Evaluation of the fully automated LIAISON®XL chemiluminescence analyzer for QuantiFERON®-CMV testing in transplant recipients.


Journal

Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology
ISSN: 1873-5967
Titre abrégé: J Clin Virol
Pays: Netherlands
ID NLM: 9815671

Informations de publication

Date de publication:
09 2023
Historique:
received: 06 04 2023
revised: 17 07 2023
accepted: 19 07 2023
medline: 16 8 2023
pubmed: 1 8 2023
entrez: 1 8 2023
Statut: ppublish

Résumé

Monitoring CMV-specific cell-mediated immunity by the QuantiFERON®-CMV (QF-CMV) may be useful in predicting the risk of CMV infection in transplant recipients (TR). As the QuantiFERON-Tuberculosis (QFT®-Plus) became available on the fully automated LIAISON®XL chemiluminescence (CLIA) analyzer, we evaluated the performance of the QF-CMV on the LIAISON®XL analyzer using the QuantiFERON®-TB Gold Plus reagent. Between 2018 and 2022, 81 samples from TR were collected at the Department of Virology of Limoges Hospital, France. Whole blood was collected into each of the three QF-CMV collection tubes: a CMV-antigen tube (QF-Ag), a mitogen tube (QF-Mg) (positive control), and a nil tube (negative control). The QF-CMV was performed on the LIAISON®XL analyzer, and results were compared with those obtained by conventional microplate ELISA. Intra- and inter-assay coefficients of variation were inferior to 20%. No inter-sample contamination was found (p=0.366). The level of concordance between CLIA and the commonly used ELISA method was 88.9%. Positive and negative percent agreements were 92.3% and 85.7%, respectively, with a very good agreement between assays (κ=0.818). Most discordances were due to indeterminate- or negative-ELISA/positive-CLIA results (most of ELISA results were borderline). Linear regression analyses demonstrated a strong correlation between both assays (QF-Ag Pearson's r=0.978, QF-Mg Pearson's r=0.963). No significant difference was observed in median QF-CMV values between both assays (QF-Ag p=0.776; QF-Mg p=0.853; Mann-Whitney U test). The Bland-Altman plots showed a minor difference in IFN-γ release (QF-Ag -0.069 IU/ml, 95% limits of agreement (LoA): -1.589; 1.451; QF-Mg 0.190 IU/ml, 95% LoA: -2.070; 2.450). Automated QF-CMV with CLIA is comparable to QF-CMV performed by ELISA with a presumably higher sensitivity for IFN-γ detection that may result in the conversion of samples close to the ELISA cut-off into positive results. Moreover, the use of a random-access analyzer allows to optimize the follow-up of TR.

Sections du résumé

BACKGROUND
Monitoring CMV-specific cell-mediated immunity by the QuantiFERON®-CMV (QF-CMV) may be useful in predicting the risk of CMV infection in transplant recipients (TR).
OBJECTIVES
As the QuantiFERON-Tuberculosis (QFT®-Plus) became available on the fully automated LIAISON®XL chemiluminescence (CLIA) analyzer, we evaluated the performance of the QF-CMV on the LIAISON®XL analyzer using the QuantiFERON®-TB Gold Plus reagent.
STUDY DESIGN
Between 2018 and 2022, 81 samples from TR were collected at the Department of Virology of Limoges Hospital, France. Whole blood was collected into each of the three QF-CMV collection tubes: a CMV-antigen tube (QF-Ag), a mitogen tube (QF-Mg) (positive control), and a nil tube (negative control). The QF-CMV was performed on the LIAISON®XL analyzer, and results were compared with those obtained by conventional microplate ELISA.
RESULTS
Intra- and inter-assay coefficients of variation were inferior to 20%. No inter-sample contamination was found (p=0.366). The level of concordance between CLIA and the commonly used ELISA method was 88.9%. Positive and negative percent agreements were 92.3% and 85.7%, respectively, with a very good agreement between assays (κ=0.818). Most discordances were due to indeterminate- or negative-ELISA/positive-CLIA results (most of ELISA results were borderline). Linear regression analyses demonstrated a strong correlation between both assays (QF-Ag Pearson's r=0.978, QF-Mg Pearson's r=0.963). No significant difference was observed in median QF-CMV values between both assays (QF-Ag p=0.776; QF-Mg p=0.853; Mann-Whitney U test). The Bland-Altman plots showed a minor difference in IFN-γ release (QF-Ag -0.069 IU/ml, 95% limits of agreement (LoA): -1.589; 1.451; QF-Mg 0.190 IU/ml, 95% LoA: -2.070; 2.450).
CONCLUSION
Automated QF-CMV with CLIA is comparable to QF-CMV performed by ELISA with a presumably higher sensitivity for IFN-γ detection that may result in the conversion of samples close to the ELISA cut-off into positive results. Moreover, the use of a random-access analyzer allows to optimize the follow-up of TR.

Identifiants

pubmed: 37527584
pii: S1386-6532(23)00173-7
doi: 10.1016/j.jcv.2023.105550
pii:
doi:

Substances chimiques

Interferon-gamma 82115-62-6

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

105550

Informations de copyright

Copyright © 2023. Published by Elsevier B.V.

Déclaration de conflit d'intérêts

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Sarah Mafi (S)

French National Reference Center for Herpesviruses, Bacteriology, Virology, Hygiene Department, CHU Limoges, F-87000 Limoges, France; INSERM, RESINFIT, U1092, F-87000, Limoges, France. Electronic address: sarah.mafi@unilim.fr.

Sophie Alain (S)

French National Reference Center for Herpesviruses, Bacteriology, Virology, Hygiene Department, CHU Limoges, F-87000 Limoges, France; INSERM, RESINFIT, U1092, F-87000, Limoges, France.

Sébastien Hantz (S)

French National Reference Center for Herpesviruses, Bacteriology, Virology, Hygiene Department, CHU Limoges, F-87000 Limoges, France; INSERM, RESINFIT, U1092, F-87000, Limoges, France. Electronic address: sebastien.hantz@unilim.fr.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH