lateral and posterior-anterior axes for translations and around the cranial-caudal axis for rotations. The maximum total point motion was measured for comparison between the two components. The effect of simulated patient motion was generally small, except in the cranial-caudal axis, but the induced imprecision associated with motion was larger in model-based RSA than it was in marker-based RSA. The mean ± standard deviation values of precision in model-based RSA were 0.035 ± 0.015 mm, 0.045 ± 0.014 mm, and 0.049 ± 0.036 mm greater than those in marker-based RSA, in accordance with the simulated motion protocol in translations along the medial-lateral axis (0.018 ± 0.004 mm; p = 0.01), along the posterior-anterior axis (0.018 ± 0.007 mm; p = 0.003), and rotations around the cranial-caudal axis (0.017 ± 0.006 mm; p = 0.02). Apparent differences in implant migration were the greatest for the maximum total point motion. The maximum total point motion increased from 0.038 ± 0.007 mm for the no-motion protocol to 1.684 ± 0.038 mm (p < 0.001) for the simulated motion protocol in marker-based RSA, and from 0.101 ± 0.027 mm for the no-motion protocol to 1.973 ± 0.442 mm (p < 0.001) for the simulated motion protocol in model-based RSA, and was the worst-case scenario regarding patient motion artifacts. Patient motion exceeding 1 mm or 1° on nonsynchronized RSA images affects measurement errors regarding the detection of migration of a tibial component. In clinical RSA studies, the effect of patient motion on the assessment of implant migration should be of particular concern, even if clinical RSA systems have acceptable precision. Specially trained radiographers are crucial for correctly acquiring radiographs, especially when simultaneous radiography exposures are not electronically automated. In general, RSA requires synchronized image acquisition, and this should be the state-of-the-art. In clinical RSA studies, precision assessed by repeat examinations may not be reliable using the current standards that are widely used in radiology departments. When assessing implant migration for reliability, comparison of the maximum total point motion between the tested (simulated motion) implant and baseline (no-motion) implant, as in this study, is advocated because of the accurate detection of patient motion artifacts.

Sections du résumé

BACKGROUND BACKGROUND
Classic (static) Roentgen stereophotogrammetric analysis (RSA) is the current gold standard to assess, in vivo, the migration of total joint arthroplasty components. To prevent potential patient motion artifacts during the acquisition of paired radiostereometric images, images must be taken by simultaneously firing both X-ray tubes. However, the influence of nonsynchronized RSA paired images or patient motion artifacts on the precision of RSA and the assessment of implant migration is not well understood.
QUESTIONS/PURPOSES OBJECTIVE
We assessed (1) the effect of possible patient motion on the precision of RSA and (2) apparent differences in implant migration among axes (in-plane and out-of-plane translations and in-plane and out-of-plane rotations) of possible motion artifacts.
METHODS METHODS
Radiographs of two tibial knee arthroplasty components, each fixed in two bone-implant models as a customized phantom, were taken in a uniplanar measurement setup. We evaluated both model-based (implant models from reversed engineering) and marker-based (additional attached implant markers) RSA approaches. Between the simulated reference and follow-up examinations, we used one of the bone-implant models to simulate patient motion and the other to simulate no patient motion in parallel. Two defined protocols were followed for each of the bone-implant models: no-motion and simulated motion protocols. RSA image pairs were analyzed using a model-based RSA software package (MBRSA 4.1, RSA core ). Precision was calculated through repeat examinations, and migration of the two components was assessed for comparison of the components with each other. Measurements were taken along the medial-lateral and posterior-anterior axes for translations and around the cranial-caudal axis for rotations. The maximum total point motion was measured for comparison between the two components.
RESULTS RESULTS
The effect of simulated patient motion was generally small, except in the cranial-caudal axis, but the induced imprecision associated with motion was larger in model-based RSA than it was in marker-based RSA. The mean ± standard deviation values of precision in model-based RSA were 0.035 ± 0.015 mm, 0.045 ± 0.014 mm, and 0.049 ± 0.036 mm greater than those in marker-based RSA, in accordance with the simulated motion protocol in translations along the medial-lateral axis (0.018 ± 0.004 mm; p = 0.01), along the posterior-anterior axis (0.018 ± 0.007 mm; p = 0.003), and rotations around the cranial-caudal axis (0.017 ± 0.006 mm; p = 0.02). Apparent differences in implant migration were the greatest for the maximum total point motion. The maximum total point motion increased from 0.038 ± 0.007 mm for the no-motion protocol to 1.684 ± 0.038 mm (p < 0.001) for the simulated motion protocol in marker-based RSA, and from 0.101 ± 0.027 mm for the no-motion protocol to 1.973 ± 0.442 mm (p < 0.001) for the simulated motion protocol in model-based RSA, and was the worst-case scenario regarding patient motion artifacts.
CONCLUSION CONCLUSIONS
Patient motion exceeding 1 mm or 1° on nonsynchronized RSA images affects measurement errors regarding the detection of migration of a tibial component. In clinical RSA studies, the effect of patient motion on the assessment of implant migration should be of particular concern, even if clinical RSA systems have acceptable precision. Specially trained radiographers are crucial for correctly acquiring radiographs, especially when simultaneous radiography exposures are not electronically automated. In general, RSA requires synchronized image acquisition, and this should be the state-of-the-art.
CLINICAL RELEVANCE CONCLUSIONS
In clinical RSA studies, precision assessed by repeat examinations may not be reliable using the current standards that are widely used in radiology departments. When assessing implant migration for reliability, comparison of the maximum total point motion between the tested (simulated motion) implant and baseline (no-motion) implant, as in this study, is advocated because of the accurate detection of patient motion artifacts.

