A Framework for Analytical Validation of Inertial-Sensor-Based Knee Kinematics Using a Six-Degrees-of-Freedom Joint Simulator.
IMU
gait analysis
joint angle
joint simulator
knee kinematics
Journal
Sensors (Basel, Switzerland)
ISSN: 1424-8220
Titre abrégé: Sensors (Basel)
Pays: Switzerland
ID NLM: 101204366
Informations de publication
Date de publication:
29 Dec 2022
29 Dec 2022
Historique:
received:
24
11
2022
revised:
23
12
2022
accepted:
24
12
2022
entrez:
8
1
2023
pubmed:
9
1
2023
medline:
11
1
2023
Statut:
epublish
Résumé
The success of kinematic analysis that relies on inertial measurement units (IMUs) heavily depends on the performance of the underlying algorithms. Quantifying the level of uncertainty associated with the models and approximations implemented within these algorithms, without the complication of soft-tissue artefact, is therefore critical. To this end, this study aimed to assess the rotational errors associated with controlled movements. Here, data of six total knee arthroplasty patients from a previously published fluoroscopy study were used to simulate realistic kinematics of daily activities using IMUs mounted to a six-degrees-of-freedom joint simulator. A model-based method involving extended Kalman filtering to derive rotational kinematics from inertial measurements was tested and compared against the ground truth simulator values. The algorithm demonstrated excellent accuracy (root-mean-square error ≤0.9°, maximum absolute error ≤3.2°) in estimating three-dimensional rotational knee kinematics during level walking. Although maximum absolute errors linked to stair descent and sit-to-stand-to-sit rose to 5.2° and 10.8°, respectively, root-mean-square errors peaked at 1.9° and 7.5°. This study hereby describes an accurate framework for evaluating the suitability of the underlying kinematic models and assumptions of an IMU-based motion analysis system, facilitating the future validation of analogous tools.
Identifiants
pubmed: 36616945
pii: s23010348
doi: 10.3390/s23010348
pmc: PMC9824828
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Références
Acta Orthop Scand. 2000 Jun;71(3):262-7
pubmed: 10919297
Sensors (Basel). 2020 Jun 11;20(11):
pubmed: 32545227
Clin Orthop Relat Res. 2010 Jan;468(1):57-63
pubmed: 19844772
Knee Surg Sports Traumatol Arthrosc. 2018 Oct;26(10):2873-2874
pubmed: 29744530
Gait Posture. 2019 Jul;72:34-39
pubmed: 31136940
J Neuroeng Rehabil. 2022 Feb 20;19(1):22
pubmed: 35184727
PLoS One. 2017 Oct 9;12(10):e0185952
pubmed: 29016647
Sensors (Basel). 2017 Feb 22;17(2):
pubmed: 28241434
J R Soc Interface. 2019 Feb 28;16(151):20180678
pubmed: 30958178
Sensors (Basel). 2022 Mar 26;22(7):
pubmed: 35408159
J Biomech Eng. 1983 May;105(2):136-44
pubmed: 6865355
J Biomech. 2020 Jun 9;106:109832
pubmed: 32517995
Hum Mov Sci. 2021 Dec;80:102866
pubmed: 34509901
IEEE Trans Med Imaging. 2016 Jan;35(1):326-36
pubmed: 26316030
Sensors (Basel). 2020 Jan 26;20(3):
pubmed: 31991862
Lancet. 1986 Feb 8;1(8476):307-10
pubmed: 2868172
Gait Posture. 2020 Mar;77:132-137
pubmed: 32035296
Sensors (Basel). 2020 Jul 22;20(15):
pubmed: 32707987
Sensors (Basel). 2014 Apr 16;14(4):6891-909
pubmed: 24743160
J Bone Joint Surg Br. 2007 Jul;89(7):893-900
pubmed: 17673581
Gait Posture. 2010 Jan;31(1):1-8
pubmed: 19853455
Gait Posture. 2009 Apr;29(3):360-9
pubmed: 19013070
EFORT Open Rev. 2017 Mar 13;1(10):368-374
pubmed: 28461915
NPJ Digit Med. 2020 Apr 14;3:55
pubmed: 32337371
J Orthop Res. 2005 Jul;23(4):726-34
pubmed: 16022983
Orthopade. 2013 Oct;42(10):858-65
pubmed: 23695195
Clin Orthop Relat Res. 2006 Nov;452:35-43
pubmed: 16967035
J Biomech. 2010 May 7;43(7):1292-301
pubmed: 20206357
J Biomech. 2009 Nov 13;42(15):2486-94
pubmed: 19682694
Gait Posture. 2012 Jul;36(3):482-6
pubmed: 22672896
Sensors (Basel). 2020 Mar 18;20(6):
pubmed: 32197330