Magnetic resonance perfusion imaging findings following flow diversion in patients with complex middle cerebral artery bifurcation aneurysms: a single-center analysis regarding the jailed cortical branches.
Aneurysm
Flow Diverter
MR perfusion
Stent
Journal
Journal of neurointerventional surgery
ISSN: 1759-8486
Titre abrégé: J Neurointerv Surg
Pays: England
ID NLM: 101517079
Informations de publication
Date de publication:
24 Mar 2023
24 Mar 2023
Historique:
received:
24
01
2023
accepted:
07
03
2023
entrez:
24
3
2023
pubmed:
25
3
2023
medline:
25
3
2023
Statut:
aheadofprint
Résumé
Flow diverter (FD) devices provide a safe and effective treatment option especially for wide-necked intracranial aneurysms. One of the main concerns in patients treated with FD devices is patency of arterial branches jailed by the stent. However, there are no long-term data from magnetic resonance perfusion (MRP) studies regarding jailed branches. In this study we aimed to reveal the MRP findings in patients with jailed middle cerebral artery (MCA) cortical branches during long-term follow-up after flow diversion. Patients who underwent FD stent treatment for MCA aneurysms with a resulting jailed cortical branch were included. Follow-up clinical, angiographic, and MRP examination findings were recorded. Different MRP parameters were measured in the MCA territory regarding the jailed branches. Eighteen patients treated endovascularly with flow diversion for a total of 20 MCA aneurysms were included. At angiographic follow-up (median 35 months, range 7-95 months) complete occlusion was observed in 13 (65%) aneurysms and partial occlusion was observed in 6 (30%). The mean transit time (MTT) prolongation, MTT ratio, time-to peak (TTP) prolongation, and TTP ratio were 1.34, 1.20, 1.18, and 1.06 s, respectively, when compared with the contralateral side in the MCA territory. MTT, TTP, and cerebral blood volume values of the patients showed statistically significant differences compared with the contralateral side (P<0.05). Flow diversion treatment of complex bifurcation aneurysms can be effective and safe. MRP examination may reveal perfusion changes in the territory vascularized via a jailed branch, and these changes are rarely accompanied by clinical findings.
Sections du résumé
BACKGROUND
BACKGROUND
Flow diverter (FD) devices provide a safe and effective treatment option especially for wide-necked intracranial aneurysms. One of the main concerns in patients treated with FD devices is patency of arterial branches jailed by the stent. However, there are no long-term data from magnetic resonance perfusion (MRP) studies regarding jailed branches. In this study we aimed to reveal the MRP findings in patients with jailed middle cerebral artery (MCA) cortical branches during long-term follow-up after flow diversion.
METHODS
METHODS
Patients who underwent FD stent treatment for MCA aneurysms with a resulting jailed cortical branch were included. Follow-up clinical, angiographic, and MRP examination findings were recorded. Different MRP parameters were measured in the MCA territory regarding the jailed branches.
RESULTS
RESULTS
Eighteen patients treated endovascularly with flow diversion for a total of 20 MCA aneurysms were included. At angiographic follow-up (median 35 months, range 7-95 months) complete occlusion was observed in 13 (65%) aneurysms and partial occlusion was observed in 6 (30%). The mean transit time (MTT) prolongation, MTT ratio, time-to peak (TTP) prolongation, and TTP ratio were 1.34, 1.20, 1.18, and 1.06 s, respectively, when compared with the contralateral side in the MCA territory. MTT, TTP, and cerebral blood volume values of the patients showed statistically significant differences compared with the contralateral side (P<0.05).
CONCLUSIONS
CONCLUSIONS
Flow diversion treatment of complex bifurcation aneurysms can be effective and safe. MRP examination may reveal perfusion changes in the territory vascularized via a jailed branch, and these changes are rarely accompanied by clinical findings.
Identifiants
pubmed: 36963822
pii: jnis-2023-020124
doi: 10.1136/jnis-2023-020124
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
© Author(s) (or their employer(s)) 2023. No commercial re-use. See rights and permissions. Published by BMJ.
Déclaration de conflit d'intérêts
Competing interests: CI has a consulting and proctoring agreement with Medtronic, Inc., outside of this work. NK has a consulting and proctoring agreement with MicroVention, Inc. and Medtronic, Inc., outside of this work. OK has a consulting and proctoring agreement with Acandis GmbH and Stryker, Inc., outside of this work. The other authors have no conflicts of interest to declare.