Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement.
Defibrillation test
Generator replacement
Impedance
PRAETORIAN score
Subcutaneous ICD
Journal
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
ISSN: 1532-2092
Titre abrégé: Europace
Pays: England
ID NLM: 100883649
Informations de publication
Date de publication:
04 07 2023
04 07 2023
Historique:
received:
05
05
2023
accepted:
22
06
2023
medline:
10
7
2023
pubmed:
28
6
2023
entrez:
28
6
2023
Statut:
ppublish
Résumé
After implantation of a subcutaneous implantable cardioverter-defibrillator (S-ICD), a defibrillation test (DFT) is performed to ensure that the device can effectively detect and terminate the induced ventricular arrhythmia. Data on DFT efficacy at generator replacement are scarce with a limited number of patients and conflicting results. This study evaluates conversion efficacy during DFT at elective S-ICD generator replacement in a large cohort from our tertiary centre. Retrospective data of patients who underwent an S-ICD generator replacement for battery depletion with subsequent DFT between February 2015 and June 2022 were collected. Defibrillation test data were collected from both implant and replacement procedures. PRAETORIAN scores at implant were calculated. Defibrillation test was defined unsuccessful when two conversions at 65 J failed. A total of 121 patients were included. The defibrillation test was successful in 95% after the first and 98% after two consecutive tests. This was comparable with success rates at implant, despite a significant rise in shock impedance (73 ± 23 vs. 83 ± 24 Ω, P < 0.001). Both patients with an unsuccessful DFT at 65 J successfully converted with 80 J. This study shows a high DFT conversion rate at elective S-ICD generator replacement, which is comparable to conversion rates at implant, despite a rise in shock impedance. Evaluating device position before generator replacement may be recommended to optimize defibrillation success at generator replacement.
Identifiants
pubmed: 37379530
pii: 7209713
doi: 10.1093/europace/euad184
pmc: PMC10325005
pii:
doi:
Types de publication
Journal Article
Comment
Langues
eng
Sous-ensembles de citation
IM
Commentaires et corrections
Type : CommentOn
Informations de copyright
© The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology.
Déclaration de conflit d'intérêts
Conflict of interest: R.E.K. reports consultancy fees and research grants from Abbott, Boston Scientific, Cardiac, and Medtronic and has stock options from AtaCor Medical Inc. A.A.M.W. is a member of the Scientific advisory board of Thryv Therapeutics and ARMGO. K.M.K. reports consultancy fees from Boston Scientific. The other authors report no conflicts of interest.
Références
Neth Heart J. 2020 Aug;28(Suppl 1):122-127
pubmed: 32780342
Am Heart J. 2019 Aug;214:167-174
pubmed: 31220775
Lancet. 2015 Feb 28;385(9970):785-91
pubmed: 25715991
Eur J Heart Fail. 2022 Dec;24(12):2238-2250
pubmed: 36394512
Europace. 2016 Jul;18(7):1010-5
pubmed: 26324840
Heart Rhythm. 2019 Mar;16(3):403-410
pubmed: 30292861
JACC Clin Electrophysiol. 2018 May;4(5):652-659
pubmed: 29798794
Europace. 2022 Jan 4;24(1):30-31
pubmed: 34339485
Europace. 2020 Dec 23;22(12):1822-1829
pubmed: 33118017
JACC Clin Electrophysiol. 2017 Apr;3(4):405-414
pubmed: 29759454
Heart Rhythm. 2020 Mar;17(3):455-459
pubmed: 31669771
N Engl J Med. 1996 Dec 26;335(26):1933-40
pubmed: 8960472
Eur Heart J. 2015 Oct 1;36(37):2500-7
pubmed: 26112885
Cardiology. 2017;136(1):29-32
pubmed: 27548370
Heart Rhythm. 2008 Mar;5(3):387-90
pubmed: 18243813
J Cardiovasc Electrophysiol. 2019 Nov;30(11):2441-2447
pubmed: 31512339
Front Cardiovasc Med. 2022 Nov 01;9:1045730
pubmed: 36386377
Europace. 2022 Feb 2;24(2):181-192
pubmed: 34453529
Europace. 2021 Mar 8;23(3):395-399
pubmed: 33197266
Eur Heart J. 2022 Dec 14;43(47):4872-4883
pubmed: 36030464
N Engl J Med. 2005 Jan 20;352(3):225-37
pubmed: 15659722
Europace. 2016 Feb;18(2):159-83
pubmed: 26585598
N Engl J Med. 2020 Aug 6;383(6):526-536
pubmed: 32757521