Pulse-sampling fluorescence lifetime imaging: evaluation of photon economy.


Journal

Optics letters
ISSN: 1539-4794
Titre abrégé: Opt Lett
Pays: United States
ID NLM: 7708433

Informations de publication

Date de publication:
01 Sep 2023
Historique:
medline: 1 9 2023
pubmed: 1 9 2023
entrez: 1 9 2023
Statut: ppublish

Résumé

This Letter presents an experimental study comparing the photon rate and photon economy of pulse sampling fluorescence lifetime imaging (PS-FLIm) with the conventional time-correlated single photon counting (TCSPC) technique. We found that PS-FLIm has a significantly higher photon detection rate (200 MHz) compared with TCSPC (2-8 MHz) but lower photon economy (4-5 versus 1-1.3). The main factor contributing to the lower photon economy in PS-FLIm is laser pulse variability. These results demonstrate that PS-FLIm offers 25× faster imaging speed than TCSPC while maintaining room light rejection in clinical settings. This makes PS-FLIm a robust technique for clinical applications.