Design modifications and evaluation of the silicone artificial finger joints.


Journal

Bio-medical materials and engineering
ISSN: 1878-3619
Titre abrégé: Biomed Mater Eng
Pays: Netherlands
ID NLM: 9104021

Informations de publication

Date de publication:
2023
Historique:
medline: 14 11 2023
pubmed: 7 8 2023
entrez: 7 8 2023
Statut: ppublish

Résumé

There are many reasons that could lead to finger joint arthroplasty, and the most familiar reason is osteoarthritis. Silicone finger joint are the most commonly used implants. However, these implants might fracture with time and cause wear which will lead to chronic inflammation and synovitis for the patient and then implant failure. The aim of this study is to improve the design of the silicone finger joint and simulate the different designs using finite element analysis (FEA) simulation. Three different designs were drawn and FEA has been used in this study using Solidworks software. The first design is the silicone finger joint design without any modification, the second one is modified design with added ribs to the junction of distal stem and hinge and the third design was added filler material inside the body of the artificial joint. An axial force with 625 N that was applied on the upper part of the distal stem which is nearly represents the maximum value of the grip strength for normal males. The results showed improvement on the design in which the concentrated stress at the junction of the distal stem and hinge of the design was distributed. In addition, the Von Mises stress was stable for the modified design with added ribs and the added filler material designs after 15°. The design modification could improve the stress distribution and stability of the artificial finger joint and increase the lifetime expectancy of these implants.

Sections du résumé

BACKGROUND BACKGROUND
There are many reasons that could lead to finger joint arthroplasty, and the most familiar reason is osteoarthritis. Silicone finger joint are the most commonly used implants. However, these implants might fracture with time and cause wear which will lead to chronic inflammation and synovitis for the patient and then implant failure.
OBJECTIVE OBJECTIVE
The aim of this study is to improve the design of the silicone finger joint and simulate the different designs using finite element analysis (FEA) simulation.
METHOD METHODS
Three different designs were drawn and FEA has been used in this study using Solidworks software. The first design is the silicone finger joint design without any modification, the second one is modified design with added ribs to the junction of distal stem and hinge and the third design was added filler material inside the body of the artificial joint. An axial force with 625 N that was applied on the upper part of the distal stem which is nearly represents the maximum value of the grip strength for normal males.
RESULTS RESULTS
The results showed improvement on the design in which the concentrated stress at the junction of the distal stem and hinge of the design was distributed. In addition, the Von Mises stress was stable for the modified design with added ribs and the added filler material designs after 15°.
CONCLUSION CONCLUSIONS
The design modification could improve the stress distribution and stability of the artificial finger joint and increase the lifetime expectancy of these implants.

Identifiants

pubmed: 37545204
pii: BME222527
doi: 10.3233/BME-222527
doi:

Substances chimiques

Silicones 0

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

481-490

Auteurs

F A Alnaimat (FA)

Department of Medical Engineering, Al-Ahliyya Amman University, Amman, Jordan.

H A Owida (HA)

Department of Medical Engineering, Al-Ahliyya Amman University, Amman, Jordan.

Jamal I Al-Nabulsi (JI)

Department of Medical Engineering, Al-Ahliyya Amman University, Amman, Jordan.

B Al-Naami (B)

Department of Biomedical Engineering, Engineering Faculty, The Hashemite University, Zarqa, Jordan.

A Al-Ghraibah (A)

Department of Medical Engineering, Al-Ahliyya Amman University, Amman, Jordan.

M Al-Ayyad (M)

Department of Medical Engineering, Al-Ahliyya Amman University, Amman, Jordan.

Muneera Altayeb (M)

Department of Electronics & Communications Engineering, Al-Ahliyya Amman University, Amman, Jordan.

Ashraf Al Sharah (A)

Department of Electrical Engineering, College of Engineering Technology, Al-Balqa Applied University, Amman, Jordan.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH