Usefulness of electrocardiogram mA modulation during the electrocardiogram-gated CT scan in paediatrics with high heart rate for different helical pitch: a phantom-based assessment study.
Journal
Radiation protection dosimetry
ISSN: 1742-3406
Titre abrégé: Radiat Prot Dosimetry
Pays: England
ID NLM: 8109958
Informations de publication
Date de publication:
21 Jul 2023
21 Jul 2023
Historique:
received:
08
01
2023
revised:
08
04
2023
accepted:
29
05
2023
medline:
24
7
2023
pubmed:
20
6
2023
entrez:
20
6
2023
Statut:
ppublish
Résumé
We investigated the effect of electrocardiographic (ECG) mA-modulation of ECG-gated scans of computed tomography (CTA) on radiation dose and image noise at high heart rates (HR) above 100 bpm between helical pitches (HP) 0.16 and 0.24. ECG mA-modulation range during ECG-gated CTA is 50-100 mA, the phase setting is 40-60% and the scan range is 90 mm for clinical data during HR for 90, 120 and 150 bpm. Radiation dose and image noise in Housfield units are measured for CT equipment during HR for 90, 120 and 150 bpm between HP 0.16 and 0.24. ECG mA-modulation, dose reduction ratio for HR 90, 120 and 150 bpm are 19.1, 13.4 and 8.7% at HP 0.16 and 17.1, 13.3 and 7.7% at HP 0.24, respectively. No significant differences were observed in image noise between both HP. Dose reductions of 8-24% are achieved with ECG mA-modulation during ECG-gated CCTA scan, which is beneficial even in high HR more than 100 bpm.
Identifiants
pubmed: 37337642
pii: 7202135
doi: 10.1093/rpd/ncad180
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
1295-1300Informations de copyright
© The Author(s) 2023. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.