Hybrid Approach Using the Cusp-Overlap Technique for Transcatheter Aortic Valve Replacement With a Balloon-Expandable Valve.


Journal

JACC. Cardiovascular interventions
ISSN: 1876-7605
Titre abrégé: JACC Cardiovasc Interv
Pays: United States
ID NLM: 101467004

Informations de publication

Date de publication:
12 12 2022
Historique:
received: 29 06 2022
revised: 23 09 2022
accepted: 21 10 2022
pubmed: 20 11 2022
medline: 15 12 2022
entrez: 19 11 2022
Statut: ppublish

Résumé

The cusp-overlap (CO) technique has recently been advocated and is being increasingly adopted for self-expandable transcatheter heart valve (THV) implantation. The aim of this study was to evaluate the feasibility, implantation depth, and outcomes of the CO technique for the balloon-expandable SAPIEN 3 THV. The CO technique was used in consecutive patients undergoing balloon-expandable THV implantation at one center between April 2021 and March 2022. Optimal fluoroscopic angles were determined from preprocedural computed tomography and confirmed on predeployment angiography. The THV radiolucent line was positioned 2 to 4 mm below the noncoronary cusp in the CO view, and positioning was confirmed in the 3-cusp view. Postdeployment THV implantation depth was assessed in both views. One-month outcomes were assessed using Valve Academic Research Consortium 3 criteria. Among 137 patients eligible for the CO technique, the CO view was not used because of unfavorable ergonomics in 27 patients (26.5%) and hemodynamic instability in 8 patients (7.8%). Among 102 patients, the mean age was 81.1 ± 6.6 years, the mean Society of Thoracic Surgeons score was 3.3% ± 2.2%, and 64.7% were men. The mean measured THV implantation depth was 3.0 ± 1.4 mm in the CO view and 2.5 ± 1.4 mm in the 3-cusp view. At 1-month follow-up, 1 patient (1.0%) had died, 1 (1.0%) had had a stroke, and 7 (6.8%) had undergone permanent pacemaker implantation. The CO technique is feasible and safe and may facilitate more accurate balloon-expandable THV positioning, especially when deep implantation needs to be avoided. Further studies are required to explore potential reduction in atrioventricular conduction block, pacemakers, or paravalvular regurgitation.

Sections du résumé

BACKGROUND
The cusp-overlap (CO) technique has recently been advocated and is being increasingly adopted for self-expandable transcatheter heart valve (THV) implantation.
OBJECTIVES
The aim of this study was to evaluate the feasibility, implantation depth, and outcomes of the CO technique for the balloon-expandable SAPIEN 3 THV.
METHODS
The CO technique was used in consecutive patients undergoing balloon-expandable THV implantation at one center between April 2021 and March 2022. Optimal fluoroscopic angles were determined from preprocedural computed tomography and confirmed on predeployment angiography. The THV radiolucent line was positioned 2 to 4 mm below the noncoronary cusp in the CO view, and positioning was confirmed in the 3-cusp view. Postdeployment THV implantation depth was assessed in both views. One-month outcomes were assessed using Valve Academic Research Consortium 3 criteria.
RESULTS
Among 137 patients eligible for the CO technique, the CO view was not used because of unfavorable ergonomics in 27 patients (26.5%) and hemodynamic instability in 8 patients (7.8%). Among 102 patients, the mean age was 81.1 ± 6.6 years, the mean Society of Thoracic Surgeons score was 3.3% ± 2.2%, and 64.7% were men. The mean measured THV implantation depth was 3.0 ± 1.4 mm in the CO view and 2.5 ± 1.4 mm in the 3-cusp view. At 1-month follow-up, 1 patient (1.0%) had died, 1 (1.0%) had had a stroke, and 7 (6.8%) had undergone permanent pacemaker implantation.
CONCLUSIONS
The CO technique is feasible and safe and may facilitate more accurate balloon-expandable THV positioning, especially when deep implantation needs to be avoided. Further studies are required to explore potential reduction in atrioventricular conduction block, pacemakers, or paravalvular regurgitation.

Identifiants

pubmed: 36402718
pii: S1936-8798(22)02006-4
doi: 10.1016/j.jcin.2022.10.035
pii:
doi:

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

2387-2395

Commentaires et corrections

Type : CommentIn

Informations de copyright

Copyright © 2022 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Déclaration de conflit d'intérêts

Funding Support and Author Disclosures Dr Akodad has received research funding from Medtronic, Biotronik, Villa M, and Federation Française de Cardiologie. Dr Blanke is a consultant to Edwards Lifesciences; and provides computed tomography core laboratory services for Edwards Lifesciences, Medtronic, Neovasc, Boston Scientific, and Abbott, for which no direct compensation is received. Dr Nestelberger has received research support from the Swiss National Science Foundation (grant P400PM_191037/1), the Swiss Heart Foundation (grant FF20079), the Prof Dr Max Cloëtta Foundation, Margarete und Walter Lichtenstein-Stiftung (grant 3MS1038), University Basel, and University Hospital Basel; and has received speaker honoraria and consulting honoraria from Siemens, Beckman Coulter, Bayer, Ortho Clinical Diagnostics and Orion Pharma, all outside the submitted work. Dr Chatfield is funded by the New Zealand Heart Foundation and John Ormiston Scholarship. Dr Chuang is funded by the Royal Australian College of Physicians Vincent Fairfax Research Entry Scholarship. Dr Leipsic holds institutional research core laboratory agreements with Medtronic, Edwards Lifesciences, Abbott, Boston Scientific, and Pi CARDIA. Dr Meier is supported by the Swiss National Science Foundation (grant P2LAP3_199561). Dr Tzimas is supported by Fondation Vaudoise de Cardiologie and the SICPA Foundation. Dr Sathananthan has received speaker fees from Edwards Lifesciences and NVT Medical; and is a consultant for Edwards Lifesciences, Boston Scientific, and Medtronic. Dr Wood is a consultant for and receives unrestricted grant support from Medtronic, Edwards Lifesciences, and Abbott Vascular. Dr Webb is a consultant for Edwards Lifesciences. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.

Auteurs

Mariama Akodad (M)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Philipp Blanke (P)

Department of Radiology, University of British Columbia, Vancouver, British Columbia, Canada.

Thomas Nestelberger (T)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Abdulmajeed Alosail (A)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Andrew G Chatfield (AG)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Ming-Yu A Chuang (MA)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Jonathon A Leipsic (JA)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Department of Radiology, University of British Columbia, Vancouver, British Columbia, Canada.

Georgios Tzimas (G)

Department of Radiology, University of British Columbia, Vancouver, British Columbia, Canada.

Youcef Lounes (Y)

Acute Care Surgery and Trauma, Division of General Surgery, Department of Surgery, University of British Columbia, Vancouver, British Columbia, Canada.

David Meier (D)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

Janarthanan Sathananthan (J)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

David A Wood (DA)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada.

John G Webb (JG)

Centres for Heart Valve Innovation and for Cardiovascular Innovation, St. Paul's and Vancouver General Hospitals, Vancouver, British Columbia, Canada; Division of Cardiology, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Centre for Heart Lung Innovation, University of British Columbia & St. Paul's Hospital, Vancouver, British Columbia, Canada. Electronic address: johngraydonwebb@gmail.com.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH