In Vitro Investigation of the Fixation Performance of a Bioabsorbable Magnesium ACL Interference Screw Compared to a Conventional Interference Screw.
ACL reconstruction
bioabsorbable
biomechanics
interference screw
magnesium
Journal
Life (Basel, Switzerland)
ISSN: 2075-1729
Titre abrégé: Life (Basel)
Pays: Switzerland
ID NLM: 101580444
Informations de publication
Date de publication:
10 Feb 2023
10 Feb 2023
Historique:
received:
29
11
2022
revised:
31
01
2023
accepted:
08
02
2023
entrez:
25
2
2023
pubmed:
26
2
2023
medline:
26
2
2023
Statut:
epublish
Résumé
An anterior cruciate ligament (ACL) reconstruction is a common treatment for patients with ACL rupture that aims to regain pre-injury knee stability and kinematics. During the ACL reconstruction, one method to fix the graft is the use of an interference screw (IS). The IS should provide initial stability and secure the graft during the healing period. In recent years, magnesium has emerged as an alternative material to permanent metal and polymer ISs. In addition, differences in designs, such as the shape of the IS, can influence the fixation performance of the IS. Therefore, in this biomechanical experiment, two different screw designs with two ligament materials were compared in an insertion and a pull-out test at a rate of 1 mm/s. The screw designs were a conventional polymer screw and a magnesium screw. Porcine tendon and nylon rope were used as ligament materials. All tests were performed in polyurethane foam blocks with 15 PCF density (Synbone AG, Switzerland). As a result, both screw designs required an insertion torque of less than 3 Nm. There was a significant difference between the porcine and nylon rope in pull-out tests for each screw design. The magnesium screw had the highest pull-out force at 412.14 ± 50.00 N for porcine tendon and 707.38 ± 21.81 N for nylon rope. There were no significant differences in tunnel widening (narrow-wide ratio) between each ligament material. The magnesium screw showed the lowest narrow-wide tunnel ratio, implying a better ability to compress the graft to the tunnel. In conclusion, a more optimized magnesium IS design resulted in better graft fixation and an improved ACL reconstruction outcome.
Identifiants
pubmed: 36836841
pii: life13020484
doi: 10.3390/life13020484
pmc: PMC9959181
pii:
doi:
Types de publication
Journal Article
Langues
eng
Références
Int Orthop. 2014 Dec;38(12):2499-503
pubmed: 25123129
Radiol Med. 2001 Oct;102(4):211-6
pubmed: 11740446
BMC Musculoskelet Disord. 2010 Jun 30;11:139
pubmed: 20591177
Knee Surg Sports Traumatol Arthrosc. 2004 Sep;12(5):486-9
pubmed: 14767640
Arthroscopy. 2010 Aug;26(8):1105-10
pubmed: 20678709
BMC Musculoskelet Disord. 2021 Jun 18;22(1):556
pubmed: 34144682
J Orthop Translat. 2019 Oct 14;20:25-30
pubmed: 31908930
Bioact Mater. 2021 Mar 12;6(10):3231-3243
pubmed: 33778201
Arch Orthop Trauma Surg. 2016 May;136(5):593-603
pubmed: 26897352
Am J Sports Med. 1999 Sep-Oct;27(5):621-5
pubmed: 10496580
Acta Biomater. 2010 May;6(5):1714-25
pubmed: 19788945
Arch Orthop Trauma Surg. 2016 Oct;136(10):1465-71
pubmed: 27475640
Clin Orthop Relat Res. 2015 Oct;473(10):3143-53
pubmed: 25940337
J Bone Joint Surg Br. 2005 May;87(5):593-603
pubmed: 15855357
J Med Syst. 2011 Oct;35(5):941-7
pubmed: 20703682
Arthroscopy. 2010 May;26(5):705-13
pubmed: 20434671
Arch Bone Jt Surg. 2015 Oct;3(4):220-40
pubmed: 26550585
Am J Sports Med. 1994 May-Jun;22(3):334-8
pubmed: 8037273
Am J Sports Med. 1998 Jan-Feb;26(1):119-26
pubmed: 9474412
Clin Orthop Relat Res. 1975 Jan-Feb;(106):216-31
pubmed: 1126079
J Orthop Translat. 2016 Oct 01;8:32-39
pubmed: 30035092
Knee Surg Sports Traumatol Arthrosc. 2011 Nov;19(11):1798-805
pubmed: 21932078
Knee Surg Sports Traumatol Arthrosc. 1998;6 Suppl 1:S70-6
pubmed: 9608467
Arthrosc Sports Med Rehabil. 2020 Apr 06;2(2):e137-e144
pubmed: 32368750
Knee. 2003 Mar;10(1):97-102
pubmed: 12649035
J Biomater Appl. 2014 Feb 12;29(2):291-302
pubmed: 24522242