Low BAU/ml values with 4 of 5 SARS CoV-2 spike-specific monoclonal antibodies in the Roche Elecsys antibody assay.

Casirivimab Etesevimab Imdevimab Monoclonal antibodies Receptor binding domain SARS CoV-2 SARS CoV-2 antibody assay Sotrovimab

Journal

Journal of virological methods
ISSN: 1879-0984
Titre abrégé: J Virol Methods
Pays: Netherlands
ID NLM: 8005839

Informations de publication

Date de publication:
11 2023
Historique:
received: 13 04 2023
revised: 21 07 2023
accepted: 26 07 2023
medline: 21 9 2023
pubmed: 30 7 2023
entrez: 29 7 2023
Statut: ppublish

Résumé

The Abbott SARS-CoV-2 IgG Quant II assay and the Roche Elecsys double antigen sandwich (DAgS) immunoassay measure SARS-CoV-2 receptor binding domain (RBD)-specific antibodies in serum samples in different ways. The IgG Quant II assay uses an antigen in combination with a secondary antibody and the DAgS assay uses two antigens. The aim of the study was to investigate whether the assays give comparable results with monoclonal antibodies. The immunoassays were tested with the RBD-specific human monoclonal antibodies (mAbs) casirivimab. imdevimab, CR3022, etesevimab and sotrovimab. The mAbs were tested at various concentrations in µg/ml, alone or in combination and the relative light units (RLU) and binding antibody units (BAU)/ml were determined. With 1 µg/ml of casirivimab, imdevimab, CR3022 and etesevimab the Abbott IgG II Quant assay yielded between 65 and 158 BAU/ml and the Elecsys assay < 0.4 - 7.1 BAU/ml. In the DAgS assay, the addition of a second and a third mAb increased the BAU/ml values synergistically. With increasing concentrations of the mAb combinations in µg/ml the Abbott IgG Quant II assay showed proportionate and the Elecsys DAgS assay disproportionate increases in BAU/ml. With 1 µg/ml sotrovimab the Abbott assay gave 39 and the Elecsys assay 136 BAU/ml. The DAgS assay showed a high dose hook effect in the µg/ml range. The secondary antibody-based and the DAgS-based SARS CoV-2 antibody assays gave very different results with 4 of 5 mAbs. This suggests that the two assays measure different binding characteristics. The ability of antibodies to cross-link multiple antigen-antibody complexes may contribute to the measurement signal in the DAgS assay.

Sections du résumé

BACKGROUND
The Abbott SARS-CoV-2 IgG Quant II assay and the Roche Elecsys double antigen sandwich (DAgS) immunoassay measure SARS-CoV-2 receptor binding domain (RBD)-specific antibodies in serum samples in different ways. The IgG Quant II assay uses an antigen in combination with a secondary antibody and the DAgS assay uses two antigens. The aim of the study was to investigate whether the assays give comparable results with monoclonal antibodies.
MATERIAL AND METHODS
The immunoassays were tested with the RBD-specific human monoclonal antibodies (mAbs) casirivimab. imdevimab, CR3022, etesevimab and sotrovimab. The mAbs were tested at various concentrations in µg/ml, alone or in combination and the relative light units (RLU) and binding antibody units (BAU)/ml were determined.
RESULTS
With 1 µg/ml of casirivimab, imdevimab, CR3022 and etesevimab the Abbott IgG II Quant assay yielded between 65 and 158 BAU/ml and the Elecsys assay < 0.4 - 7.1 BAU/ml. In the DAgS assay, the addition of a second and a third mAb increased the BAU/ml values synergistically. With increasing concentrations of the mAb combinations in µg/ml the Abbott IgG Quant II assay showed proportionate and the Elecsys DAgS assay disproportionate increases in BAU/ml. With 1 µg/ml sotrovimab the Abbott assay gave 39 and the Elecsys assay 136 BAU/ml. The DAgS assay showed a high dose hook effect in the µg/ml range.
CONCLUSIONS
The secondary antibody-based and the DAgS-based SARS CoV-2 antibody assays gave very different results with 4 of 5 mAbs. This suggests that the two assays measure different binding characteristics. The ability of antibodies to cross-link multiple antigen-antibody complexes may contribute to the measurement signal in the DAgS assay.

Identifiants

pubmed: 37516369
pii: S0166-0934(23)00111-8
doi: 10.1016/j.jviromet.2023.114786
pii:
doi:

Substances chimiques

Antibodies, Viral 0
Antibodies, Monoclonal 0
Immunoglobulin G 0

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

114786

Informations de copyright

Copyright © 2023 Elsevier B.V. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Christian Jassoy (C)

Institute for Medical Microbiology and Virology, University Hospital and Medical Faculty, University of Leipzig, Leipzig, Germany. Electronic address: Christian.jassoy@medizin.uni-leipzig.de.

Ronald Biemann (R)

Institute for Laboratory Medicine, Clinical Chemistry and Molecular Diagnostics, University Hospital and Medical Faculty, University of Leipzig, Leipzig, Germany.

Mario Hönemann (M)

Institute for Medical Microbiology and Virology, University Hospital and Medical Faculty, University of Leipzig, Leipzig, Germany.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH