Short- and Mid-Term Results of Pulmonary Valve Replacement with the Inspiris Valve.
Journal
The Annals of thoracic surgery
ISSN: 1552-6259
Titre abrégé: Ann Thorac Surg
Pays: Netherlands
ID NLM: 15030100R
Informations de publication
Date de publication:
23 Aug 2023
23 Aug 2023
Historique:
received:
19
05
2023
revised:
24
06
2023
accepted:
17
07
2023
pubmed:
26
8
2023
medline:
26
8
2023
entrez:
25
8
2023
Statut:
aheadofprint
Résumé
Various bioprosthetic valves are used off-label for pulmonary valve replacement (PVR), but there is no consensus on whether a particular valve is best for this application. Recently, the Inspiris Resilia valve (Edwards Lifesciences Inc) was approved for aortic valve replacement, and surgeons have begun using it for PVR. There is limited evidence on the performance of the Inspiris valve compared with other valves in the pulmonary position. This study reviewed all patients who underwent PVR with a size 19- to 29-mm Inspiris valve or Mosaic valve (Medtronic Inc) from 2007 to 2022 at Lucile Packard Children's Hospital Stanford (Palo Alto, CA). Midterm outcomes included freedom from moderate or severe pulmonary regurgitation (PR), a maximum Doppler gradient ≥36 mm Hg, and freedom from reintervention. A total of 225 consecutive patients who underwent PVR with a size 19- to 29-mm Mosaic (n = 163) or Inspiris (n = 62) valve were included. There was no difference in baseline characteristics. Early postoperative gradients were low in both groups but higher in the Mosaic cohort, and neither group had more than mild PR on discharge. On univariable and multivariable analysis, Inspiris valves were significantly more likely to develop moderate or greater PR over time. There was no significant difference between the valves in freedom from reintervention or from a maximum gradient ≥36 mm Hg. Early and short-term gradients were similar in patients undergoing PVR with Inspiris and Mosaic valves, but significant PR was more common in patients who received an Inspiris valve. These preliminary findings suggest that the durability of the Inspiris valve in the pulmonary position may not be superior to that of other bioprosthetic valves used for PVR.
Sections du résumé
BACKGROUND
BACKGROUND
Various bioprosthetic valves are used off-label for pulmonary valve replacement (PVR), but there is no consensus on whether a particular valve is best for this application. Recently, the Inspiris Resilia valve (Edwards Lifesciences Inc) was approved for aortic valve replacement, and surgeons have begun using it for PVR. There is limited evidence on the performance of the Inspiris valve compared with other valves in the pulmonary position.
METHODS
METHODS
This study reviewed all patients who underwent PVR with a size 19- to 29-mm Inspiris valve or Mosaic valve (Medtronic Inc) from 2007 to 2022 at Lucile Packard Children's Hospital Stanford (Palo Alto, CA). Midterm outcomes included freedom from moderate or severe pulmonary regurgitation (PR), a maximum Doppler gradient ≥36 mm Hg, and freedom from reintervention.
RESULTS
RESULTS
A total of 225 consecutive patients who underwent PVR with a size 19- to 29-mm Mosaic (n = 163) or Inspiris (n = 62) valve were included. There was no difference in baseline characteristics. Early postoperative gradients were low in both groups but higher in the Mosaic cohort, and neither group had more than mild PR on discharge. On univariable and multivariable analysis, Inspiris valves were significantly more likely to develop moderate or greater PR over time. There was no significant difference between the valves in freedom from reintervention or from a maximum gradient ≥36 mm Hg.
CONCLUSIONS
CONCLUSIONS
Early and short-term gradients were similar in patients undergoing PVR with Inspiris and Mosaic valves, but significant PR was more common in patients who received an Inspiris valve. These preliminary findings suggest that the durability of the Inspiris valve in the pulmonary position may not be superior to that of other bioprosthetic valves used for PVR.