PEPI Lab: a flexible compact multi-modal setup for X-ray phase-contrast and spectral imaging.
Journal
Scientific reports
ISSN: 2045-2322
Titre abrégé: Sci Rep
Pays: England
ID NLM: 101563288
Informations de publication
Date de publication:
14 Mar 2023
14 Mar 2023
Historique:
received:
14
10
2022
accepted:
21
02
2023
entrez:
15
3
2023
pubmed:
16
3
2023
medline:
16
3
2023
Statut:
epublish
Résumé
This paper presents a new flexible compact multi-modal imaging setup referred to as PEPI (Photon-counting Edge-illumination Phase-contrast imaging) Lab, which is based on the edge-illumination (EI) technique and a chromatic detector. The system enables both X-ray phase-contrast (XPCI) and spectral (XSI) imaging of samples on the centimeter scale. This work conceptually follows all the stages in its realization, from the design to the first imaging results. The setup can be operated in four different modes, i.e. photon-counting/conventional, spectral, double-mask EI, and single-mask EI, whereby the switch to any modality is fast, software controlled, and does not require any hardware modification or lengthy re-alignment procedures. The system specifications, ranging from the X-ray tube features to the mask material and aspect ratio, have been quantitatively studied and optimized through a dedicated Geant4 simulation platform, guiding the choice of the instrumentation. The realization of the imaging setup, both in terms of hardware and control software, is detailed and discussed with a focus on practical/experimental aspects. Flexibility and compactness (66 cm source-to-detector distance in EI) are ensured by dedicated motion stages, whereas spectral capabilities are enabled by the Pixirad-1/Pixie-III detector in combination with a tungsten anode X-ray source operating in the range 40-100 kVp. The stability of the system, when operated in EI, has been verified, and drifts leading to mask misalignment of less than 1 [Formula: see text]m have been measured over a period of 54 h. The first imaging results, one for each modality, demonstrate that the system fulfills its design requirements. Specifically, XSI tomographic images of an iodine-based phantom demonstrate the system's quantitativeness and sensibility to concentrations in the order of a few mg/ml. Planar XPCI images of a carpenter bee specimen, both in single and double-mask modes, demonstrate that refraction sensitivity (below 0.6 [Formula: see text]rad in double-mask mode) is comparable with other XPCI systems based on microfocus sources. Phase CT capabilities have also been tested on a dedicated plastic phantom, where the phase channel yielded a 15-fold higher signal-to-noise ratio with respect to attenuation.
Identifiants
pubmed: 36918574
doi: 10.1038/s41598-023-30316-5
pii: 10.1038/s41598-023-30316-5
pmc: PMC10014955
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
4206Informations de copyright
© 2023. The Author(s).
Références
Med Phys. 2002 Jul;29(7):1634-6
pubmed: 12148746
Opt Express. 2014 Jun 30;22(13):15514-29
pubmed: 24977810
J Synchrotron Radiat. 2018 Jul 1;25(Pt 4):1068-1077
pubmed: 29979168
Nat Mater. 2008 Feb;7(2):134-7
pubmed: 18204454
Phys Med Biol. 2019 Nov 26;64(23):235005
pubmed: 31569079
Phys Med Biol. 2013 Jan 7;58(1):R1-35
pubmed: 23220766
Phys Med. 2018 May;49:58-76
pubmed: 29866345
IEEE Trans Med Imaging. 2020 Nov;39(11):3278-3289
pubmed: 32340940
Rev Sci Instrum. 2022 May 1;93(5):053706
pubmed: 35649794
J Opt Soc Am A Opt Image Sci Vis. 2021 Jan 1;38(1):71-79
pubmed: 33362154
Opt Express. 2014 Sep 22;22(19):23480-8
pubmed: 25321817
Mol Imaging Biol. 2020 Jun;22(3):539-548
pubmed: 31250331
Opt Express. 2022 Nov 21;30(24):42995-43011
pubmed: 36523008
Opt Express. 2020 May 11;28(10):14522-14535
pubmed: 32403491
Opt Express. 2021 Jan 18;29(2):2049-2064
pubmed: 33726406
Eur Radiol. 2010 Sep;20(9):2126-34
pubmed: 20309554
Radiology. 2017 Jun;283(3):723-728
pubmed: 27918709
J Phys Condens Matter. 2021 Jul 13;33(36):
pubmed: 34167096
Phys Med Biol. 2020 Mar 06;65(5):055016
pubmed: 31995530
Phys Rev Lett. 2014 Jun 27;112(25):253903
pubmed: 25014818
IEEE Trans Radiat Plasma Med Sci. 2021 Jul;5(4):441-452
pubmed: 34485784
Sci Rep. 2018 Nov 6;8(1):16394
pubmed: 30401876
Opt Express. 2021 Feb 1;29(3):3438-3457
pubmed: 33770942
J Synchrotron Radiat. 2019 Jul 1;26(Pt 4):1343-1353
pubmed: 31274463
Opt Express. 2020 Oct 26;28(22):33390-33412
pubmed: 33115004
Opt Express. 2022 Oct 10;30(21):38695-38708
pubmed: 36258428
Sci Rep. 2017 Nov 16;7(1):15722
pubmed: 29146965
Radiology. 2018 Nov;289(2):293-312
pubmed: 30179101