Evaluation of Accuracy and Performance of a Novel, Fully Gantry Integrated 3D Laser System for Computed Tomography Guided Needle Placement: A Phantom Study.

computed tomography intervention laser-guidance phantom radiation exposure safety

Journal

Diagnostics (Basel, Switzerland)
ISSN: 2075-4418
Titre abrégé: Diagnostics (Basel)
Pays: Switzerland
ID NLM: 101658402

Informations de publication

Date de publication:
12 Jan 2023
Historique:
received: 18 11 2022
revised: 21 12 2022
accepted: 10 01 2023
entrez: 21 1 2023
pubmed: 22 1 2023
medline: 22 1 2023
Statut: epublish

Résumé

The purpose of this phantom study was to compare the accuracy, speed and technical performance of CT guided needle placement using a conventional technique versus a novel, gantry integrated laser guidance system for both an expert and a novice. A total of 80 needle placements were performed in an abdominal phantom using conventional CT guidance and a laser guidance system. Analysis of pooled results of expert and novice showed a significant reduction of time (277 vs. 204 s, p = 0.001) and of the number of needle corrections (3.28 vs. 1.58, p < 0.001) required when using laser guidance versus conventional technique. No significant improvement in absolute (3.81 vs. 3.41 mm, p = 0.213) or angular deviation (2.85 vs. 2.18°, p = 0.079) was found. With either approach, the expert was significantly faster (conventional guidance: 207 s vs. 346 s, p < 0.001; laser guidance: 144 s vs. 264 s, p < 0.001) and required fewer needle corrections (conventional guidance: 4 vs. 3, p = 0.027; laser guidance: 2 vs. 1, p = 0.001) than the novice. The laser guidance system helped both the expert and the novice to perform CT guided interventions in a phantom faster and with fewer needle corrections compared to the conventional technique, while achieving similar accuracy.

Identifiants

pubmed: 36673092
pii: diagnostics13020282
doi: 10.3390/diagnostics13020282
pmc: PMC9858339
pii:
doi:

Types de publication

Journal Article

Langues

eng

Références

Invest Radiol. 2017 Jun;52(6):374-380
pubmed: 28141614
Cardiovasc Intervent Radiol. 2022 Jan;45(1):62-68
pubmed: 34414495
Cardiovasc Intervent Radiol. 2021 Sep;44(9):1323-1353
pubmed: 34231007
Sci Rep. 2020 Jan 31;10(1):1618
pubmed: 32005871
Cardiovasc Intervent Radiol. 2021 Jun;44(6):968-975
pubmed: 33474604
Cardiovasc Intervent Radiol. 2015 Oct;38(5):1252-60
pubmed: 25376924
J Am Coll Radiol. 2015 Jan;12(1):82-9
pubmed: 25086959
Cardiovasc Intervent Radiol. 2018 Apr;41(4):639-644
pubmed: 29159685
Semin Intervent Radiol. 2019 Mar;36(1):26-28
pubmed: 30936613
Cardiovasc Intervent Radiol. 2014 Feb;37(1):241-6
pubmed: 23778886
PLoS One. 2021 Nov 3;16(11):e0258896
pubmed: 34731178
Eur J Radiol. 2015 Oct;84(10):1970-5
pubmed: 26184039
Sci Rep. 2020 Nov 2;10(1):18835
pubmed: 33139835
Abdom Radiol (NY). 2021 Oct;46(10):5007-5016
pubmed: 34146132
J Vasc Interv Radiol. 2022 Nov;33(11):1416-1423.e4
pubmed: 35970505
Eur Radiol. 2020 Mar;30(3):1342-1349
pubmed: 31773299
Cardiovasc Intervent Radiol. 2021 Mar;44(3):500-506
pubmed: 33409545
Hepatology. 2019 Sep;70(3):840-850
pubmed: 30520063
Eur Radiol. 2012 Apr;22(4):930-7
pubmed: 22071776

Auteurs

Katharina Weigel (K)

Department of Radiology, Baden Cantonal Hospital, 5404 Baden, Switzerland.
Barking, Havering and Redbridge University Hospitals NHS Trust, Romford RM7 0A, UK.

Rahel A Kubik-Huch (RA)

Department of Radiology, Baden Cantonal Hospital, 5404 Baden, Switzerland.

Matthias Baer-Beck (M)

Siemens Healthineers, 91301 Forchheim, Germany.

Christian Canstein (C)

Siemens Healthineers, 91301 Forchheim, Germany.

Michael Kostrzewa (M)

Department of Radiology, Baden Cantonal Hospital, 5404 Baden, Switzerland.
Institute of Clinical Radiology and Nuclear Medicine, University Hospital Mannheim, Heidelberg University, 68167 Mannheim, Germany.

Classifications MeSH