Evaluation of Quantitative Hepatitis C Virus Core Antigen Testing as a Tool for Screening and Confirmation of HCV Infection.
Journal
Clinical laboratory
ISSN: 1433-6510
Titre abrégé: Clin Lab
Pays: Germany
ID NLM: 9705611
Informations de publication
Date de publication:
01 Sep 2023
01 Sep 2023
Historique:
medline:
14
9
2023
pubmed:
13
9
2023
entrez:
13
9
2023
Statut:
ppublish
Résumé
Hepatitis C is a common viral infection worldwide. Finding the most effective diagnostic methods with low cost is always needed for laboratory improvement. In this study, we evaluated the performance of a quantitative chemiluminescent hepatitis C virus core antigen (HCV cAg) test by comparing it with the HCV confirmatory antibody line immunoblot assay (HCV Ab-LIA) test as well as the HCV quantitate reverse transcription-polymerase chain reaction (qRT-PCR) test. A total of 394 samples were enrolled in the retrospective study. Of these, 225 samples were tested using HCV Ab screening and confirmatory Ab-LIA along with chemiluminescent HCV core Ag testing, while 169 samples were tested using qRT-PCR for HCV RNA and chemiluminescent HCV core Ag testing. Out of these, 225 positive samples tested by HCV Ab screening test were analyzed using the confirmatory Ab LIA and HCV cAg assays, a total of 183 samples (81.3 %) were confirmed to be Ab-positive, and among those, 77 samples (42.1%) were also positive for HCV cAg. Thirty-eight samples (20.76%) were HCV Ab indeterminate, and all of them were HCV cAg negative. Four samples (1.8%) were HCV Ab LIA-negative and negative for HCV cAg. Moreover, 169 samples were measured for qRT-PCR HCV viral load and quantitative HCV cAg test. One hundred and three samples were positive for HCV RNA, while 66 were negative. Among the positives, 96/103 samples were HCV cAg positive and 7/103 samples were negative. Out of the negatives, 4/66 samples were HCV cAg positive but 62/66 samples were negative. The HCV cAg results were concordant with the qRT-PCR results in 158 samples (93.5%); however, 11 samples (6.5%) were found to be discrepant. The sensitivity, specificity, positive predictive value, and negative predictive value of the quantitative HCV cAg were found to be 93%, 94%, 96%, and 90%, respectively. The overall coefficient of correlation between the HCV RNA levels and HCV cAg data was determined to be r2 = 0.9. The HCV cAg test showed a high correlation with the HCV RNA levels and may potentially be used as a more cost-effective alternative to the HCV RNA qRT-PCR test.
Sections du résumé
BACKGROUND
BACKGROUND
Hepatitis C is a common viral infection worldwide. Finding the most effective diagnostic methods with low cost is always needed for laboratory improvement. In this study, we evaluated the performance of a quantitative chemiluminescent hepatitis C virus core antigen (HCV cAg) test by comparing it with the HCV confirmatory antibody line immunoblot assay (HCV Ab-LIA) test as well as the HCV quantitate reverse transcription-polymerase chain reaction (qRT-PCR) test.
METHODS
METHODS
A total of 394 samples were enrolled in the retrospective study. Of these, 225 samples were tested using HCV Ab screening and confirmatory Ab-LIA along with chemiluminescent HCV core Ag testing, while 169 samples were tested using qRT-PCR for HCV RNA and chemiluminescent HCV core Ag testing.
RESULTS
RESULTS
Out of these, 225 positive samples tested by HCV Ab screening test were analyzed using the confirmatory Ab LIA and HCV cAg assays, a total of 183 samples (81.3 %) were confirmed to be Ab-positive, and among those, 77 samples (42.1%) were also positive for HCV cAg. Thirty-eight samples (20.76%) were HCV Ab indeterminate, and all of them were HCV cAg negative. Four samples (1.8%) were HCV Ab LIA-negative and negative for HCV cAg. Moreover, 169 samples were measured for qRT-PCR HCV viral load and quantitative HCV cAg test. One hundred and three samples were positive for HCV RNA, while 66 were negative. Among the positives, 96/103 samples were HCV cAg positive and 7/103 samples were negative. Out of the negatives, 4/66 samples were HCV cAg positive but 62/66 samples were negative. The HCV cAg results were concordant with the qRT-PCR results in 158 samples (93.5%); however, 11 samples (6.5%) were found to be discrepant. The sensitivity, specificity, positive predictive value, and negative predictive value of the quantitative HCV cAg were found to be 93%, 94%, 96%, and 90%, respectively. The overall coefficient of correlation between the HCV RNA levels and HCV cAg data was determined to be r2 = 0.9.
CONCLUSIONS
CONCLUSIONS
The HCV cAg test showed a high correlation with the HCV RNA levels and may potentially be used as a more cost-effective alternative to the HCV RNA qRT-PCR test.