Ultrafast measurement of laser-induced shock waves.
Fiber optic hydrophone
Laser-induced shockwave
Shockwave rise time
Viscosity
Journal
Photoacoustics
ISSN: 2213-5979
Titre abrégé: Photoacoustics
Pays: Germany
ID NLM: 101622604
Informations de publication
Date de publication:
Apr 2023
Apr 2023
Historique:
received:
21
09
2022
revised:
01
02
2023
accepted:
21
02
2023
entrez:
6
3
2023
pubmed:
7
3
2023
medline:
7
3
2023
Statut:
epublish
Résumé
We present measurements of laser-induced shockwave pressure rise time in liquids on a sub-nanosecond scale, using custom-designed single-mode fiber optic hydrophone. The measurements are aimed at the study of the shockwave generation process, helping to improve the effectiveness of various applications and decrease possible accidental damage from shockwaves. The developed method allows measurement of the fast shockwave rise time as close as 10 µm from an 8 µm sized laser-induced plasma shockwave source, significantly improving the spatial and temporal resolution of the pressure measurement over other types of hydrophones. The spatial and temporal limitations of the presented hydrophone measurements are investigated theoretically, with actual experimental results agreeing well with the predictions. To demonstrate the capabilities of the fast sensor, we were able to show that the shockwave rise time is linked to liquid viscosity exhibiting logarithmic dependency in the low viscosity regime (from 0.4 cSt to 50 cSt). Additionally, the shockwave rise time dependency on propagation distance close to the source in water was investigated, with shock wave rise times measured down to only 150 ps. It was found that at short propagation distances in water halving the shock wave peak pressure results in the rise time increase by approximately factor of 1.6. These results extend the understanding of shockwave behavior in low viscosity liquids.
Identifiants
pubmed: 36874590
doi: 10.1016/j.pacs.2023.100465
pii: S2213-5979(23)00018-6
pmc: PMC9974415
doi:
Types de publication
Journal Article
Langues
eng
Pagination
100465Informations de copyright
© 2023 The Authors.
Déclaration de conflit d'intérêts
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Références
Ultrasound Med Biol. 1995;21(7):961-7
pubmed: 7491750
Ultrasound Med Biol. 1996;22(2):151-64
pubmed: 8735525
Graefes Arch Clin Exp Ophthalmol. 2017 May;255(5):945-953
pubmed: 28101654
IEEE Trans Ultrason Ferroelectr Freq Control. 2022 Jan;69(1):369-376
pubmed: 34559647
Ultrasound Med Biol. 2019 Aug;45(8):2188-2204
pubmed: 31085030
Lasers Surg Med. 1996;19(3):249-59
pubmed: 8923421
Opt Lett. 2020 Mar 15;45(6):1547-1550
pubmed: 32164013
J Biomed Opt. 2019 Jun;24(6):1-10
pubmed: 31230425
Lasers Med Sci. 2008 Jul;23(3):217-28
pubmed: 17674122
Phys Rev Lett. 2000 Feb 7;84(6):1328-30
pubmed: 11017510
Photoacoustics. 2021 Jul 21;23:100289
pubmed: 34386348
Lasers Surg Med. 2000;27(5):451-64
pubmed: 11126439
Photoacoustics. 2021 Nov 10;25:100316
pubmed: 34926158
Nat Mater. 2020 Sep;19(9):946-958
pubmed: 32807918
Ultrason Sonochem. 2022 Nov;90:106131
pubmed: 36274417
J Biomed Opt. 2005 Nov-Dec;10(6):064029
pubmed: 16409094