Identifiants

pubmed: 36398327
doi: 10.1097/CORR.0000000000002453
pii: 00003086-202302000-00039
pmc: PMC9831201
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

400-412

Informations de copyright

Copyright © 2022 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the Association of Bone and Joint Surgeons.

Déclaration de conflit d'intérêts

Each author certifies that there are no funding or commercial associations (consultancies, stock ownership, equity interest, patent/licensing arrangements, etc.) that might pose a conflict of interest in connection with the submitted article related to the author or any immediate family members. All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research® editors and board members are on file with the publication and can be viewed on request.

Références

Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310.
Bojan AJ, Bragdon C, Jönsson A, Ekholm C, Kärrholm J. Three-dimensional bone-implant movements in trochanteric hip fractures: precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res. 2015:33:705-711.
Bojan AJ, Jönsson A, Granhed H, Ekholm C, Kärrholm J. Trochanteric fracture-implant motion during healing - a radiostereometry (RSA) study. Injury. 2018;49:673-679.
Broberg JS, Yuan X, Teeter MG. Radiostereometric analysis using clinical radiographic views: development of a universal calibration object. J Biomech. 2018;73:238-242.
Brodén C, Sandberg O, Olivecrona H, Emery R, Sköldenberg O. Precision of CT-based micromotion analysis is comparable to radiostereometry for early migration measurements in cemented acetabular cups. Acta Orthop. 2021;92:419-423.
Cai R, Yuan X, Rorabeck C, Bourne RB, Holdsworth DW. Development of an RSA calibration system with improved accuracy and precision. J Biomech. 2008;41:907-911.
Christensen R, Petersen ET, Jurgens-Lahnstein J, et al. Assessment of knee kinematics with dynamic radiostereometry: validation of an automated model-based method of analysis using bone models. J Orthop Res . 2021;39:597-608.
Crockarell RJ, Guyton LJ. Arthroplasty of the knee. In: Canale ST, Beaty HJ, eds. Campbell's Operative Orthopaedics. 11th ed. Mosby Elsevier; 2007:246-255.
Grewal R, Rimmer MG, Freeman MA. Early migration of prostheses related to long-term survivorship. Comparison of tibial components in knee replacement. J Bone Joint Surg Br. 1992;74:239-242.
Gudnason A, Adalberth G, Nilsson KG, Hailer NP. Tibial component rotation around the transverse axis measured by radiostereometry predicts aseptic loosening better than maximal total point motion. Acta Orthop . 2017;88:282-287.
Hasan S, Marang-van de Mheen PJ, Kaptein BL, Nelissen R, Pijls BG. RSA-tested TKA implants on average have lower mean 10-year revision rates than non-RSA-tested designs. Clin Orthop Relat Res. 2020;478:1232-1241.
Hurschler C, Seehaus F, Emmerich J, Kaptein BL, Windhagen H. Accuracy of model-based RSA contour reduction in a typical clinical application. Clin Orthop Relat Res. 2008;466:1978-1986.
ISO16087:2013(E). Implants for surgery — Roentgen stereophotogrammetric analysis for the assessment of migration of orthopaedic implants. International Organization for Standardization, Switzerland; 2013.
Jürgens-Lahnstein JH, Petersen ET, Laursen M, et al. Development, construction, and validation of a thinner uniplanar calibration cage for radiostereometry. J Orthop Res. 2022;40:1645-1663.
Kaptein BL, Valstar ER, Stoel BC, Reiber HC, Nelissen RG. Clinical validation of model-based RSA for a total knee prosthesis. Clin Orthop Relat Res. 2007;464:205-209.
Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JH. A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech . 2003;36:873-882.
Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JH. A new type of model-based Roentgen stereophotogrammetric analysis for solving the occluded marker problem. J Biomech . 2005;38:2330-2334.
Kärrholm J, Borssen B, Lowenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br. 1994;76:912-917.
Kärrholm J, Gill RH, Valstar ER. The history and future of radiostereometric analysis. Clin Orthop Relat Res. 2006;448:10-21.
Kärrholm J, Herberts P, Hultmark P, Malchau H, Nivbrant B, Thanner J. Radiostereometry of hip prostheses. Review of methodology and clinical results. Clin Orthop Relat Res. 1997;344:94-110.
