The effects of topping-off instrumentation on biomechanics of sacroiliac joint after lumbosacral fusion.


Journal

Computers in biology and medicine
ISSN: 1879-0534
Titre abrégé: Comput Biol Med
Pays: United States
ID NLM: 1250250

Informations de publication

Date de publication:
09 2023
Historique:
received: 29 04 2023
revised: 02 08 2023
accepted: 12 08 2023
medline: 11 9 2023
pubmed: 17 8 2023
entrez: 16 8 2023
Statut: ppublish

Résumé

Lumbar/lumbosacral fusion supplemented with topping-off devices has been proposed with the aim of avoiding adjacent segment degeneration proximal to the fusion construct. However, it remains unclear how the biomechanics of the sacroiliac joint (SIJ) are altered after topping-off surgery. The objective of this study was to investigate the biomechanical effects of topping-off instrumentation on SIJ after lumbosacral fusion. The validated finite element model of an intact lumbar spine-pelvis segment was modified to simulate L5-S1 interbody fusion fixed with a pedicle screw system. An interspinous spacer, Device for Intervertebral Assisted Motion (DIAM), was used as a topping-off device and placed between interspinous processes of the L4 and L5 segments. Range of motion (ROM), von-Mises stress distribution, and ligament strain at SIJ were compared between fusion (without DIAM) and topping-off (fusion with DIAM) models under moments of four physiological motions. ROM at the left and right SIJs in the topping-off model was higher by 26.9% and 27.5% in flexion, 16.8% and 16.1% in extension, 18.8% and 15.8% in lateral bending, and 3.7% and 7.4% in axial rotation, respectively, compared to those in the fusion model. The predicted stress and strain data showed that under all physiological loads, the topping-off model exhibited higher stress and ligament strain at the SIJs than the fusion model. Motion, stress, and ligament strain at SIJ increase when supplementing lumbosacral fusion with topping-off devices, suggesting that topping-off surgery may be associated with higher risks of SIJ degeneration and pain than fusion alone.

Sections du résumé

BACKGROUND
Lumbar/lumbosacral fusion supplemented with topping-off devices has been proposed with the aim of avoiding adjacent segment degeneration proximal to the fusion construct. However, it remains unclear how the biomechanics of the sacroiliac joint (SIJ) are altered after topping-off surgery. The objective of this study was to investigate the biomechanical effects of topping-off instrumentation on SIJ after lumbosacral fusion.
METHODS
The validated finite element model of an intact lumbar spine-pelvis segment was modified to simulate L5-S1 interbody fusion fixed with a pedicle screw system. An interspinous spacer, Device for Intervertebral Assisted Motion (DIAM), was used as a topping-off device and placed between interspinous processes of the L4 and L5 segments. Range of motion (ROM), von-Mises stress distribution, and ligament strain at SIJ were compared between fusion (without DIAM) and topping-off (fusion with DIAM) models under moments of four physiological motions.
RESULTS
ROM at the left and right SIJs in the topping-off model was higher by 26.9% and 27.5% in flexion, 16.8% and 16.1% in extension, 18.8% and 15.8% in lateral bending, and 3.7% and 7.4% in axial rotation, respectively, compared to those in the fusion model. The predicted stress and strain data showed that under all physiological loads, the topping-off model exhibited higher stress and ligament strain at the SIJs than the fusion model.
CONCLUSIONS
Motion, stress, and ligament strain at SIJ increase when supplementing lumbosacral fusion with topping-off devices, suggesting that topping-off surgery may be associated with higher risks of SIJ degeneration and pain than fusion alone.

Identifiants

pubmed: 37586205
pii: S0010-4825(23)00822-3
doi: 10.1016/j.compbiomed.2023.107357
pii:
doi:

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

107357

Informations de copyright

Copyright © 2023 Elsevier Ltd. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Wei Fan (W)

School of Mechanical Engineering and Automation, Northeastern University, Shenyang, China. Electronic address: fanwei@mail.neu.edu.cn.

Chi Zhang (C)

School of Mechanical Engineering and Automation, Northeastern University, Shenyang, China.

Qing-Dong Wang (QD)

Department of Mechanical Engineering, Tsinghua University, Beijing, China.

Li-Xin Guo (LX)

School of Mechanical Engineering and Automation, Northeastern University, Shenyang, China.

Ming Zhang (M)

Department of Biomedical Engineering, The Hong Kong Polytechnic University, Hong Kong, China; Research Institute for Sports Science and Technology, The Hong Kong Polytechnic University, Hong Kong, China.

Articles similaires

Failed radial head arthroplasty treated by removal of the implant.

Juan Ameztoy Gallego, Blanca Diez Sanchez, Afonso Vaquero-Picado et al.
1.00
Humans Male Female Middle Aged Range of Motion, Articular
Humans Female Male Middle Aged Spinal Fusion
Humans Arthroplasty, Replacement, Shoulder Male Female Retrospective Studies
Humans Arthroplasty, Replacement, Knee Osteoarthritis, Knee Awards and Prizes Biomechanical Phenomena

Classifications MeSH