Real-World Evidence of the Cambridge Hybrid Closed-Loop App With a Novel Real-Time Continuous Glucose Monitoring System.
CamAPS FX
automated insulin delivery
hybrid closed loop
real world evidence
Journal
Journal of diabetes science and technology
ISSN: 1932-2968
Titre abrégé: J Diabetes Sci Technol
Pays: United States
ID NLM: 101306166
Informations de publication
Date de publication:
28 Jul 2023
28 Jul 2023
Historique:
medline:
28
7
2023
pubmed:
28
7
2023
entrez:
28
7
2023
Statut:
aheadofprint
Résumé
We evaluated the performance of the interoperable Cambridge hybrid closed-loop app with FreeStyle Libre 3 glucose sensor, and YpsoPump insulin pump in a real-world setting. Data from 100 users (63 adults [mean ± SD age 41.9 ± 14.0 years], 15 children [8.6 ± 5.2 years)] and 22 users of unreported age) for a period of 28 days were analyzed. Time in range (3.91- 10.0mmol/L) was 72.6 ± 11.1% overall. Time below range (<3.9mmol/L) was 3.1% (1.4-5.1) (median [interquartile range]). Auto-mode was active for 95.8% (91.8-97.9) of time. This real-world analysis suggests that the performance of Cambridge hybrid closed-loop app with this glucose sensor is comparable to other commercially available hybrid closed-loop systems.