Conformational changes during in vitro balloon fracture of internal aortic annuloplasty ring.
HAART 300 annuloplasty ring
aortic annuloplasty ring
aortic valve repair
ring fracture
valve-in-ring
Journal
JTCVS techniques
ISSN: 2666-2507
Titre abrégé: JTCVS Tech
Pays: United States
ID NLM: 101768546
Informations de publication
Date de publication:
Aug 2023
Aug 2023
Historique:
received:
13
01
2023
revised:
14
03
2023
accepted:
18
03
2023
medline:
9
8
2023
pubmed:
9
8
2023
entrez:
9
8
2023
Statut:
epublish
Résumé
HAART 300 300 (BioStable Science and Engineering, Inc) aortic annuloplasty rings restore physiologic annular geometry during aortic valve repair. Transcatheter valve-in-ring implantation is appealing for recurrent valve dysfunction but may necessitate balloon fracture of downsized annuloplasty rings. We characterized the feasibility of ring fracture and changes in ring geometry preceding fracture. The 19-mm, 21-mm, and 23-mm HAART 300 annuloplasty rings were obtained, and 23-mm, 24-mm, 25-mm, and 26-mm valvuloplasty balloons were obtained. Under continuous fluoroscopy and video recording, a 23-mm balloon was inflated within a 19-mm ring at 1 atm/s until ring fracture or balloon failure occurred. If balloon failure occurred, experiments were sequentially repeated with 1-mm upsized balloons until ring fracture occurred or no larger-sized balloons were available. Upon balloon inflation, all rings exhibited an irreversible conformational change from an elliptical, annular geometry to a circular shape with ring posts flaring outward. A 23-mm balloon burst at 21 atm without fracturing the 19-mm ring. The 24-mm balloon fractured the 19-mm ring at 15 atm. Likewise, a 24-mm balloon ruptured at 18 atm without fracturing the 21-mm annuloplasty ring. A 25-mm balloon fractured the 21-mm ring at 18 atm. Finally, a 26-mm balloon burst at 20 atm without fracturing a 23-mm annuloplasty ring, but it did elicit the confirmational changes described. All fractures occurred along the upslope of a ring post. The exposed metal frame was visible after the 21-mm ring fracture. Fracture of HAART 300 aortic annuloplasty rings is possible with an oversized, high-pressure balloon. However, the geometrical changes in the ring and subsequent rupture of its fabric covering may be obstacles to safe, in vivo ring fracture.
Identifiants
pubmed: 37555046
doi: 10.1016/j.xjtc.2023.03.009
pii: S2666-2507(23)00107-4
pmc: PMC10405091
doi:
Types de publication
Journal Article
Langues
eng
Pagination
24-29Informations de copyright
© 2023 The Author(s).
Références
Curr Cardiol Rep. 2014;16(6):490
pubmed: 24743899
Innovations (Phila). 2019 Jun;14(3):199-208
pubmed: 31084444
J Thorac Cardiovasc Surg. 2019 Nov;158(5):1317-1328.e1
pubmed: 30857820
J Card Surg. 2022 Jan;37(1):70-75
pubmed: 34669217
Catheter Cardiovasc Interv. 2021 Oct;98(4):756-764
pubmed: 33991385
Ann Thorac Surg. 2020 Nov;110(5):e375-e376
pubmed: 32376351
Circ Cardiovasc Interv. 2017 Jul;10(7):
pubmed: 28698291
J Card Surg. 2022 Jul;37(7):1947-1956
pubmed: 35384050
Ann Thorac Surg. 2018 Mar;105(3):739-748
pubmed: 29273323
Ann Cardiothorac Surg. 2021 Sep;10(5):564-570
pubmed: 34733685
Ann Thorac Surg. 2020 Mar;109(3):e163-e165
pubmed: 31445912
Interact Cardiovasc Thorac Surg. 2019 Jun 1;28(6):908-915
pubmed: 30649330
Catheter Cardiovasc Interv. 2019 Nov 15;94(6):856-858
pubmed: 30773825
Ann Thorac Surg. 2015 Nov;100(5):1556-62
pubmed: 26212514
Ann Thorac Surg. 2017 Nov;104(5):1501-1508
pubmed: 28669505
JTCVS Tech. 2022 Jun 09;14:55-65
pubmed: 35967205
Eur J Cardiothorac Surg. 2016 Mar;49(3):987-93
pubmed: 26156945