Clinical Accuracy of a Glucose Oxidase-Based Blood Glucose Test-Strip Across Extremes of Oxygen Partial Pressure.
blood glucose monitor
capillary
clinical accuracy
diabetes
partial pressure of oxygen
Journal
Journal of diabetes science and technology
ISSN: 1932-2968
Titre abrégé: J Diabetes Sci Technol
Pays: United States
ID NLM: 101306166
Informations de publication
Date de publication:
06 Mar 2023
06 Mar 2023
Historique:
entrez:
7
3
2023
pubmed:
8
3
2023
medline:
8
3
2023
Statut:
aheadofprint
Résumé
Glucose oxidase (GOx)-based blood glucose monitors (BGMs) are influenced by the partial pressure of oxygen (Po Clinical accuracy data were collected as part of a BGM manufacturer's ongoing post-market surveillance program for a commercially available GOx-based BGM test-strip. The data set comprised 29 901 paired BGM-comparator readings and corresponding Po A linear regression-calculated bias range of 5.22% (+0.72% [low Po Data from this large-scale clinical study, performed on unmanipulated fingertip capillary bloods from a diverse diabetes population, indicate Po
Sections du résumé
BACKGROUND
UNASSIGNED
Glucose oxidase (GOx)-based blood glucose monitors (BGMs) are influenced by the partial pressure of oxygen (Po
METHOD
UNASSIGNED
Clinical accuracy data were collected as part of a BGM manufacturer's ongoing post-market surveillance program for a commercially available GOx-based BGM test-strip. The data set comprised 29 901 paired BGM-comparator readings and corresponding Po
RESULTS
UNASSIGNED
A linear regression-calculated bias range of 5.22% (+0.72% [low Po
CONCLUSIONS
UNASSIGNED
Data from this large-scale clinical study, performed on unmanipulated fingertip capillary bloods from a diverse diabetes population, indicate Po