Keller MC, Hurschler C, Schwarze M. Experimental evaluation of precision and accuracy of RSA in the lumbar spine. Eur Spine J. 2021;30:2060-2068.
Lombardi AV Jr, Berend KR, Adams JB. Why knee replacements fail in 2013: patient, surgeon, or implant? Bone Joint J . 2014;96:101-104.
Nazari-Farsani S, Finnilä S, Moritz N, Mattila K, Alm JJ, Aro HT. Is model-based radiostereometric analysis suitable for clinical trials of a cementless tapered wedge femoral stem? Clin Orthop Relat Res. 2016;474:2246-2253.
Nazari-Farsani S, Vuopio M, Löyttyniemi E, Aro HT. Contributing factors to the initial femoral stem migration in cementless total hip arthroplasty of postmenopausal women. J Biomech . 2021;117:110262.
Nelissen RG, Pijls BG, Karrholm J, Malchau H, Nieuwenhuijse MJ, Valstar ER. RSA and registries: the quest for phased introduction of new implants. J Bone Joint Surg Am. 2011;93:62-65.
Niesen AE, Hull ML. Propagation of registration error into maximum total point motion to analyze tibial baseplate stability at six months using marker-based and model-based RSA. J Biomech . 2021;127:110651.
Nilsson KG, Kärrholm J. RSA in the assessment of aseptic loosening. J Bone Joint Surg Br. 1996;78:1-3.
Øhrn FD, Van Leeuwen J, Tsukanaka M, Röhrl SM. A 2-year RSA study of the Vanguard CR total knee system: a randomized controlled trial comparing patient-specific positioning guides with conventional technique. Acta Orthop. 2018:89:418-424.
Petersen ET, Rytter S, Koppens D, et al. Patients with knee osteoarthritis can be divided into subgroups based on tibiofemoral joint kinematics of gait - an exploratory and dynamic radiostereometric study. Osteoarthritis Cartilage . 2022;30:249-259.
Pijls BG, Plevier J, Nelissen R. RSA migration of total knee replacements. Acta Orthop. 2018;89:320-328.
Pijls BG, Valstar ER, Nouta KA, et al. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop. 2012;83:614-624.
Ryd L, Albrektsson BEJ, Carlsson L, et al. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br. 1995;77:377-383.
Seehaus F, Olender GD, Kaptein BL, Ostermeier S, Hurschler C. Markerless Roentgen stereophotogrammetric analysis for in vivo implant migration measurement using three dimensional surface models to represent bone. J Biomech . 2012;45:1540-1545.
Sundfeldt M, Carlsson LV, Johansson CB, Thomsen P, Gretzer C. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77:177-197.
Ten Brinke B, Beumer A, Koenraadt K, Eygendaal D, Kraan GA, Mathijssen N. The accuracy and precision of radiostereometric analysis in upper limb arthroplasty. Acta Orthop. 2017;88:320-325.
Valstar ER. RSA in practical steps. In: Digital RSA Development, Validation, and Clinical Application . Leiden University Press; 2001.
Valstar ER, Gill RH. Radiostereometric analysis in orthopaedic surgery: editorial comment. Clin Orthop Relat Res. 2006;448:2.
Valstar ER, Gill RH, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563-572.
Valstar ER, Nelissen RGHH, Reiber JHC, Rozing PM. The use of Roentgen stereophotogrammetry to study micromotion of orthopaedic implants. ISPRS Journal of Photogrammetry and Remote Sensing. 2002;56:376-389.
van Hamersveld KT, Marang-van de Mheen PJ, Nelissen RGHH. The effect of coronal alignment on tibial component migration following total knee arthroplasty: a cohort study with long-term radiostereometric analysis results. J Bone Joint Surg Am. 2019;101:1203-1212.
Yuan X, Ryd L. Accuracy analysis for RSA: a computer simulation study on 3D marker reconstruction. J Biomech. 2000;33:493-498.

Auteurs

Han Cao (H)

Department of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
Department of Orthopaedic Surgery, Taizhou Second People's Hospital, Taizhou, PR China.

Stefan Sesselmann (S)

Institute for Medical Technology, Ostbayerische Technische Hochschule Amberg-Weiden, Weiden, Germany.

Jing Xu (J)

Department of Orthopaedic Surgery, Sun Yat-Sen Memorial Hospital, Sun Yat-Sen University, Guangzhou, PR China.

Frank Seehaus (F)

Department of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
Department of Orthopaedic and Trauma Surgery, University Hospital Erlangen, Erlangen, Germany.

Raimund Forst (R)

Department of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH