[Study of Phantom Length Correction for kV-Cone Beam CT Dose Evaluation Method Using Farmer-type Ionization Chamber and Cylindrical Phantom].

Farmer-type ionization chamber cone-beam computed tomography (CBCT) dose evaluation image-guided radiotherapy (IGRT) phantom length

Journal

Nihon Hoshasen Gijutsu Gakkai zasshi
ISSN: 1881-4883
Titre abrégé: Nihon Hoshasen Gijutsu Gakkai Zasshi
Pays: Japan
ID NLM: 7505722

Informations de publication

Date de publication:
20 Jun 2023
Historique:
medline: 23 6 2023
pubmed: 17 4 2023
entrez: 16 4 2023
Statut: ppublish

Résumé

In this study, we investigated the necessary phantom length for dose evaluation of kV-CBCT mounted on the linear accelerator using a Farmer-type ionization chamber and cylindrical PMMA phantom, which many radiotherapy facilities own. Furthermore, a phantom length correction factor was proposed to compensate for the lack of scattered radiation contributed from the craniocaudal direction due to the inadequate phantom length. The air-absorbed dose at the center of a cylindrical PMMA phantom measured by a Farmer-type ionization chamber indicated that the contribution of scattered radiation saturates when the phantom length is 300 mm or longer. The phantom length correction factor was calculated from an approximate curve showing the relationship of the charge ratio measured using phantoms of 300 mm and 150 mm in length to the X-ray beam width. The air-absorbed dose measured with the 150-mm length phantom, corrected by the phantom length correction factor, showed a 1.61% dose difference from the air-absorbed dose measured with the 300-mm length phantom. In this study, the air-absorbed dose at the center of the phantom could be estimated over a wide X-ray beam width only using a 150-mm length cylindrical PMMA phantom. The method proposed in this study could be used in any radiation therapy facility.

Identifiants

pubmed: 37062712
doi: 10.6009/jjrt.2023-1349
doi:

Substances chimiques

Polymethyl Methacrylate 9011-14-7

Types de publication

English Abstract Journal Article

Langues

jpn

Sous-ensembles de citation

IM

Pagination

544-554

Auteurs

Masataka Sakamoto (M)

Department of Radiology, Hamamatsu University Hospital.

Koji Sasaki (K)

Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences.

Hayato Tsuno (H)

Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH