Neutron time-of-flight detectors (nTOF) used at Sandia's Z-Machine.
Journal
The Review of scientific instruments
ISSN: 1089-7623
Titre abrégé: Rev Sci Instrum
Pays: United States
ID NLM: 0405571
Informations de publication
Date de publication:
01 Nov 2022
01 Nov 2022
Historique:
entrez:
3
12
2022
pubmed:
4
12
2022
medline:
4
12
2022
Statut:
ppublish
Résumé
Neutron time-of-flight (nTOF) detectors have been used on Sandia National Laboratories' Z-Machine for inertial confinement fusion and magnetized liner fusion experiments to infer physics parameters including the apparent fuel-ion temperature, neutron yield, the magnetic-radius product (BR), and the liner rho-r. Single-paddle, dual-paddle, and co-axial scintillation nTOF detectors are used in axial lines-of-sight (LOS) and LOS that are 12° from the midplane. Detector fabrication, characterization, and calibration are discussed.
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM