Transcranial ultrasound simulation with uncertainty estimation.
Journal
JASA express letters
ISSN: 2691-1191
Titre abrégé: JASA Express Lett
Pays: United States
ID NLM: 101775177
Informations de publication
Date de publication:
01 05 2023
01 05 2023
Historique:
received:
09
12
2022
accepted:
24
04
2023
medline:
15
5
2023
pubmed:
11
5
2023
entrez:
11
5
2023
Statut:
ppublish
Résumé
Transcranial ultrasound simulations are increasingly used to predict in situ exposure parameters for ultrasound therapies in the brain. However, there can be considerable uncertainty in estimating the acoustic medium properties of the skull and brain from computed tomography (CT) images. This paper shows how the resulting uncertainty in the simulated acoustic field can be predicted in a computationally efficient way using linear uncertainty propagation. Results for a representative transcranial simulation using a focused bowl transducer at 500 kHz show good agreement with unbiased uncertainty estimates obtained using Monte Carlo.
Identifiants
pubmed: 37166991
pii: 2890225
doi: 10.1121/10.0019380
pii:
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
© 2023 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).