Résumé

Multiple techniques have been described to treat humeral diaphyseal bone tumors requiring curettage or excision. Recent studies have suggested that carbon fiber-reinforced polyetheretherketone (CFR-PEEK) intramedullary nails (IMNs) may be preferable to titanium IMNs for patients with musculoskeletal tumors due to CFR-PEEK's high tensile strength, radiolucency, a modulus of elasticity closer to native bone, and improved postoperative surveillance/radiation dosing. In this study, we describe the rate of fixation failure for both CFR-PEEK and titanium humeral IMNs when used for humeral diaphyseal bone tumors requiring curettage or excision. This was a single-institution retrospective cohort study including 81 patients (27 CFR-PEEK and 54 titanium) treated for a humeral diaphyseal bone tumor using an IMN ± methylmethacrylate between January 2017 and December 2022. Primary outcome was revision surgery due to soft tissue complications, nonunions, structural complications such as periprosthetic fracture or IMN breakage, periprosthetic infection, tumor progression, and implant failure due to rejection or fatigue. No failures were observed in either patients treated with titanium nails or patients treated with CFR-PEEK not requiring curettage. Fixation failure due to implant failure was observed in 2 cases-at 214 days and 469 days after surgery-where CFR-PEEK IMN was used for stabilization after a wide segmental resection for oncologic control with a cement spacer reconstruction. In both cases, the resection was larger than 6 cm, the remaining distal humerus was less than 5 cm, and failures occurred at the interface of the residual bone and spacer. Both patients were revised using a titanium distal posterolateral humeral plate fixed with screws and cables without any subsequent complications. One additional CFR-PEEK IMN required revision surgery after 744 days due to progression of the tumor and subsequent nonunion. One revision surgery was observed after 63 days for the titanium IMN because of nonunion and tumor progression. Humeral diaphyseal bone tumors requiring large segmental resection with small residual bone and a large cement spacer may fail via tension due to bending forces at the distal portion. In this clinical scenario, the use of larger-diameter CFR-PEEK IMNs may be indicated when available. In the interim, use of intercalary allografts instead of cement spacers, additional fixation with a titanium plate distally, or the use of a titanium nail when using a cement spacer may be considered.

Sections du résumé

BACKGROUND BACKGROUND
Multiple techniques have been described to treat humeral diaphyseal bone tumors requiring curettage or excision. Recent studies have suggested that carbon fiber-reinforced polyetheretherketone (CFR-PEEK) intramedullary nails (IMNs) may be preferable to titanium IMNs for patients with musculoskeletal tumors due to CFR-PEEK's high tensile strength, radiolucency, a modulus of elasticity closer to native bone, and improved postoperative surveillance/radiation dosing. In this study, we describe the rate of fixation failure for both CFR-PEEK and titanium humeral IMNs when used for humeral diaphyseal bone tumors requiring curettage or excision.
METHODS METHODS
This was a single-institution retrospective cohort study including 81 patients (27 CFR-PEEK and 54 titanium) treated for a humeral diaphyseal bone tumor using an IMN ± methylmethacrylate between January 2017 and December 2022. Primary outcome was revision surgery due to soft tissue complications, nonunions, structural complications such as periprosthetic fracture or IMN breakage, periprosthetic infection, tumor progression, and implant failure due to rejection or fatigue.
RESULTS RESULTS
No failures were observed in either patients treated with titanium nails or patients treated with CFR-PEEK not requiring curettage. Fixation failure due to implant failure was observed in 2 cases-at 214 days and 469 days after surgery-where CFR-PEEK IMN was used for stabilization after a wide segmental resection for oncologic control with a cement spacer reconstruction. In both cases, the resection was larger than 6 cm, the remaining distal humerus was less than 5 cm, and failures occurred at the interface of the residual bone and spacer. Both patients were revised using a titanium distal posterolateral humeral plate fixed with screws and cables without any subsequent complications. One additional CFR-PEEK IMN required revision surgery after 744 days due to progression of the tumor and subsequent nonunion. One revision surgery was observed after 63 days for the titanium IMN because of nonunion and tumor progression.
CONCLUSIONS CONCLUSIONS
Humeral diaphyseal bone tumors requiring large segmental resection with small residual bone and a large cement spacer may fail via tension due to bending forces at the distal portion. In this clinical scenario, the use of larger-diameter CFR-PEEK IMNs may be indicated when available. In the interim, use of intercalary allografts instead of cement spacers, additional fixation with a titanium plate distally, or the use of a titanium nail when using a cement spacer may be considered.

Identifiants

pubmed: 37263478
pii: S1058-2746(23)00418-4
doi: 10.1016/j.jse.2023.04.023
pii:
doi:

Substances chimiques

Carbon Fiber 0
polyetheretherketone 31694-16-3
Titanium D1JT611TNE
Polyethylene Glycols 3WJQ0SDW1A
Ketones 0
Carbon 7440-44-0

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2286-2295

Informations de copyright

Copyright © 2023 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.

Auteurs

Abhiram R Bhashyam (AR)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Caleb Yeung (C)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Alisha Sodhi (A)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Raylin F Xu (RF)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Olivier Q Groot (OQ)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Sean Kelly (S)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA.

Santiago Lozano-Calderon (S)

Department of Orthopaedics, Massachusetts General Hospital, Boston, MA, USA. Electronic address: slozanocalderon@mgh.harvard.edu.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH