Quantification of Free Immunoglobulin Light Chains in Urine.
Journal
The journal of applied laboratory medicine
ISSN: 2576-9456
Titre abrégé: J Appl Lab Med
Pays: England
ID NLM: 101693884
Informations de publication
Date de publication:
02 11 2023
02 11 2023
Historique:
received:
23
03
2023
accepted:
26
06
2023
medline:
8
11
2023
pubmed:
20
9
2023
entrez:
19
9
2023
Statut:
ppublish
Résumé
The serum-free immunoglobulin light chain assay has been recommended as a screening test for monoclonal gammopathy. We evaluated the usefulness of urine free immunoglobulin light concentration for selection of specimens for immunofixation electrophoresis. Using kits from The Binding Site for Freelite ®, we validated examination of urine for measuring free κ and λ light chains. The results of urine free light chain concentrations were evaluated to ascertain if the results could be used to reduce the number of specimens requiring urine protein immunofixation electrophoresis. In the 515 specimens examined, there was no evidence of monoclonal gammopathy or history of monoclonal gammopathy in 331. Monoclonal κ or λ light chains were detectable in 42 and 30 specimens, respectively. There was history of κ or λ chain associated monoclonal gammopathy in 62 and 50 patients, respectively. In the 38 monoclonal κ positive urine specimens, with light chain data, κ/λ ratio was >5.83 in all specimens. In 27 specimens positive for monoclonal λ light chains, with light chain data, the urine λ/κ ratio was > 0.17 in 24 of 27 specimens and > 0.041 in all specimens. In patients without monoclonal gammopathy all specimens had a κ/λ ratio of >5.83 or λ/κ ratio >0.17. The Freelite ® assay from The Binding Site is suitable for quantification of free light chains in urine. In patients with known history of monoclonal gammopathy, urine immunofixation electrophoresis may be omitted in specimens with κ/λ ratio of <5.83 for κ associated lesions and λ/κ ratio of <0.041 for λ associated lesions. However, the results do not support using this test for first-time urine testing for monoclonal light chains as it is not predictive of positive result, nor does it exclude a monoclonal light chain in urine.
Sections du résumé
BACKGROUND
The serum-free immunoglobulin light chain assay has been recommended as a screening test for monoclonal gammopathy. We evaluated the usefulness of urine free immunoglobulin light concentration for selection of specimens for immunofixation electrophoresis.
METHODS
Using kits from The Binding Site for Freelite ®, we validated examination of urine for measuring free κ and λ light chains. The results of urine free light chain concentrations were evaluated to ascertain if the results could be used to reduce the number of specimens requiring urine protein immunofixation electrophoresis.
RESULTS
In the 515 specimens examined, there was no evidence of monoclonal gammopathy or history of monoclonal gammopathy in 331. Monoclonal κ or λ light chains were detectable in 42 and 30 specimens, respectively. There was history of κ or λ chain associated monoclonal gammopathy in 62 and 50 patients, respectively. In the 38 monoclonal κ positive urine specimens, with light chain data, κ/λ ratio was >5.83 in all specimens. In 27 specimens positive for monoclonal λ light chains, with light chain data, the urine λ/κ ratio was > 0.17 in 24 of 27 specimens and > 0.041 in all specimens. In patients without monoclonal gammopathy all specimens had a κ/λ ratio of >5.83 or λ/κ ratio >0.17.
CONCLUSIONS
The Freelite ® assay from The Binding Site is suitable for quantification of free light chains in urine. In patients with known history of monoclonal gammopathy, urine immunofixation electrophoresis may be omitted in specimens with κ/λ ratio of <5.83 for κ associated lesions and λ/κ ratio of <0.041 for λ associated lesions. However, the results do not support using this test for first-time urine testing for monoclonal light chains as it is not predictive of positive result, nor does it exclude a monoclonal light chain in urine.
Identifiants
pubmed: 37725944
pii: 7277362
doi: 10.1093/jalm/jfad055
doi:
Substances chimiques
Immunoglobulin Light Chains
0
Immunoglobulin lambda-Chains
0
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
1101-1114Informations de copyright
© Association for Diagnostics & Laboratory Medicine 2023.