Impact of Median Sternotomy on Safety and Efficacy of the Subcutaneous Implantable Cardioverter Defibrillator.
arrhythmia, cardiac
defibrillator, implantable
shock
sternotomy
Journal
Circulation. Arrhythmia and electrophysiology
ISSN: 1941-3084
Titre abrégé: Circ Arrhythm Electrophysiol
Pays: United States
ID NLM: 101474365
Informations de publication
Date de publication:
08 2023
08 2023
Historique:
medline:
17
8
2023
pubmed:
24
7
2023
entrez:
24
7
2023
Statut:
ppublish
Résumé
Subcutaneous implantable cardioverter defibrillators (S-ICDs) are an attractive alternative to transvenous ICDs among those not requiring pacing. However, the risks of damage to the S-ICD electrode during sternotomy and adverse interactions with sternal wires remain unclear. We sought to determine the rates of damage to the S-ICD lead during sternotomy, inappropriate shocks from electrical noise due to interaction with sternal wires, and failure to terminate spontaneous or induced ventricular arrhythmias. Retrospective, multicenter study of patients undergoing sternotomy before or after S-ICD implantation. Clinical, procedural, and device-related data were collected by each center and analyzed by the coordinating center. These data were compared with a historical control cohort of nonsternotomy patients. Of 196 identified patients (52±16 years, 47 women), 166 underwent S-ICD implantation after sternotomy and 30 sternotomy after S-ICD. There was no damage to any lead among those who underwent sternotomy after S-ICD. Defibrillation threshold testing was performed in 63% at implant, with 91% first shock success. During a median follow-up of 29 months (range, 1-188), S-ICD first shocks successfully terminated spontaneous ventricular arrhythmias in 31 of 32 patients (97%). Inappropriate shocks occurred in 22 patients, most commonly related to T wave oversensing (n=14). Compared with the nonsternotomy controls, there were no differences in rates of first shock success for induced or spontaneous arrhythmias or rate of inappropriate shocks. Sternotomy before or after S-ICD does not confer additional risk relative to a historical control group without sternotomy.
Sections du résumé
BACKGROUND
Subcutaneous implantable cardioverter defibrillators (S-ICDs) are an attractive alternative to transvenous ICDs among those not requiring pacing. However, the risks of damage to the S-ICD electrode during sternotomy and adverse interactions with sternal wires remain unclear. We sought to determine the rates of damage to the S-ICD lead during sternotomy, inappropriate shocks from electrical noise due to interaction with sternal wires, and failure to terminate spontaneous or induced ventricular arrhythmias.
METHODS
Retrospective, multicenter study of patients undergoing sternotomy before or after S-ICD implantation. Clinical, procedural, and device-related data were collected by each center and analyzed by the coordinating center. These data were compared with a historical control cohort of nonsternotomy patients.
RESULTS
Of 196 identified patients (52±16 years, 47 women), 166 underwent S-ICD implantation after sternotomy and 30 sternotomy after S-ICD. There was no damage to any lead among those who underwent sternotomy after S-ICD. Defibrillation threshold testing was performed in 63% at implant, with 91% first shock success. During a median follow-up of 29 months (range, 1-188), S-ICD first shocks successfully terminated spontaneous ventricular arrhythmias in 31 of 32 patients (97%). Inappropriate shocks occurred in 22 patients, most commonly related to T wave oversensing (n=14). Compared with the nonsternotomy controls, there were no differences in rates of first shock success for induced or spontaneous arrhythmias or rate of inappropriate shocks.
CONCLUSIONS
Sternotomy before or after S-ICD does not confer additional risk relative to a historical control group without sternotomy.
Identifiants
pubmed: 37485688
doi: 10.1161/CIRCEP.123.011867
doi:
Types de publication
Multicenter Study
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM